Airway and body surface sensors for triggering in neonatal ventilation.

J John1, L J Björklund, N W Svenningsen, B Jonson.   

Abstract

Failure of neonatal patient triggered ventilation may reflect a delay in delivery of flow relative to the inspiratory effort of the infant. Transmission of diaphragmatic contraction to the sensor site (patient delay) and further transmission to and within the sensing device (device delay) both contribute to the delay in triggering. Patient and device delays were studied for different sensing systems in 36 infants, 24 of whom were intubated. Device delay was long (> 40 ms) with a conventional apnoea monitor compared with sensors placed at the airway opening (2 ms), the inspiratory (12 ms) and expiratory (3 ms) pressure transducers of the ventilator, the Graseby capsule (8 ms), strain gauges (3 ms) and oesophageal pressure (6 ms). In near normal infants, the sum of patient and device delays for the latter sensors was less than 20 ms and a minor component of the total delay. However, in severe lung disease the total delay may be more than 100 ms even for airway sensors.

Entities:  

Mesh:

Year:  1994        PMID: 7819683     DOI: 10.1111/j.1651-2227.1994.tb13168.x

Source DB:  PubMed          Journal:  Acta Paediatr        ISSN: 0803-5253            Impact factor:   2.299


  4 in total

Review 1.  Aerosol delivery to ventilated newborn infants: historical challenges and new directions.

Authors:  Jan Mazela; Richard A Polin
Journal:  Eur J Pediatr       Date:  2010-09-28       Impact factor: 3.183

2.  Accuracy of volume measurements in mechanically ventilated newborns: a comparative study of commercial devices.

Authors:  K Roske; B Foitzik; R R Wauer; G Schmalisch
Journal:  J Clin Monit Comput       Date:  1998-08       Impact factor: 2.502

3.  Comparison of NIV-NAVA and NCPAP in facilitating extubation for very preterm infants.

Authors:  Byoung Kook Lee; Seung Han Shin; Young Hwa Jung; Ee-Kyung Kim; Han-Suk Kim
Journal:  BMC Pediatr       Date:  2019-08-28       Impact factor: 2.125

Review 4.  Synchronized mechanical ventilation for respiratory support in newborn infants.

Authors:  Anne Greenough; Thomas E Rossor; Adesh Sundaresan; Vadivelam Murthy; Anthony D Milner
Journal:  Cochrane Database Syst Rev       Date:  2016-09-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.