Literature DB >> 8923072

Bedside burr hole for intracranial pressure monitoring performed by intensive care physicians. A 5-year experience.

M Bochicchio1, N Latronico, S Zappa, A Beindorf, A Candiani.   

Abstract

OBJECTIVE: To assess the results of a 5-year experience with bedside burr hole for intracranial pressure (ICP) monitoring performed by intensive care physicians.
DESIGN: Prospective, observational study in 120 patients.
SETTING: A general-neurologic Intensive Care Unit in a University Hospital. PATIENTS: Patients admitted for acute neural lesion requiring ICP monitoring.
METHOD: A 2.71 mm burr hole was made with positioning of a subarachnoid screw, through which a miniaturized fiberoptic, tip transducer device (Camino) was advanced and inserted 2 mm in the frontal cortex. MAIN
RESULTS: Over a 5-year period 120 patients, mainly with severe head trauma, underwent ICP monitoring. None of the planned patients was excluded because of technical difficulties. No life-threatening complications were reported, and the overall morbidity rate related to the ICP monitor was 3.3%. Complications were infectious in nature, with 2.5% wound infections and 0.8% meningitis. Although seven patients bled when opening the dura, no intracranial hematomas were recorded due to the ICP monitor. The fiberoptic device was left in place for 5 +/- 1.6 (SD) days (range 1-12 days). Five patients (4.1%) required catheter substitution due to breakage of the system components (fiberoptics).
CONCLUSIONS: Bedside insertion of a ICP monitor performed by intensive care physicians is a safe procedure, with a complication rate comparable to other series published by neurosurgeons. The overall morbidity rate is comparable to, or even lower than, that caused by central vein catheterization.

Entities:  

Mesh:

Year:  1996        PMID: 8923072     DOI: 10.1007/bf01699230

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  21 in total

1.  Evaluation of a fiberoptic intracranial pressure monitor.

Authors:  J S Crutchfield; R K Narayan; C S Robertson; L H Michael
Journal:  J Neurosurg       Date:  1990-03       Impact factor: 5.115

2.  Reye's syndrome: assessment of intracranial monitoring.

Authors:  J G Jenkins; J F Glasgow; G W Black; T F Fannin; E M Hicks; S R Keilty; P M Crean
Journal:  Br Med J (Clin Res Ed)       Date:  1987-02-07

3.  The outcome with aggressive treatment in severe head injuries. Part I: the significance of intracranial pressure monitoring.

Authors:  L F Marshall; R W Smith; H M Shapiro
Journal:  J Neurosurg       Date:  1979-01       Impact factor: 5.115

4.  Intracranial pressure: to monitor or not to monitor? A review of our experience with severe head injury.

Authors:  R K Narayan; P R Kishore; D P Becker; J D Ward; G G Enas; R P Greenberg; A Domingues Da Silva; M H Lipper; S C Choi; C G Mayhall; H A Lutz; H F Young
Journal:  J Neurosurg       Date:  1982-05       Impact factor: 5.115

5.  Intracranial pressure and outcome in adult encephalitis.

Authors:  G H Barnett; A H Ropper; J Romeo
Journal:  J Neurosurg       Date:  1988-04       Impact factor: 5.115

6.  ICP monitoring: complications and associated factors.

Authors:  M J Rosner; D P Becker
Journal:  Clin Neurosurg       Date:  1976

7.  The safety and value of extradural intracranial pressure monitors in fulminant hepatic failure.

Authors:  R T Keays; G J Alexander; R Williams
Journal:  J Hepatol       Date:  1993-06       Impact factor: 25.083

8.  Clinical experience with a fiberoptic intracranial pressure monitor.

Authors:  J S Yablon; H J Lantner; T M McCormack; S Nair; E Barker; P Black
Journal:  J Clin Monit       Date:  1993-07

9.  Aggressive use of ICP monitoring is safe and alters patient care.

Authors:  V A Eddy; J L Vitsky; E J Rutherford; J A Morris
Journal:  Am Surg       Date:  1995-01       Impact factor: 0.688

10.  Infectious complications and duration of intracranial pressure monitoring.

Authors:  R K Kanter; L B Weiner; A M Patti; L K Robson
Journal:  Crit Care Med       Date:  1985-10       Impact factor: 7.598

View more
  2 in total

Review 1.  The Insertion and Management of External Ventricular Drains: An Evidence-Based Consensus Statement : A Statement for Healthcare Professionals from the Neurocritical Care Society.

Authors:  Herbert I Fried; Barnett R Nathan; A Shaun Rowe; Joseph M Zabramski; Norberto Andaluz; Adarsh Bhimraj; Mary McKenna Guanci; David B Seder; Jeffrey M Singh
Journal:  Neurocrit Care       Date:  2016-02       Impact factor: 3.210

2.  Direct cost comparison of minimally invasive punch technique versus traditional approaches for percutaneous bone anchored hearing devices.

Authors:  Yaeesh Sardiwalla; Nicholas Jufas; David P Morris
Journal:  J Otolaryngol Head Neck Surg       Date:  2017-06-12
  2 in total