Literature DB >> 8946293

Optimum parameters for short-wavelength automated perimetry.

P A Sample1, C A Johnson, G Haegerstrom-Portnoy, A J Adams.   

Abstract

PURPOSE: To determine the optimum parameters for short-wavelength automated perimetry (SWAP) and to recommend these for standardization of the procedure.
METHODS: We used a variety of stimulus and background configurations to determine the optimum background spectral distribution and luminance, and the optimum target spectral distribution, maximum luminance, and duration. We measured threshold versus intensity curves to determine which combination provided (a) the greatest isolation of the short-wavelength sensitive mechanisms and (b) the largest dynamic range for perimetry. We also evaluated the effect of lens absorption and cataract on these two factors.
RESULTS: A broad-band yellow background at 100 candela/m2 with a narrow-band 440-nm (27-nm half-bandwidth), 1.8 degrees diameter (Goldmann size V) stimulus presented for 200 ms was optimum at all retinal eccentricities. Specific recommendations for how to modify existing perimeters are given.
CONCLUSION: Agreement regarding the optimum parameters for SWAP should lead to standardization of the test that will facilitate comparison of results from different centers. Normative data can be collected at several sites and incorporated into statistical analysis packages currently available with various perimeters. This will greatly improve the clinical utility of this test.

Entities:  

Mesh:

Year:  1996        PMID: 8946293

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  20 in total

Review 1.  Acquired colour vision defects in glaucoma-their detection and clinical significance.

Authors:  M Pacheco-Cutillas; D F Edgar; A Sahraie
Journal:  Br J Ophthalmol       Date:  1999-12       Impact factor: 4.638

2.  African Descent and Glaucoma Evaluation Study (ADAGES): III. Ancestry differences in visual function in healthy eyes.

Authors:  Lyne Racette; Jeffrey M Liebmann; Christopher A Girkin; Linda M Zangwill; Sonia Jain; Lida M Becerra; Felipe A Medeiros; Christopher Bowd; Robert N Weinreb; Catherine Boden; Pamela A Sample
Journal:  Arch Ophthalmol       Date:  2010-05

3.  A comparison of perimetric results with the Medmont and Humphrey perimeters.

Authors:  J Landers; A Sharma; I Goldberg; S Graham
Journal:  Br J Ophthalmol       Date:  2003-06       Impact factor: 4.638

4.  Pattern electroretinogram and psychophysical tests of visual function for discriminating between healthy and glaucoma eyes.

Authors:  Ali Tafreshi; Lyne Racette; Robert N Weinreb; Pamela A Sample; Linda M Zangwill; Felipe A Medeiros; Christopher Bowd
Journal:  Am J Ophthalmol       Date:  2010-03       Impact factor: 5.258

Review 5.  A multifocal electroretinogram model predicting the development of diabetic retinopathy.

Authors:  Marcus A Bearse; Anthony J Adams; Ying Han; Marilyn E Schneck; Jason Ng; Kevin Bronson-Castain; Shirin Barez
Journal:  Prog Retin Eye Res       Date:  2006-09-01       Impact factor: 21.198

6.  Topography of the frequency doubling perimetry visual field compared with that of short wavelength and achromatic automated perimetry visual fields.

Authors:  J Landers; A Sharma; I Goldberg; S Graham
Journal:  Br J Ophthalmol       Date:  2006-01       Impact factor: 4.638

7.  Correlation of a scanning laser derived oedema index and visual function following grid laser treatment for diabetic macular oedema.

Authors:  C Hudson; J G Flanagan; G S Turner; H C Chen; L B Young; D McLeod
Journal:  Br J Ophthalmol       Date:  2003-04       Impact factor: 4.638

8.  Comparing the full-threshold and Swedish interactive thresholding algorithms for short-wavelength automated perimetry.

Authors:  Minna Ng; Lyne Racette; John P Pascual; Jeffrey M Liebmann; Christopher A Girkin; Sarah L Lovell; Linda M Zangwill; Robert N Weinreb; Pamela A Sample
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-12-13       Impact factor: 4.799

9.  Variability in short-wavelength automated perimetry among peri- or postmenopausal women: a dependence on phyto-oestrogen consumption?

Authors:  Alvin Eisner; Shaban Demirel
Journal:  Acta Ophthalmol       Date:  2011-05       Impact factor: 3.761

10.  Is there evidence for continued learning over multiple years in perimetry?

Authors:  Stuart K Gardiner; Shaban Demirel; Chris A Johnson
Journal:  Optom Vis Sci       Date:  2008-11       Impact factor: 1.973

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.