Literature DB >> 8915414

Improved and simplified methods for specifying positions of the electrode bands of a cochlear implant array.

L T Cohen1, J Xu, S A Xu, G M Clark.   

Abstract

OBJECTIVE: To develop techniques for measuring the positions of the individual electrodes of a multiple channel cochlear implant and for estimating associated characteristic frequencies.
BACKGROUND: Information concerning the positions of the individual electrodes of a cochlear implant array is important for analyzing speech perception or psychophysical data and for optimizing speech-processing strategies. This study presents two techniques for obtaining such information from postoperative plain film radiographs.
METHODS: A template spiral shape, derived from analysis of the radiographs of 30 cochlear implant patients, is used to obtain measurements of the angular positions of the electrode bands within scala tympani. A research technique measures angular positions and estimates characteristic frequencies for all electrode bands but requires that the positions of two cochlear landmarks and all electrode bands be digitized. A clinical technique provides similar angle and frequency estimates but requires a minimum of information to be extracted visually from the radiograph. The lateral positions of the bands are estimated, in the research technique, using mean outer and inner wall functions obtained from 11 Silastic molds of scala tympani.
RESULTS: The mean position of the implanted array relative to the mean scala tympani outer wall function was consistent with published histologic observations of implanted temporal bones. Measured angles did not vary greatly with experimenter or with rotation of the cochlea relative to the radiographic beam by up to 20 degrees.
CONCLUSIONS: The techniques described allow, principally, measurement of the longitudinal positions of the bands of a cochlear implant in scala tympani and estimation of corresponding characteristic frequencies.

Entities:  

Mesh:

Year:  1996        PMID: 8915414

Source DB:  PubMed          Journal:  Am J Otol        ISSN: 0192-9763


  27 in total

1.  Combined spectral and temporal enhancement to improve cochlear-implant speech perception.

Authors:  Aparajita Bhattacharya; Andrew Vandali; Fan-Gang Zeng
Journal:  J Acoust Soc Am       Date:  2011-11       Impact factor: 1.840

2.  Tonotopic relationships reveal the charge density varies along the lateral wall of outer hair cells.

Authors:  Christian Corbitt; Federica Farinelli; William E Brownell; Brenda Farrell
Journal:  Biophys J       Date:  2012-06-19       Impact factor: 4.033

3.  Frequency map for the human cochlear spiral ganglion: implications for cochlear implants.

Authors:  Olga Stakhovskaya; Divya Sridhar; Ben H Bonham; Patricia A Leake
Journal:  J Assoc Res Otolaryngol       Date:  2007-02-21

4.  Psychophysical versus physiological spatial forward masking and the relation to speech perception in cochlear implants.

Authors:  Michelle L Hughes; Lisa J Stille
Journal:  Ear Hear       Date:  2008-06       Impact factor: 3.570

5.  High-resolution secondary reconstructions with the use of flat panel CT in the clinical assessment of patients with cochlear implants.

Authors:  M S Pearl; A Roy; C J Limb
Journal:  AJNR Am J Neuroradiol       Date:  2013-12-26       Impact factor: 3.825

6.  Bionic ear imaging.

Authors:  R Cerini; N Faccioli; M Barillari; M De Iorio; M Carner; V Colletti; R Pozzi Mucelli
Journal:  Radiol Med       Date:  2008-04-02       Impact factor: 3.469

7.  Partial withdrawal of deeply inserted cochlear electrodes: observations of two patients.

Authors:  Maria-Izabel Kos; Colette Boex; Jean-Philippe Guyot; Marco Pelizzone
Journal:  Eur Arch Otorhinolaryngol       Date:  2007-06-12       Impact factor: 2.503

8.  Estimation of insertion depth angle based on cochlea diameter and linear insertion depth: a prediction tool for the CI422.

Authors:  Annett Franke-Trieger; Dirk Mürbe
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-11-02       Impact factor: 2.503

9.  Investigation of ultra-low insertion speeds in an inelastic artificial cochlear model using custom-made cochlear implant electrodes.

Authors:  Silke Hügl; Katharina Rülander; Thomas Lenarz; Omid Majdani; Thomas S Rau
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-10-09       Impact factor: 2.503

10.  Initial Operative Experience and Short-term Hearing Preservation Results With a Mid-scala Cochlear Implant Electrode Array.

Authors:  Maja Svrakic; J Thomas Roland; Sean O McMenomey; Mario A Svirsky
Journal:  Otol Neurotol       Date:  2016-12       Impact factor: 2.311