A wall-less vessel phantom for Doppler ultrasound studies.

D W Rickey1, P A Picot, D A Christopher, A Fenster.   

Abstract

Doppler ultrasound flow measurement techniques are often validated using phantoms that simulate the vasculature, surrounding tissue and blood. Many researchers use rubber tubing to mimic blood vessels because of the realistic acoustic impedance, robust physical properties and wide range of available sizes. However, rubber tubing has a very high acoustic attenuation, which may introduce artefacts into the Doppler measurements. We describe the construction of a wall-less vessel phantom that eliminates the highly attenuating wall and reduces impedance mismatches between the vessel lumen and tissue mimic. An agar-based tissue mimic and a blood mimic are described and their acoustic attenuation coefficients and velocities are characterised. The high attenuation of the latex rubber tubing resulted in pronounced shadowing in B-mode images; however, an image of a wall-less vessel phantom did not show any shadowing. We show that the effects of the highly attenuating latex rubber vessels on Doppler amplitude spectra depend on the vessel diameter and ultrasound beam width. In this study, only small differences were observed in spectra obtained from 0.6 cm inside diameter thin-wall latex, thick-wall latex and wall-less vessel phantoms. However, a computer model predicted that the spectrum obtained from a 0.3-cm inside diameter latex-wall vessel would be significantly different than the spectrum obtained from a wall-less vessel phantom, thus resulting in an overestimation of the average fluid velocity. These results suggest that care must be taken to ensure that the Doppler measurements are not distorted by the highly attenuating wall material. In addition, the results show that a wall-less vessel phantom is preferable when measuring flow in small vessels.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8849831     DOI: 10.1016/0301-5629(95)00044-5

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  22 in total

1.  Development of an in vivo tumor-mimic model for learning radiofrequency ablation.

Authors:  M D Scott DJ; W N Young; L M Watumull; G Lindberg; J B Fleming; R V Rege; R J Brown; D B Jones
Journal:  J Gastrointest Surg       Date:  2000 Nov-Dec       Impact factor: 3.452

2.  A novel realistic three-layer phantom for intravascular ultrasound imaging.

Authors:  J Brunette; R Mongrain; G Cloutier; M Bertrand; O F Bertrand; J C Tardif
Journal:  Int J Cardiovasc Imaging       Date:  2001-10       Impact factor: 2.357

3.  High-frequency dynamics of ultrasound contrast agents.

Authors:  Yang Sun; Dustin E Kruse; Paul A Dayton; Katherine W Ferrara
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2005-11       Impact factor: 2.725

4.  Pulsatile flow phantom for ultrasound image-guided HIFU treatment of vascular injuries.

Authors:  Robyn Greaby; Vesna Zderic; Shahram Vaezy
Journal:  Ultrasound Med Biol       Date:  2007-04-27       Impact factor: 2.998

5.  Evaluation of guidewire path reproducibility.

Authors:  Sebastian Schafer; Kenneth R Hoffmann; Peter B Noël; Ciprian N Ionita; Jacek Dmochowski
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

Review 6.  Developments in cardiovascular ultrasound. Part 2: Arterial applications.

Authors:  P R Hoskins; P J Fish; W N McDicken; C Moran
Journal:  Med Biol Eng Comput       Date:  1998-05       Impact factor: 2.602

7.  Pipe Phantoms With Applications in Molecular Imaging and System Characterization.

Authors:  Shiying Wang; Elizabeth B Herbst; Stephen D Pye; Carmel M Moran; John A Hossack
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2016-11-09       Impact factor: 2.725

8.  Pulsed focused ultrasound-induced displacements in confined in vitro blood clots.

Authors:  Cameron C Wright; Kullervo Hynynen; David E Goertz
Journal:  IEEE Trans Biomed Eng       Date:  2011-12-21       Impact factor: 4.538

9.  Ultrasound settings significantly alter arterial lumen and wall thickness measurements.

Authors:  Kathleen Potter; Christopher J Reed; Daniel J Green; Graeme J Hankey; Leonard F Arnolda
Journal:  Cardiovasc Ultrasound       Date:  2008-01-22       Impact factor: 2.062

10.  Poly(vinyl alcohol) gel ultrasound phantom with durability and visibility of internal flow.

Authors:  Kenichi Funamoto; Osamu Yamashita; Toshiyuki Hayase
Journal:  J Med Ultrason (2001)       Date:  2014-08-05       Impact factor: 1.314

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.