Cochlear implants: in vitro investigation of electromagnetic interference at MR imaging--compatibility and safety aspects.

C Teissl1, C Kremser, E S Hochmair, I J Hochmair-Desoyer.   

Abstract

In vitro experiments were performed to determine the compatibility of a cochlear implant at 0.2- and 1.5-T magnetic resonance (MR) imaging. Except for the torque at 1.5 T, all electromagnetic interferences remained within acceptable limits. MR imaging should be performed only if there is a strong medical indication, by following appropriate safety procedures. MR imaging at 0.2 T should be safe; at 1.5 T, however, the relative risks will have to be assessed.

Mesh:

Year:  1998        PMID: 9722849     DOI: 10.1148/radiology.208.3.9722849

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  12 in total

1.  Noninvasive direct stimulation of the cochlear nerve for functional MR imaging of the auditory cortex.

Authors:  E Hofmann; C Preibisch; C Knaus; J Müller; C Kremser; C Teissl
Journal:  AJNR Am J Neuroradiol       Date:  1999 Nov-Dec       Impact factor: 3.825

Review 2.  Is magnetic resonance imaging still a contraindication in cochlear-implanted patients?

Authors:  Sebastien Schmerber; Emile Reyt; Jean-Pierre Lavieille
Journal:  Eur Arch Otorhinolaryngol       Date:  2003-01-16       Impact factor: 2.503

3.  Magnetic resonance imaging with the Digisonic SP Neurelec cochlear implant.

Authors:  C Vincent; I Ruzza; F M Vaneecloo; F Dubrulle
Journal:  Eur Arch Otorhinolaryngol       Date:  2008-02-23       Impact factor: 2.503

4.  Artifacts caused by cochlear implants with non-removable magnets in 3T MRI: phantom and cadaveric studies.

Authors:  Omid Majdani; Thomas S Rau; Friedrich Götz; Martin Zimmerling; Minoo Lenarz; Thomas Lenarz; Robert Labadie; Martin Leinung
Journal:  Eur Arch Otorhinolaryngol       Date:  2009-07-23       Impact factor: 2.503

5.  Cochlear implant with a non-removable magnet: preliminary research at 3-T MRI.

Authors:  F Dubrulle; A Sufana Iancu; C Vincent; G Tourrel; O Ernst
Journal:  Eur Radiol       Date:  2013-01-10       Impact factor: 5.315

Review 6.  [Magnetic resonance imaging in patients with magnetic hearing implants: overview and procedural management].

Authors:  S Nospes; W Mann; A Keilmann
Journal:  Radiologe       Date:  2013-11       Impact factor: 0.635

7.  Cortical activation patterns to spatially presented pure tone stimuli with different intensities measured by functional near-infrared spectroscopy.

Authors:  Günther Bauernfeind; Selina C Wriessnegger; Sabine Haumann; Thomas Lenarz
Journal:  Hum Brain Mapp       Date:  2018-03-08       Impact factor: 5.038

8.  The High Rate CIS Auditory Brainstem Implant for Restoration of Hearing in NF-2 Patients.

Authors:  Robert Behr; Joachim Müller; Wafaa Shehata-Dieler; Hans-Peter Schlake; Jan Helms; Klaus Roosen; Norfrid Klug; Bernd Hölper; Artur Lorens
Journal:  Skull Base       Date:  2007-03

Review 9.  The cochlear implant: historical aspects and future prospects.

Authors:  Adrien A Eshraghi; Ronen Nazarian; Fred F Telischi; Suhrud M Rajguru; Eric Truy; Chhavi Gupta
Journal:  Anat Rec (Hoboken)       Date:  2012-10-08       Impact factor: 2.064

Review 10.  Safety of active auditory implants in magnetic resonance imaging.

Authors:  Guy Fierens; Nina Standaert; Ronald Peeters; Christ Glorieux; Nicolas Verhaert
Journal:  J Otol       Date:  2021-01-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.