Literature DB >> 8950588

New device for artificial increasing intraocular pressure.

P R Preussner1, A Duran.   

Abstract

BACKGROUND: For certain special investigations, particularly in glaucoma diagnostics, intraocular pressure has to be increased. Mostly, a suction cup apparatus is used for that purpose. These devices, however, have some disadvantages. They are at least uncomfortable, sometimes risky for the patient and induce a strong astigmatism which can influence the results of the corresponding investigation.
METHOD: A very simple mechanical device has been constructed and tested, by means of which the intraocular pressure can be increased for diagnostic purposes. A ring on a lever, which is moved towards the eye by small weights, compresses the eye through the lids without the need of even local anesthetics. The device can be easily mounted to any head fixation which is normally used in combination with ophthalmological equipment. e.g. slit lamp, visual field analyzer or fundus camera. Tests have been performed in four different groups: one group of 19 healthy volunteers, one of 21 patients with Graves' disease, one of 21 patients with high myopia and one of 20 patients with open-angle glaucoma.
RESULTS: Pressure elevations could be achieved in all patients with very good reproducibility with respect to the applied forces. However, in patients with exophthalmos (Graves' disease), the elevation effect was somewhat higher than in normal eyes. The induced astigmatism was only about 0.5 dpt for pressure elevations of about 20 mmHg.
CONCLUSION: The tests performed showed that the new device is easy to handle and comfortable for the patients. If very high accuracy is needed concerning the pressure values achieved, an individual calibration should be performed.

Entities:  

Mesh:

Year:  1996        PMID: 8950588     DOI: 10.1007/bf00292354

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  14 in total

1.  Artificially increasing intraocular pressure causes flicker sensitivity losses.

Authors:  V Van Toi; P A Grounauer; C W Burckhardt
Journal:  Invest Ophthalmol Vis Sci       Date:  1990-08       Impact factor: 4.799

2.  [Relation of negative pressure difference and artificially elevated intraocular pressure using the suction cup method].

Authors:  R Stodtmeister; L Pillunat; A Mattern; E Polly
Journal:  Klin Monbl Augenheilkd       Date:  1989-03       Impact factor: 0.700

3.  Visual evoked cortical potentials during pressure-blinding.

Authors:  R Fox; R Blake; J R Bourne
Journal:  Vision Res       Date:  1973-02       Impact factor: 1.886

4.  [A test of pressure tolerance of the optic nerve head in ocular hypertension].

Authors:  L E Pillunat; R Stodtmeister; I Wilmanns; T Christ
Journal:  Klin Monbl Augenheilkd       Date:  1986-01       Impact factor: 0.700

5.  Reliability of intraocular pressure measurement with the Goldmann applanation tonometer in epidemiological studies.

Authors:  I Dielemans; J R Vingerling; A Hofman; D E Grobbee; P T de Jong
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1994-03       Impact factor: 3.117

6.  Refraction changes during elevation of intraocular pressure by suction cup, their reflection in the pattern visual evoked cortical potential and their compensation.

Authors:  A Bernd; W D Ulrich; H Teubel; F Rohrwacher; T Barth
Journal:  Doc Ophthalmol       Date:  1993       Impact factor: 2.379

7.  [Effect of change in refraction by increasing intraocular pressure with the suction cup method on pattern reversal VECP].

Authors:  F Rohrwacher; A Bernd; C Ulrich; H Teubel; T Barth; W D Ulrich
Journal:  Ophthalmologe       Date:  1994-04       Impact factor: 1.059

8.  [Biomicroscopic pneumatic ophthalmodynamography (author's transl)].

Authors:  P A Grounauer; V van Toi; E B Streiff
Journal:  Klin Monbl Augenheilkd       Date:  1980-04       Impact factor: 0.700

9.  [Modification of refraction and visual noise perception by suction cup oculocompression].

Authors:  U Schiefer; H Wilhelm; E Zrenner; E Aulhorn
Journal:  Fortschr Ophthalmol       Date:  1991

10.  [White-noise field campimetry before and after artificial increase of intraocular pressure. Possible applications in diagnosis and evaluation of glaucoma].

Authors:  U Schiefer; W D Ulrich; C Ulrich; H Wilhelm; E Aulhorn
Journal:  Ophthalmologe       Date:  1992-12       Impact factor: 1.059

View more
  2 in total

1.  Two-year efficacy after first transscleral controlled cyclophotocoagulation in patients with and without pseudoexfoliation.

Authors:  Markus Lenzhofer; Melchior Hohensinn; Wolfgang Hitzl; Veit Steiner; Armin Motaabbed; Karolina Motloch; Hans Peter Colvin; Herbert A Reitsamer; Sarah Moussa
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-04-02       Impact factor: 3.117

2.  MicroPulse Transscleral Laser Therapy Demonstrates Similar Efficacy with a Superior and More Favorable Safety Profile Compared to Continuous-Wave Transscleral Cyclophotocoagulation.

Authors:  Enrico Bernardi; Marc Töteberg-Harms
Journal:  J Ophthalmol       Date:  2022-02-08       Impact factor: 1.909

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.