| Literature DB >> 27279013 |
Charlotte Verbeek1, Henriëtte A van Zanten1, Jeroen J van Vonderen1, Marcus J Kitchen2, Stuart B Hooper3, Arjan B Te Pas4.
Abstract
UNLABELLED: This study aimed to test the accuracy in volume measurements of three available respiratory function monitors (RFMs) for neonatal resuscitation and the effect of changing gas conditions. The Florian, New Life Box Neo-RSD (NLB Neo-RSD) and NICO RFM were tested on accuracy with volumes of 10 and 20 mL and on changes in volume measurements under changing gas conditions (oxygen level 21-100 % and from cold dry air (24 ± 2 °C) to heated humidified air (37 °C). Volume differences >10 % were considered clinically relevant. We found that the mean (SD) volume difference was clinically acceptable for all devices (10, 20 mL): Florian (+8.4 (1.2)%, +8.4 (0.5)%); NLB Neo-RSD (+5.8 (1.1)%, +4.3 (1.4)%); and NICO (-8.2 (0.9)%, -8.7 (0.8)%). Changing from cold dry to heated humidified air increased the volume difference using the Florian (cold dry air, heated humidified air (+5.2 (1.2)%, +12.2 (0.9)%) but not NLB Neo-RSD (+2.0(1.6)%, +3.4(2.8)%) and NICO (-2.3 % (0.8), +0.1 (0.6)%). Similarly, when using heated humidified air, increasing oxygen enlarged increased the volume difference using the Florian (oxygen 21 %, 100 %: +12.2(1.0)%, +19.8(1.1)%), but not NLB Neo-RSD (+0.2(1.9)%, +1.1(2.8)%) and NICO (-5.6(0.9)%, -3.7(0.9)%). Clinically relevant changes occurred when changing both gas conditions (Florian +25.7(1.7)%; NLB Neo-RSD +3.8(2.4)%; NICO -5.7(1.4)%).Entities:
Keywords: Neonatology; Respiratory function monitor; Resuscitation; Tidal volume
Mesh:
Year: 2016 PMID: 27279013 PMCID: PMC4930469 DOI: 10.1007/s00431-016-2739-1
Source DB: PubMed Journal: Eur J Pediatr ISSN: 0340-6199 Impact factor: 3.183
Fig. 1The experimental set-up to determine the accuracy of the three RFMs
Fig. 2The experimental set-up to test the deviation under different gas conditions
Mean (SD) deviation when measuring 10 and 20 mL in percentages (mean (SD))
| RFM | 10 mL deviation Vt (%) | 20 mL deviation Vt (%) |
|---|---|---|
| Florian | 8.4 (1.2) | 8.4 (0.5) |
| NLB Neo-RSD | 5.8 (1.1) | 3.8 (0.8) |
| NICO | −8.2 (0.9) | −8.7 (0.8) |
Mean (SD) change in percentage during changing gas conditions
| Change | FiO2 0.21–1.0 | 24 ± 2 °C → 37 °C | FiO2 0.21–1.0 | ||
|---|---|---|---|---|---|
| Cold dry air | Heated humidified air | FiO2 0.21 | FiO2 1.0 | ||
| Florian | 5.2 (1.2) % | 12.2 (0.9) % | 12.2 (1.0) % | 19.8 (1.1) % | 25.7 (1.7) % |
| NLB | 2.0 (1.6) % | 3.4 (2.8) % | 0.2 (1.9) % | 1.1 (2.8) % | 3.8 (2.4) % |
| NICO | −2.3 (0.8) % | 0.1 (0.6) % | −5.6 (0.9) % | −3.7 (0.9) % | −5.7 (1.4) % |
|
|
| • |
|
|
| • |