| Literature DB >> 21081886 |
Andrea Fidanzio1, Francesca Greco, Alessandra Mameli, Luigi Azario, Mario Balducci, Maria Antonietta Gambacorta, Vincenzo Frascino, Savino Cilla, Domenico Sabatino, Angelo Piermattei.
Abstract
An electronic portal imaging device (EPID) is an effective detector for in vivo transit dosimetry. In fact, it supplies two-dimensional information, does not require special efforts to be used during patient treatment, and can supply data in real time. In the present paper, a new procedure has been proposed to improve the EPID in vivo dosimetry accuracy by taking into account the patient setup variations. The procedure was applied to the breast tangential irradiation for the reconstruction of the dose at the breast midpoint, Dm. In particular, the patient setup variations were accounted for by comparing EPID images versus digitally reconstructed radiographies. In this manner, EPID transit signals were obtained corresponding to the geometrical projections of the breast midpoint on the EPID for each therapy session. At the end, the ratios R between D(m) and the doses computed by the treatment planning system (TPS) at breast midpoints, D(m,TPS), were determined for 800 therapy sessions of 20 patients. Taking into account the method uncertainty, tolerance levels equal to ± 5% have been determined for the ratio R.The improvement of in vivo dosimetry results obtained (taking into account patient misalignment) has been pointed out comparing the R values obtained with and with-out considering patient setup variations. In particular, when patient misalignments were taken into account, the R values were within ± 5% for 93% of the checks; when patient setup variations were not taken into account, the R values were within ± 5% in 72% of the checks. This last result points out that the transit dosimetry method overestimates the dose discrepancies if patient setup variations are not taken into account for dose reconstruction. In this case, larger tolerance levels have to be adopted as a trade-off between workload and ability to detect errors, with the drawback being that some errors (such as the ones in TPS implementation or in beam calibration) cannot be detected, limiting the in vivo dosimetry efficacy.The paper also reports preliminary results about the possibility of reconstructing a dose profile perpendicular to the beam central axis reaching from the apex to the lung and passing through the middle point of the breast by an algorithm, similar to the one used for dose reconstruction at breast midpoint. In particular, the results have shown an accuracy within ± 3% for the dose profile reconstructed in the breast (excluding the interface regions) and an underestimation of the lung dose.Entities:
Mesh:
Year: 2010 PMID: 21081886 PMCID: PMC5720411 DOI: 10.1120/jacmp.v11i4.3275
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Experimental set‐up used to determine the correlation function F(w) and the dose profile along the segment . The cylindrical phantom section shown contains the mid‐point P of the chord w positioned at the SAD along the beam's central axis (dotted line) and the point at the intersection between the segment and the chord . The transit signals and were determined at the geometrical intersection between beam axis and the fan line (dashed line) and the EPID respectively. The continuous lines represent the geometrical field edges.
Figure 2Scheme for dose reconstruction at points situated inside water and inside a low density tissue (shadowed area). P is the mid‐point on the chord along the beam's central axis B (dotted line). The P' points are at the half radiological thickness along fan lines A and C (dashed lines). The points represent the intersection between fan lines A and C and the segment ; d is the distance between P and the SAD line; is the distance between P' and the SAD line; and is the distance between P' and . The transit signals and were determined at the geometrical intersection between beam axis and the fan line (dashed line) and the EPID respectively. The continuous lines represent the contours of the geometrical field.
Figure 3Portal image of the medial beam acquired for patient #7 with correct set‐up (a); portal image of the medial beam acquired when the patent's misalignment was 2 cm (b). profiles (c) obtained along the ED line of image 3(a) (continuous line) and of image 3(b) (dashed line); (d) the same profiles of aligned with respect to the DRR.
Coefficients of the correlation function .
|
|
|
|
|---|---|---|
|
|
|
|
(r) values obtained for the four cylindrical phantoms.
|
|
|
|
|
|
|---|---|---|---|---|
|
|
|
|
|
|
Figure 4Planned dose profile (continuous line) and average reconstructed dose profile (□); the error bars are representative of the average standard deviation (3%) of the dose points along the profile. Reconstructed dose profile (Δ) and TPS dose profile (dashed line) obtained for the therapy session with a misalignment of 2 cm ((obtained simulating the patient displacement by the TPS). The points at distances of less than 1 cm from the interfaces were not considered for the dose reconstruction.
Figure 5Frequency distribution (a) of the R ratios obtained without taking into account the patient set‐up variations for the 800 DPIs examined; frequency distribution (b) of the R ratios obtained aligning the DPI versus the DRR for the same in‐vivo dosimetry checks as shown in Fig. 5(a).