Posterior lumbar interbody fusion. A biomechanical comparison, including a new threaded cage.

D S Brodke1, J C Dick, D N Kunz, R McCabe, T A Zdeblick.   

Abstract

STUDY
DESIGN: In vitro biomechanical testing was performed on eight lumbar calf spines.
OBJECTIVES: To compare the initial stiffness of a standard method of posterior lumbar interbody fusion using structural autograft with the same procedure using additional posterior instrumentation. These constructs also were compared to a new titanium implant. SUMMARY OF BACKGROUND DATA: Posterior lumbar interbody fusion is gaining wide acceptance for the treatment of segmental spinal instability, spondylolisthesis, and discogenic pain. Many methods have been described, including use of autograft or allograft bone, in either structural or nonstructural form, with or without additional fixation. A new threaded titanium interbody implant has been designed to increase initial stability while allowing bony ingrowth for fusion.
METHODS: Eight lumbar calf spines were subjected to axial compression, sagittal moments (flexion-extension), and axial torque while displacement was measured. Stiffness was calculated from the load displacement curves for each construct under each load pattern.
RESULTS: The posterior lumbar interbody fusion by bone graft alone was the least stiff construct of all modes tested. In two of eight specimens the bone graft dislodged posteriorly into the canal during torsional testing. The titanium interbody implant was similar in stiffness to the bone graft posterior lumbar interbody fusion with posterior instrumentation group in all three modes. They were both significantly stiffer than the normal spine, the destabilized spine, and the posterior lumbar interbody fusion by bone graft alone (P < 0.05).
CONCLUSIONS: In this model, the posterior lumbar interbody fusion with bone graft alone had less initial stiffness than that of the intact spine. The addition of posterior instrumentation or interbody implants can increase initial stiffness significantly.

Entities:  

Mesh:

Year:  1997        PMID: 9122778     DOI: 10.1097/00007632-199701010-00005

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  39 in total

Review 1.  [Vertebral body replacement in spine surgery].

Authors:  F Kandziora; K J Schnake; C K Klostermann; N P Haas
Journal:  Unfallchirurg       Date:  2004-05       Impact factor: 1.000

2.  [Biodegradable cage. Osteointegration in spondylodesis of the sheep cervical spine].

Authors:  R Pflugmacher; T Eindorf; M Scholz; S Gumnior; C Krall; P Schleicher; N P Haas; F Kandziora
Journal:  Chirurg       Date:  2004-10       Impact factor: 0.955

3.  A history of spine biomechanics. Focus on 20th century progress.

Authors:  T R Oxland
Journal:  Unfallchirurg       Date:  2015-12       Impact factor: 1.000

4.  Minimally-invasive technique for transforaminal lumbar interbody fusion (TLIF).

Authors:  Burak M Ozgur; Kevin Yoo; Gerardo Rodriguez; William R Taylor
Journal:  Eur Spine J       Date:  2005-09-08       Impact factor: 3.134

5.  Primary stability of anterior lumbar stabilization: interdependence of implant type and endplate retention or removal.

Authors:  Christian H Flamme; Nadine von der Heide; Caroline Heymann; Christof Hurschler
Journal:  Eur Spine J       Date:  2005-08-10       Impact factor: 3.134

6.  Comparison of two interbody fusion cages for posterior lumbar interbody fusion in a cadaveric model.

Authors:  Shih-Tien Wang; Vijay K Goel; Chong-Yau Fu; Shinichiro Kubo; Woosung Choi; Chien-Lin Liu; Tain-Hsiung Chen
Journal:  Int Orthop       Date:  2006-02-28       Impact factor: 3.075

7.  Porcine models in spinal research: calibration and comparative finite element analysis of various configurations during flexion-extension.

Authors:  Hadi N Aziz; Fabio Galbusera; Chiara Maria Bellini; Giuseppe Vincenzo Mineo; Alessandro Addis; Riccardo Pietrabissa; Marco Brayda-Bruno
Journal:  Comp Med       Date:  2008-04       Impact factor: 0.982

8.  Current concepts on spinal arthrodesis in degenerative disorders of the lumbar spine.

Authors:  Marios G Lykissas; Alexander Aichmair
Journal:  World J Clin Cases       Date:  2013-04-16       Impact factor: 1.337

9.  Assessment of BioPlex interbody fusion device in a sheep lumbar fusion model.

Authors:  Douglas C Fredericks; Anup A Gandhi; Nicole M Grosland; Joseph D Smucker
Journal:  Iowa Orthop J       Date:  2013

10.  Repeated migration of a fusion cage after posterior lumbar interbody fusion.

Authors:  Jun Gue Lee; Sung Myung Lee; Seok Won Kim; Ho Shin
Journal:  Korean J Spine       Date:  2013-03-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.