Literature DB >> 8922577

European multicentre evaluation of a commercial system for identification of methicillin-resistant Staphylococcus aureus.

G Zambardi1, J Fleurette, G C Schito, R Auckenthaler, E Bergogne-Berezin, R Hone, A King, W Lenz, C Lohner, A Makristhatis, F Marco, C Müller-Serieys, C Nonhoff, I Phillips, P Rohner, M Rotter, K P Schaal, M Struelens, A Viebahn.   

Abstract

A commercial system for the rapid detection of methicillin-resistant Staphylococcus aureus, the BBL Crystal MRSA test (C-MRSA ID; Becton Dickinson, USA), was evaluated prospectively and compared with a polymerase chain reaction test for the presence of the mecA gene. Ten European centres tested a total of 676 isolates of Staphylococcus aureus from blood cultures. The system correctly identified 661 (97.8%) isolates within 4 h. All but three mecA gene-negative isolates (99.4% specificity) yielded a negative C-MRSA ID reaction, and 158 of 170 mecA gene-positive isolates were accurately detected (92.9% sensitivity). After repeated testing of discrepant results, sensitivity and specificity increased to 99% and 100%, respectively.

Entities:  

Mesh:

Year:  1996        PMID: 8922577     DOI: 10.1007/bf01691964

Source DB:  PubMed          Journal:  Eur J Clin Microbiol Infect Dis        ISSN: 0934-9723            Impact factor:   3.267


  8 in total

Review 1.  Methicillin-resistant staphylococci.

Authors:  H F Chambers
Journal:  Clin Microbiol Rev       Date:  1988-04       Impact factor: 26.132

2.  Laboratory diagnosis of oxacillin resistance in Staphylococcus aureus by a multiplex-polymerase chain reaction assay.

Authors:  G Zambardi; M E Reverdy; S Bland; M Bes; J Freney; J Fleurette
Journal:  Diagn Microbiol Infect Dis       Date:  1994-05       Impact factor: 2.803

3.  Low-affinity penicillin-binding protein associated with beta-lactam resistance in Staphylococcus aureus.

Authors:  B J Hartman; A Tomasz
Journal:  J Bacteriol       Date:  1984-05       Impact factor: 3.490

4.  Molecular cloning of the gene of a penicillin-binding protein supposed to cause high resistance to beta-lactam antibiotics in Staphylococcus aureus.

Authors:  M Matsuhashi; M D Song; F Ishino; M Wachi; M Doi; M Inoue; K Ubukata; N Yamashita; M Konno
Journal:  J Bacteriol       Date:  1986-09       Impact factor: 3.490

5.  Evaluation of differential inoculum disk diffusion method and Vitek GPS-SA card for detection of oxacillin-resistant staphylococci.

Authors:  C C Knapp; M D Ludwig; J A Washington
Journal:  J Clin Microbiol       Date:  1994-02       Impact factor: 5.948

6.  Evaluation of MicroScan rapid gram-positive panels for detection of oxacillin-resistant staphylococci.

Authors:  G L Woods; D LaTemple; C Cruz
Journal:  J Clin Microbiol       Date:  1994-04       Impact factor: 5.948

7.  Evaluation of BBL crystal MRSA ID system.

Authors:  C C Knapp; M D Ludwig; J A Washington
Journal:  J Clin Microbiol       Date:  1994-10       Impact factor: 5.948

8.  Evaluation of Rapid ATB Staph for 5-hour antimicrobial susceptibility testing of Staphylococcus aureus. Groupement pour le Dépistage, L'Etude et la Prévention des Infections Hospitalières-Groep ter Opsporing, Studie en Preventie van Infecties in de Ziekenhuizen.

Authors:  M J Struelens; C Nonhoff; P van der Auwera; R Mertens; E Serruys
Journal:  J Clin Microbiol       Date:  1995-09       Impact factor: 5.948

  8 in total
  7 in total

1.  Evaluation of the BBL Crystal MRSA ID System for detection of oxacillin resistance in Staphylococcus aureus.

Authors:  G Kampf; C Lecke; A K Cimbal; K Weist; H Rüden
Journal:  J Clin Pathol       Date:  1999-03       Impact factor: 3.411

2.  Detection of methicillin-resistant Staphylococcus aureus using the BBL crystal MRSA kit.

Authors:  J M Hamilton-Miller
Journal:  Eur J Clin Microbiol Infect Dis       Date:  1997-06       Impact factor: 3.267

3.  Rapid slide latex agglutination test for detection of methicillin resistance in Staphylococcus aureus.

Authors:  A van Griethuysen; M Pouw; N van Leeuwen; M Heck; P Willemse; A Buiting; J Kluytmans
Journal:  J Clin Microbiol       Date:  1999-09       Impact factor: 5.948

4.  Oxacillin susceptibility testing of staphylococci directly from Bactec Plus blood cultures by the BBL Crystal MRSA ID system.

Authors:  M Kubina; B Jaulhac; X Delabranche; C Lindenmann; Y Piemont; H Monteil
Journal:  J Clin Microbiol       Date:  1999-06       Impact factor: 5.948

5.  Comparison of ATB staph, rapid ATB staph, Vitek, and E-test methods for detection of oxacillin heteroresistance in staphylococci possessing mecA.

Authors:  N B Frebourg; D Nouet; L Lemée; E Martin; J F Lemeland
Journal:  J Clin Microbiol       Date:  1998-01       Impact factor: 5.948

6.  Rapid identification of Staphylococcus aureus by using fluorescent staphylocoagulase assays.

Authors:  M G Holliday; M Ford; J D Perry; F K Gould
Journal:  J Clin Microbiol       Date:  1999-04       Impact factor: 5.948

7.  Biosensor for detection of antibiotic resistant Staphylococcus bacteria.

Authors:  Rajesh Guntupalli; Iryna Sorokulova; Eric Olsen; Ludmila Globa; Oleg Pustovyy; Vitaly Vodyanoy
Journal:  J Vis Exp       Date:  2013-05-08       Impact factor: 1.355

  7 in total