Literature DB >> 9609180

Interactive MR imaging-guided biopsy and aspiration with a modified clinical C-arm system.

J S Lewin1, C A Petersilge, S F Hatem, J L Duerk, G Lenz, M E Clampitt, M L Williams, K R Kaczynski, C F Lanzieri, A L Wise, J R Haaga.   

Abstract

OBJECTIVE: This investigation was performed to test the hypotheses that interactive guidance of MR image acquisition during needle-directed procedures using a clinical 0.2-T C-arm open MR imaging system integrated with a frameless optically linked stereotaxy system is feasible, and that procedure times can be sufficiently short to be well tolerated by the patient. SUBJECTS AND METHODS: One hundred six MR-guided procedures were performed in 86 patients (ranging in age from 5 months to 88 years) using a clinical C-arm imaging system supplemented with an in-room RF-shielded liquid crystal display monitor, a frameless stereotaxy system, rapid gradient-echo sequences for needle guidance, and MR-compatible monitoring and surgical lighting equipment. We performed 50 biopsies and aspirations of the head and neck in 37 patients, 23 biopsies of musculoskeletal lesions in 22 patients, 16 biopsies of abdominal sites in 10 patients, six biopsies of the thoracolumbar spine or sacrum in six patients, and 11 shoulder joint injections for MR arthrography in 11 patients, in addition to 38 MR arthrographic injections on the same imaging system described in a previous report. Tissue sampling included fine-needle aspiration (n = 90) and cutting needle core biopsy (n = 41). Thirty-five patients underwent both procedures. Procedures were evaluated for success of needle placement, procedure time, and complications.
RESULTS: Needle placement was successful in all cases, and no complications occurred. Tissue was sufficient for pathologic diagnosis for all but eight patients. Passes per patient averaged 2.1. For fine-needle aspiration, instrument time averaged 7 min 42 sec per pass, cutting needle core biopsy averaged 6 min 24 sec, and shoulder injection averaged 8 min.
CONCLUSION: MR imaging guidance for needle procedures on a clinical 0.2-T C-arm system with supplemental interventional accessories is feasible, with relatively rapid needle placement.

Entities:  

Mesh:

Year:  1998        PMID: 9609180     DOI: 10.2214/ajr.170.6.9609180

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  18 in total

Review 1.  Image-guided percutaneous spinal biopsy.

Authors:  Gajan Rajeswaran; Qaiser Malik; Asif Saifuddin
Journal:  Skeletal Radiol       Date:  2012-06-04       Impact factor: 2.199

2.  MRI-guided abdominal biopsy in a 0.23-T open-configuration MRI system.

Authors:  J Kariniemi; R Blanco Sequeiros; R Ojala; O Tervonen
Journal:  Eur Radiol       Date:  2004-12-31       Impact factor: 5.315

Review 3.  Interventional and intraoperative MR: review and update of techniques and clinical experience.

Authors:  Thomas Schulz; Silvia Puccini; Jens-Peter Schneider; Thomas Kahn
Journal:  Eur Radiol       Date:  2004-10-06       Impact factor: 5.315

Review 4.  Image-guided tumor ablation: standardization of terminology and reporting criteria.

Authors:  S Nahum Goldberg; Clement J Grassi; John F Cardella; J William Charboneau; Gerald D Dodd; Damian E Dupuy; Debra Gervais; Alice R Gillams; Robert A Kane; Fred T Lee; Tito Livraghi; John McGahan; David A Phillips; Hyunchul Rhim; Stuart G Silverman
Journal:  Radiology       Date:  2005-04-21       Impact factor: 11.105

5.  CT-guided aspirations in the head and neck: assessment of the first 216 cases.

Authors:  Paul M Sherman; David M Yousem; Laurie A Loevner
Journal:  AJNR Am J Neuroradiol       Date:  2004-10       Impact factor: 3.825

6.  Real-time MR fluoroscopy-navigated lumbar facet joint injections: feasibility and technical properties.

Authors:  J Fritz; S Clasen; A Boss; C Thomas; C W König; C D Claussen; P L Pereira
Journal:  Eur Radiol       Date:  2008-02-26       Impact factor: 5.315

7.  Device localization and dynamic scan plane selection using a wireless magnetic resonance imaging detector array.

Authors:  Matthew J Riffe; Stephen R Yutzy; Yun Jiang; Michael D Twieg; Colin J Blumenthal; Daniel P Hsu; Li Pan; Wesley D Gilson; Jeffrey L Sunshine; Christopher A Flask; Jeffrey L Duerk; Dean Nakamoto; Vikas Gulani; Mark A Griswold
Journal:  Magn Reson Med       Date:  2013-07-30       Impact factor: 4.668

8.  Evaluation of MR imaging guided steroid injection of the sacroiliac joints for the treatment of children with refractory enthesitis-related arthritis.

Authors:  J Fritz; N Tzaribachev; C Thomas; J A Carrino; C D Claussen; J S Lewin; P L Pereira
Journal:  Eur Radiol       Date:  2010-10-29       Impact factor: 5.315

9.  Feasibility of MR-temperature mapping of ultrasonic heating from a CMUT.

Authors:  Serena H Wong; Ronald D Watkins; Mario Kupnik; Kim Butts Pauly; Butrus T Khuri-Yakub
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2008-04       Impact factor: 2.725

10.  A technique for MRI-guided transrectal deep pelvic abscess drainage.

Authors:  Sherif Gamal Nour; Jamal J Derakhshan; Nila J Akhtar; Martin A Ayres; Mark E Clampitt; Thomas A Stellato; Jeffrey L Duerk
Journal:  AJR Am J Roentgenol       Date:  2008-10       Impact factor: 3.959

View more