Literature DB >> 9482025

Characteristics of the Orthopantomograph OP 100.

W C Scarfe1, F E Eraso, A G Farman.   

Abstract

OBJECTIVE: To investigate empirically the imaging characteristics of the Orthopantomograph OP 100 and compare them to the average form of the dental arch.
METHODS: Repeatable radiographs were taken of a lead resolution grid positioned at 1 mm increments along known angular intervals of the projected X-ray beam. Focal trough thickness was determined by visibility of the 1.5 lp mm-1 resolution limits. The path of the effective rotation center was determined using a film positioned at right angles to the slit beam. The vertical magnification and horizontal magnification and distortion index, corrected for the position of the tomographic layer, were calculated using a reference object placed at various resolution limits of the focal trough. The beam projection angle was calculated with respect to the central plane of the focal trough and the average dental arch shape compared with the average proximal contact angle.
RESULTS: The maximum resolution observed at the central plane of the focal trough was 5 lp mm-1. The width of the focal trough varied from 17 mm in the anterior region to 44 mm in the posterior region. The path of the effective rotation center was found to translate between three fixed centers of rotation, with the effective anterior projection radius of 24 mm. The vertical magnification factor within the focal trough showed a linear increase along the beam path from 1.24-1.37 and the horizontal magnification varied from 1.01-1.63. The distortion index varied from 0.84-1.24. The beam projection angle increased from 90 degrees anteriorly to 115 degrees in the premolar area with respect to the central plane of the focal trough and from 90 degrees anteriorly to 110 degrees in the premolar area compared with the average dental arch.
CONCLUSIONS: The OP 100 provides a focal trough conforming well to the overall geometry of the dental arch and provides adequate spatial resolution.

Mesh:

Year:  1998        PMID: 9482025     DOI: 10.1038/sj.dmfr.4600310

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  9 in total

1.  Influence of the intergonial distance on image distortion in panoramic radiographs.

Authors:  D B S Ladeira; A D Cruz; S M Almeida; F N Bóscolo
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

2.  Strut analysis for osteoporosis detection model using dental panoramic radiography.

Authors:  Jae Joon Hwang; Jeong-Hee Lee; Sang-Sun Han; Young Hyun Kim; Ho-Gul Jeong; Yoon Jeong Choi; Wonse Park
Journal:  Dentomaxillofac Radiol       Date:  2017-07-14       Impact factor: 2.419

3.  Object position and image magnification in dental panoramic radiography: a theoretical analysis.

Authors:  H Devlin; J Yuan
Journal:  Dentomaxillofac Radiol       Date:  2012-08-29       Impact factor: 2.419

4.  Accuracy of vertical height measurements on direct digital panoramic radiographs using posterior mandibular implants and metal balls as reference objects.

Authors:  L Vazquez; Y Nizamaldin; C Combescure; R Nedir; M Bischof; D M Dohan Ehrenfest; J-P Carrel; U C Belser
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

5.  Lateral lingual vascular canals of the mandible: a CBCT study of 500 cases.

Authors:  Halil Sahman; Ahmet Ercan Sekerci; Elif Tarim Ertas
Journal:  Surg Radiol Anat       Date:  2014-01-29       Impact factor: 1.246

6.  A new bite block for panoramic radiographs of anterior edentulous patients: A technical report.

Authors:  Jong-Woong Park; Khanthaly Symkhampha; Kyung-Hoe Huh; Won-Jin Yi; Min-Suk Heo; Sam-Sun Lee; Soon-Chul Choi
Journal:  Imaging Sci Dent       Date:  2015-06-19

Review 7.  Advanced dental implant placement techniques.

Authors:  Alex M Greenberg
Journal:  J Istanb Univ Fac Dent       Date:  2017-12-02

8.  Comparison of the reproducibility of panoramic radiographs between dentulous and edentulous patients.

Authors:  Jong-Woong Park; Kyung-Hoe Huh; Won-Jin Yi; Min-Suk Heo; Sam-Sun Lee; Soon-Chul Choi
Journal:  Imaging Sci Dent       Date:  2014-06-11

9.  Automatic analysis algorithm for acquiring standard dental and mandibular shape data using cone-beam computed tomography.

Authors:  Jae Joon Hwang; Sang-Sun Han; Chena Lee; Yun-Hoa Jung
Journal:  Sci Rep       Date:  2018-09-10       Impact factor: 4.379

  9 in total