Hemolysis in vivo from exposure to pulsed ultrasound.

D Dalecki1, C H Raeman, S Z Child, C Cox, C W Francis, R S Meltzer, E L Carstensen.   

Abstract

Ultrasonically induced hemolysis in vivo when a commercial ultrasound contrast agent, Albunex, was present in the blood. Murine hearts were exposed for 5 min at either 1.15 or 2.35 MHz with a pulse length of 10 microseconds and pulse repetition frequency of 100 Hz. During the exposure period, four boluses of Albunex were injected into a tail vein for a total of approximately 0.1 mL of Albunex. Following exposure, blood was collected by heart puncture and centrifuged, and the plasma was analyzed for hemoglobin concentration. With Albunex present in the blood, the threshold for hemolysis at 1.15 MHz was 3.0 +/- 0.8 MPa (mean +/- SD) peak positive pressure (approximately 1.9 MPa negative pressure, approximately 180 W cm-2 pulse average intensity). For the highest exposure levels (10 MPa peak positive pressure at the surface of the animal), the mean value for hemolysis was approximately 4% at 1.15 MHz and 0.46% at 2.35 MHz, i.e., the threshold at 2.35 MHz is > 10 MPa peak positive pressure. In contrast, hemolysis in control mice receiving saline injections at 10 MPa or sham-exposed (0 MPa) mice receiving Albunex was approximately 0.4%.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9140187     DOI: 10.1016/s0301-5629(96)00203-7

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  15 in total

Review 1.  Section 8--clinical relevance. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

Review 2.  Section 6--mechanical bioeffects in the presence of gas-carrier ultrasound contrast agents. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

Review 3.  Section 7--discussion of the mechanical index and other exposure parameters. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

Review 4.  Section 4--bioeffects in tissues with gas bodies. American Institute of Ultrasound in Medicine.

Authors: 
Journal:  J Ultrasound Med       Date:  2000-02       Impact factor: 2.153

5.  Diagnostic ultrasound activation of contrast agent gas bodies induces capillary rupture in mice.

Authors:  D L Miller; J Quddus
Journal:  Proc Natl Acad Sci U S A       Date:  2000-08-29       Impact factor: 11.205

Review 6.  Ultrasound-biophysics mechanisms.

Authors:  William D O'Brien
Journal:  Prog Biophys Mol Biol       Date:  2006-08-08       Impact factor: 3.667

7.  A new strategy to enhance cavitational tissue erosion using a high-intensity, Initiating sequence.

Authors:  Zhen Xu; J Brian Fowlkes; Charles A Cain
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2006-08       Impact factor: 2.725

8.  Evaluation of the threshold for lung hemorrhage by diagnostic ultrasound and a proposed new safety index.

Authors:  Charles C Church; William D O'Brien
Journal:  Ultrasound Med Biol       Date:  2007-03-26       Impact factor: 2.998

9.  Sonothrombolysis: an emerging modality for the treatment of acute ischemic and hemorrhagic stroke.

Authors:  Azita Soltani; Wayne M Clark; Douglas R Hansmann
Journal:  Transl Stroke Res       Date:  2011-04-15       Impact factor: 6.829

Review 10.  [Ultrasound contrast agents. Pharmaceutical drug safety and bioeffects].

Authors:  M Krix; J W Jenne
Journal:  Radiologe       Date:  2007-09       Impact factor: 0.635

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.