Cervical sagittal range-of-motion analysis using three methods. Cervical range-of-motion device, 3space, and radiography.

N R Ordway1, R Seymour, R G Donelson, L Hojnowski, E Lee, W T Edwards.   

Abstract

STUDY
DESIGN: Cervical flexion, extension, protrusion, and retraction were comparatively measured in volunteers using a cervical range-of-motion device (Performance Attainment Associates, Roseville, MN), a 3space system (Polhemus, Colchester, VT), and lateral radiographs.
OBJECTIVES: To compare the outcomes of three methods of measurement of cervical flexion, extension, protrusion, and retraction. SUMMARY OF BACKGROUND DATA: Few studies compare cervical range-of-motion measurement devices with radiographic measurements, and no studies have compared methods of measurement for cervical protrusion and retraction measurement.
METHODS: In 20 asymptomatic volunteers, four end-range sagittal cervical positions (flexion, extension, protrusion, and retraction) were measured simultaneously using a cervical range-of-motion device, a 3Space and lateral cervical radiographs. Measurements were compared, and differences were analyzed.
RESULTS: There were no significant differences for flexion and extension measurements between the cervical range-of-motion device and that radiographic angle determined by an occipital line and the vertical, nor were there any between the 3Space and that radiographic angle between this same occipital line and C7. The cervical range-of-motion device and the 3Space measurements for flexion and extension, however, differed significantly from one another (P < 0.05). For protrusion and retraction, there was no significant difference between the 3Space and radiographic measurements, but these two both differed significantly from the cervical range-of-motion device (P < 0.05).
CONCLUSIONS: Available methods of measurement differ as to whether the cervical spine is isolated or includes upper thoracic motion. Protrusion and retraction can be measured reliably with all three methods studied, but without measurement consistency between devices. Because end-range cervical flexion and extension-cannot occur in isolation from upper thoracic motion, true cervical motion can be measured only with an internally referenced, or landmark-based, methodology such as the 3Space. Even though the cervical range-of-motion device cannot measure isolated cervical flexion and extension, it is nevertheless a reliable clinical tool in measuring flexion and extension as well as protrusion and retraction as long as patient thoracic positioning is standardized to minimize the upper thoracic contribution.

Entities:  

Mesh:

Year:  1997        PMID: 9076881     DOI: 10.1097/00007632-199703010-00007

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  15 in total

1.  Collegiate Football Players Display More Active Cervical Spine Mobility Than High School Football Players.

Authors:  John Nyland; Darren Johnson
Journal:  J Athl Train       Date:  2004-06       Impact factor: 2.860

2.  Evidence-based protocol for structural rehabilitation of the spine and posture: review of clinical biomechanics of posture (CBP) publications.

Authors:  Paul A Oakley; Donald D Harrison; Deed E Harrison; Jason W Haas
Journal:  J Can Chiropr Assoc       Date:  2005-12

3.  Preliminary study: reliability of the spinal wheel. A novel device to measure spinal postures applied to sitting and standing.

Authors:  Liba Sheeran; Valerie Sparkes; Monica Busse; Robert van Deursen
Journal:  Eur Spine J       Date:  2009-12-15       Impact factor: 3.134

4.  Neck range of motion measurements using a new three-dimensional motion analysis system: validity and repeatability.

Authors:  Haruhi Inokuchi; Michio Tojima; Hiroshi Mano; Yuki Ishikawa; Naoshi Ogata; Nobuhiko Haga
Journal:  Eur Spine J       Date:  2015-04-07       Impact factor: 3.134

5.  A 3D motion analysis study comparing the effectiveness of cervical spine orthoses at restricting spinal motion through physiological ranges.

Authors:  Nicholas Rhys Evans; Georgina Hooper; Rachel Edwards; Gemma Whatling; Valerie Sparkes; Cathy Holt; Sashin Ahuja
Journal:  Eur Spine J       Date:  2013-01-04       Impact factor: 3.134

6.  On the relevance of surrogate parameter deduction in biomedical research: mediated regression analysis for variance explanation of cervical range of motion.

Authors:  Daniel Niederer; Lutz Vogt; Jan Wilke; Winfried Banzer
Journal:  Eur Spine J       Date:  2016-06-21       Impact factor: 3.134

7.  Age-related cutoffs for cervical movement behaviour to distinguish chronic idiopathic neck pain patients from unimpaired subjects.

Authors:  Daniel Niederer; Lutz Vogt; Jan Wilke; Marcus Rickert; Winfried Banzer
Journal:  Eur Spine J       Date:  2014-12-10       Impact factor: 3.134

8.  Three-dimensional analysis of cervical spine motion: reliability of a computer assisted magnetic tracking device compared to inclinometer.

Authors:  Ioannis D Gelalis; Louis E DeFrate; Kosmas S Stafilas; Emilios E Pakos; James D Kang; Lars G Gilbertson
Journal:  Eur Spine J       Date:  2008-12-19       Impact factor: 3.134

9.  Intra-instrument reliability of 4 goniometers.

Authors:  R Kevin Pringle
Journal:  J Chiropr Med       Date:  2003

10.  Immediate effects of inhibitive distraction on active range of cervical flexion in patients with neck pain: a pilot study.

Authors:  Kristín Briem; Peter Huijbregts; Maria Thorsteinsdottir
Journal:  J Man Manip Ther       Date:  2007
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.