Literature DB >> 9659238

The importance of "like to like" ISI calibrations with freeze dried plasmas. European Concerted Action on Anticoagulation.

L Poller1, A M van den Besselaar, J Jespersen, A Tripodi, D Houghton.   

Abstract

AIM: To assess reliability of like to like and cross species calibrations using two types of certified freeze dried plasma calibrants--artificially depleted of vitamin K clotting factors, and from coumarin treated patients.
METHODS: Six ECAA national control laboratories provided certified values for the freeze dried plasmas in terms of the human plain international reference preparation (IRP) (BCT/441) with the manual prothrombin time technique. Eight other ECAA national laboratories determined international sensitivity index (ISI) values in full fresh plasma same species and cross species WHO calibrations against a low ISI human IRP (BCT/441) of the ECAA low ISI human thromboplastin and high ISI ECAA rabbit thromboplastin. Parallel calibrations were performed using the certified values.
RESULTS: Calibrations on fresh plasmas of the human ECAA reference thromboplastin (stated ISI = 0.95) gave ISI of 0.957 against the human IRP and 1.66 against the rabbit IRP. The ECAA rabbit (stated ISI = 1.67) gave an identical value on the fresh plasma calibration v the human IRP. With freeze dried depleted plasmas certified in terms of the human IRP, the ISI of the ECAA human was 1.01, but the ECAA rabbit (stated ISI = 1.67) gave a low ISI of 1.47. The freeze dried coumarin plasmas gave an ISI of 0.943 for the ECAA human but only 1.493 for the ECAA rabbit.
CONCLUSIONS: Fresh plasmas give reliable ISI when calibrating thromboplastins in same species and cross species calibrations. Freeze dried plasmas certified in terms of a single IRP, whether artificially depleted or of coumarin plasma origin, cannot be used for calibration of dissimilar thromboplastins.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9659238      PMCID: PMC500669          DOI: 10.1136/jcp.51.4.275

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  9 in total

1.  Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.

Authors:  K Clarke; D A Taberner; J M Thomson; J A Morris; L Poller
Journal:  J Clin Pathol       Date:  1992-01       Impact factor: 3.411

2.  Minimum lyophilized plasma requirement for ISI calibration. European Concerted Action on Anticoagulation.

Authors:  L Poller; T W Barrowcliffe; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Am J Clin Pathol       Date:  1998-02       Impact factor: 2.493

3.  European Concerted Action on Anticoagulation (ECAA)-the multicentre calibration of rabbit and human ECAA reference thromboplastins. Steering Committee.

Authors:  L Poller; T W Barrowcliffe; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Thromb Haemost       Date:  1996-12       Impact factor: 5.249

4.  A proposed scheme for calibration of international reference preparations of thromboplastin for the prothrombin time. On behalf of the Subcommittee on Control of Anticoagulation.

Authors:  A Tripodi; L Poller; A M van den Besselaar; P M Mannucci
Journal:  Thromb Haemost       Date:  1995-11       Impact factor: 5.249

5.  A comparison of lyophilized artificially depleted plasmas and lyophilized plasmas from patients receiving warfarin in correcting for coagulometer effects on international normalized ratios.

Authors:  L Poller; D A Triplett; J Hirsh; J Carroll; K Clarke
Journal:  Am J Clin Pathol       Date:  1995-03       Impact factor: 2.493

6.  A collaborative calibration study of reference materials for thromboplastins.

Authors:  J Hermans; A M van den Besselaar; E A Loeliger; E A van der Velde
Journal:  Thromb Haemost       Date:  1983-10-31       Impact factor: 5.249

7.  The correction of coagulometer effects on international normalized ratios: a multicentre evaluation.

Authors:  L Poller; J M Thomson; D A Taberner; D K Clarke
Journal:  Br J Haematol       Date:  1994-01       Impact factor: 6.998

8.  Use of lyophilized calibration plasmas and control blood for international normalized ratio calculation in external quality assessment of the prothrombin time.

Authors:  A M van den Besselaar
Journal:  Am J Clin Pathol       Date:  1994-07       Impact factor: 2.493

9.  Results of a multicenter study assessing the status of standardization of a recombinant thromboplastin for the control of oral anticoagulant therapy.

Authors:  A Tripodi; V Chantarangkul; M Braga; L Poller; J W ten Cate; A M van den Besselaar; P M Mannucci
Journal:  Thromb Haemost       Date:  1994-08       Impact factor: 5.249

  9 in total
  1 in total

1.  European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors.

Authors:  L Poller; M Keown; N Chauhan; A M H P van den Besselaar; A Tripodi; C Shiach; J Jespersen
Journal:  J Clin Pathol       Date:  2003-02       Impact factor: 3.411