Literature DB >> 9917597

Failure of total hip arthroplasty with a precoated prosthesis. 4- to 11-year results.

J E Dowd1, C W Cha, S Trakru, S Y Kim, I H Yang, H E Rubash.   

Abstract

A series of early femoral component failures prompted a detailed retrospective clinical and radiographic review of 176 hybrid cemented total hip arthroplasties using a polymethyl-methacrylate coated femoral prosthesis. All surgeries were performed using third generation cement techniques. Average length of followup was 6.3 years (range, 3-12 years). Twenty-one patients died, and one underwent revision surgery because of sepsis. Of the remaining 154 total hip arthroplasties, 23 (15%) of the femoral components failed (21 revised, two definitely loose). The average time to revision was 3.9 years. None of the acetabular components failed. Comparison between the failure and nonfailure groups revealed that poor cement mantles (Grades C or D) with distal cement mantle deficiencies were statistically significant predictors of femoral failure. The most common mechanism of failure was progressive, circumferential cement-bone interface osteolysis with relative preservation of the cement-metal interface. Debonding of the cement column from the prosthesis was a late finding and occurred in only 45% of failed cases. Incorporating the techniques of centralization and centrifugation significantly improved clinical results. Strengthening of the cement-prosthesis interface may magnify the deleterious effects of a poor cement mantle and predisposes the cement-bone interface to failure.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9917597     DOI: 10.1097/00003086-199810000-00013

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  5 in total

1.  High failure rate of a modern, proximally roughened, cemented stem for total hip arthroplasty.

Authors:  A Grose; A González Della Valle; P Bullough; S Lyman; I Tomek; P Pellicci
Journal:  Int Orthop       Date:  2006-04-29       Impact factor: 3.075

2.  Favourable mid-term results of the VerSys CT polished cemented femoral stem for total hip arthroplasty.

Authors:  Alejandro González Della Valle; Fernando Comba; Adriana Zoppi; Eduardo A Salvati
Journal:  Int Orthop       Date:  2006-03-31       Impact factor: 3.075

3.  Functional interface micromechanics of 11 en-bloc retrieved cemented femoral hip replacements.

Authors:  Kenneth A Mann; Mark A Miller; Nico Verdonschot; Timothy H Izant; Amos Race
Journal:  Acta Orthop       Date:  2010-06       Impact factor: 3.717

4.  Fifteen-year results of precoated femoral stem in primary hybrid total hip arthroplasty.

Authors:  Dong Hun Suh; Ho Hyun Yun; Sung Kwang Chun; Won Yong Shon
Journal:  Clin Orthop Surg       Date:  2013-05-15

5.  Massive femoral osteolysis secondary to loosening of a cemented roughened long stem: a case report.

Authors:  Yasuaki Tamaki; Tomohiro Goto; Daisuke Hamada; Toshihiko Nishisho; Kiminori Yukata; Suzue Naoto; Hiroshi Egawa; Koichi Sairyo
Journal:  Case Rep Orthop       Date:  2014-06-23
  5 in total