| Literature DB >> 8027310 |
G L Woods1, D LaTemple, C Cruz.
Abstract
The ability of MicroScan rapid panels (Baxter MicroScan, West Sacramento, Calif.) to detect oxacillin resistance in 92 clinical isolates of Staphylococcus aureus and 103 coagulase-negative staphylococci was evaluated by comparing results with those of MicroScan 24-h MIC panels and, to resolve discrepancies, oxacillin agar screening. Both panels were interpreted by the MicroScan WalkAway-96 system. Rapid panels detected 96.7% of resistant S. aureus isolates and 72% of resistant coagulase-negative staphylococci, 22 of which did not grow in the panels.Entities:
Mesh:
Substances:
Year: 1994 PMID: 8027310 PMCID: PMC267181 DOI: 10.1128/jcm.32.4.1058-1059.1994
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948