First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose.

R Boellaard1, M van Herk, H Uiterwaal, B Mijnheer.   

Abstract

BACKGROUND AND
PURPOSE: Recently, algorithms have been developed to derive the patient dose from portal dose measurements using a liquid-filled electronic portal imaging device. These algorithms have already been validated for several phantom geometries irradiated under clinical conditions. It is the aim of the present study to investigate the applicability of a liquid-filled electronic portal imaging device in combination with these algorithms for two-dimensional midplane dose verification in clinical practice. MEASUREMENTS AND METHODS: Portal dose images were obtained during several patient treatments under routine clinical conditions. Measurements were performed to verify the midplane dose during radiotherapy of larynx cancer with 4 MV beams, breast and lung cancer with 8 MV beams and prostate cancer with both 8 and 18 MV beams. Midplane doses, determined from portal dose measurements and analyzed with our algorithms, were compared with midplane doses calculated with our three-dimensional (3D) treatment planning system (TPS).
RESULTS: For the larynx treatment the measured 2D midplane dose agreed within 2.0% with TPS calculations in most parts of the field. Larger differences were found in a small region below the skin due to the absence of electron equilibrium, which is not taken into account in our portal dose analysis. For breast irradiations the measured midplane dose showed a homogeneous distribution in the AP direction in the axial plane, while high dose regions were observed in the cranial and caudal part of the breast. Portal dose measurements and TPS calculations agreed within 2.5% for most of the prostate and lung irradiations. For a few of the prostate and lung treatments larger local differences were found due to differences between the actual patient anatomy and the planning CT data, e.g. as a result of variable gas filling in the rectum and anatomical changes in the lung.
CONCLUSIONS: Portal dose measurements with a liquid-filled electronic portal imaging device can be used to determine the 2D midplane dose for various treatment sites in clinical practice. Portal in vivo dosimetry has proven to be important in detecting changes in the patient's anatomy and its influence on the dose delivery. It is concluded that portal dosimetry is an excellent tool for accurate and independent verification of the dose in the entire (2D) midplane during patient treatment. However, a limited number of patients were involved in this study and the results are therefore preliminary. More research is needed to fully assess the clinical value of portal dose measurements.

Entities:  

Mesh:

Year:  1998        PMID: 9681895     DOI: 10.1016/s0167-8140(98)00008-5

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  10 in total

1.  In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy.

Authors:  Angelo Piermattei; Andrea Fidanzio; Luigi Azario; Francesca Greco; Alessandra Mameli; Savino Cilla; Luca Grimaldi; Guido D'Onofrio; Boris Giuseppe Augelli; Gerardina Stimato; Diego Gaudino; Sara Ramella; Rolando D'Angelillo; Francesco Cellini; Lucio Trodella
Journal:  Med Biol Eng Comput       Date:  2009-02-17       Impact factor: 2.602

2.  Tomotherapy as a tool in image-guided radiation therapy (IGRT): theoretical and technological aspects.

Authors:  S Yartsev; T Kron; J Van Dyk
Journal:  Biomed Imaging Interv J       Date:  2007-01-01

3.  Evaluation of relative transmitted dose for a step and shoot head and neck intensity modulated radiation therapy using a scanning liquid ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak
Journal:  J Med Phys       Date:  2012-01

4.  Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification.

Authors:  Ali Jomehzadeh; Parvaneh Shokrani; Mohammad Mohammadi; Alireza Amouheidari
Journal:  Adv Biomed Res       Date:  2016-11-28

5.  Guide to clinical use of electronic portal imaging.

Authors:  M G Herman; J J Kruse; C R Hagness
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

6.  Breast in vivo dosimetry by EPID.

Authors:  Andrea Fidanzio; Francesca Greco; Alessandra Mameli; Luigi Azario; Mario Balducci; Maria Antonietta Gambacorta; Vincenzo Frascino; Savino Cilla; Domenico Sabatino; Angelo Piermattei
Journal:  J Appl Clin Med Phys       Date:  2010-09-02       Impact factor: 2.102

7.  Electronic Portal Imaging Device-Based Three-Dimensional Volumetric Dosimetry for Intensity-modulated Radiotherapy Pretreatment Quality Assurance.

Authors:  Manikandan Arjunan; Sureka Chandra Sekaran; Biplab Sarkar; Saran Kumar Manavalan
Journal:  J Med Phys       Date:  2019 Jul-Sep

8.  The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak; Paul Reich
Journal:  J Appl Clin Med Phys       Date:  2006-05-15       Impact factor: 2.102

9.  Comparison of measured and computed portal dose for IMRT treatment.

Authors:  Savino Cilla; Pietro Viola; Luigi Azario; Luca Grimaldi; Maurizio Craus; Guido D'Onofrio; Andrea Fidanzio; Alessio Giuseppe Morganti; Angelo Piermattei
Journal:  J Appl Clin Med Phys       Date:  2006-08-24       Impact factor: 2.102

10.  Using the frame averaging of aS500 EPID for IMRT verification.

Authors:  J Chang; C C Ling
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.