Use of the selective linogram in cardiac tomography quality control.

J W Wallis1.   

Abstract

BACKGROUND: Quality control for detection of patient motion is essential in tomographic myocardial imaging. Despite significant limitations, the summation image or conventional "linogram" has long been advocated as a useful image in the detection of vertical motion. In this study a new quality control image entitled the "selective linogram" is proposed to replace the summation image in routine cardiac single-photon emission cardiac tomography (SPECT) quality control. The selective linogram is constructed in a manner somewhat analogous to the sinogram. In the sinogram, each row represents a different projection angle; in the selective linogram each column represents a different projection angle. METHODS AND
RESULTS: After selection of eight motion-free studies from acquisitions at our clinical center, vertical motion of various types (bounces, shifts, and creep) were added to the projection frames. Summation image and selective linogram quality control images from these motion-containing studies and the original motion-free studies were presented in a blinded manner to two observers for scoring of patient motion. The selective linogram was significantly more accurate in allowing detection of vertical motion than was the summation image (accuracy 89% vs 47%).
CONCLUSIONS: The selective linogram image is markedly superior to the summation image for the detection of vertical patient motion during cardiac SPECT. This new technique can be a valuable aid in SPECT quality control.

Entities:  

Mesh:

Year:  1995        PMID: 9420804     DOI: 10.1016/s1071-3581(05)80074-1

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  12 in total

1.  Orbit-related variation in spatial resolution as a source of artifactual defects in thallium-201 SPECT.

Authors:  P J Maniawski; H T Morgan; F J Wackers
Journal:  J Nucl Med       Date:  1991-05       Impact factor: 10.057

2.  Optimal specificity of thallium-201 SPECT through recognition of imaging artifacts.

Authors:  E G DePuey; E V Garcia
Journal:  J Nucl Med       Date:  1989-04       Impact factor: 10.057

3.  Sensitivity of SPECT thallium-201 myocardial perfusion imaging to patient motion.

Authors:  R L Eisner
Journal:  J Nucl Med       Date:  1992-08       Impact factor: 10.057

4.  Use of cross-correlation function to detect patient motion during SPECT imaging.

Authors:  R L Eisner; T Noever; D Nowak; W Carlson; D Dunn; J Oates; K Cloninger; H A Liberman; R E Patterson
Journal:  J Nucl Med       Date:  1987-01       Impact factor: 10.057

5.  Effect of motion on thallium-201 SPECT studies: a simulation and clinical study.

Authors:  F M Prigent; M Hyun; D S Berman; A Rozanski
Journal:  J Nucl Med       Date:  1993-11       Impact factor: 10.057

6.  How to detect and avoid myocardial perfusion SPECT artifacts.

Authors:  E G DePuey
Journal:  J Nucl Med       Date:  1994-04       Impact factor: 10.057

7.  Effect of patient motion on tomographic myocardial perfusion imaging.

Authors:  J A Cooper; P H Neumann; B K McCandless
Journal:  J Nucl Med       Date:  1992-08       Impact factor: 10.057

8.  Detection and correction of patient motion in dynamic and static myocardial SPECT using a multi-detector camera.

Authors:  G Germano; T Chua; P B Kavanagh; H Kiat; D S Berman
Journal:  J Nucl Med       Date:  1993-08       Impact factor: 10.057

9.  Detection of patient motion during tomographic myocardial perfusion imaging.

Authors:  J A Cooper; P H Neumann; B K McCandless
Journal:  J Nucl Med       Date:  1993-08       Impact factor: 10.057

10.  Quantitative analysis of the tomographic thallium-201 myocardial bullseye display: critical role of correcting for patient motion.

Authors:  R Eisner; A Churchwell; T Noever; D Nowak; K Cloninger; D Dunn; W Carlson; J Oates; J Jones; D Morris
Journal:  J Nucl Med       Date:  1988-01       Impact factor: 10.057

View more
  1 in total

1.  Quantitative evaluation of a comprehensive motion, resolution, and attenuation correction program: initial experience.

Authors:  P Rigo; P Van Boxem; J Foulon; M Safi; J Engdahl; J Links
Journal:  J Nucl Cardiol       Date:  1998 Sep-Oct       Impact factor: 5.952

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.