Measurement and comparison of head scatter factor for 7 MV unflattened (FFF) and 6 MV flattened photon beam using indigenously designed columnar mini phantom.

Sigamani Ashokkumar1, Arunai Nambiraj2, Sujit Nath Sinha3, Girigesh Yadav3, Kothanda Raman3, Manindra Bhushan3, Rajesh Thiyagarajan1.   

Abstract

AIM: To measure and compare the head scatter factor for 7 MV unflattened and 6 MV flattened photon beam using a home-made designed mini phantom.
BACKGROUND: The head scatter factor (Sc) is one of the important parameters for MU calculation. There are multiple factors that influence the Sc values, like accelerator head, flattening filter, primary and secondary collimators.
MATERIALS AND METHODS: A columnar mini phantom was designed as recommended by AAPM Task Group 74 with high and low atomic number material for measurement of head scatter factors at 10 cm and d max dose water equivalent thickness.
RESULTS: The Sc values measured with high-Z are higher than the low-Z mini phantoms observed for both 6MV-FB and 7MV-UFB photon energies. Sc values of 7MV-UFB photon beams were smaller than those of the 6MV-FB photon beams (0.6-2.2% (Primus), 0.2-1.4% (Artiste) and 0.6-3.7% (Clinac iX (2300CD))) for field sizes ranging from 10 cm × 10 cm to 40 cm × 40 cm. The SSD had no influence on head scatter for both flattened and unflattened beams. The presence of wedge filters influences the Sc values. The collimator exchange effects showed that the opening of the upper jaw increases Sc irrespective of FF and FFF.
CONCLUSIONS: There were significant differences in Sc values measured for 6MV-FB and unflattened 7MV-UFB photon beams over the range of field sizes from 10 cm × 10 cm to 40 cm × 04 cm. Different results were obtained for measurements performed with low-Z and high-Z mini phantoms.

Entities:  

Keywords:  6MV-FB; 7MV-UFB; Linear accelerators; Mini phantom; Sc factor

Year:  2015        PMID: 25949220      PMCID: PMC4418585          DOI: 10.1016/j.rpor.2015.02.001

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  42 in total

1.  Clinical dosimetry for implementation of a multileaf collimator.

Authors:  A L Boyer; T G Ochran; C E Nyerick; T J Waldron; C J Huntzinger
Journal:  Med Phys       Date:  1992 Sep-Oct       Impact factor: 4.071

2.  Measurements of in-air output ratios for a linear accelerator with and without the flattening filter.

Authors:  X R Zhu; Y Kang; M T Gillin
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

3.  Monte Carlo study of backscatter in a flattening filter free clinical accelerator.

Authors:  U Titt; O N Vassiliev; F Pönisch; S F Kry; R Mohan
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

4.  Characteristics of the high-energy photon beam of a 25-MeV accelerator.

Authors:  G Luxton; M A Astrahan
Journal:  Med Phys       Date:  1988 Jan-Feb       Impact factor: 4.071

5.  The fraction of photons undergoing head scatter in x-ray beams.

Authors:  T C Zhu; B E Bjärngard
Journal:  Phys Med Biol       Date:  1995-06       Impact factor: 3.609

6.  Revision of tissue-maximum ratio and scatter-maximum ratio concepts for cobalt 60 and higher energy x-ray beams.

Authors:  F M Khan; W Sewchand; J Lee; J F Williamson
Journal:  Med Phys       Date:  1980 May-Jun       Impact factor: 4.071

7.  Characteristics of an 18 MV photon beam from a Therac 20 Medical Linear Accelerator.

Authors:  M S Patterson; P C Shragge
Journal:  Med Phys       Date:  1981 May-Jun       Impact factor: 4.071

8.  Electron contamination of a high-energy X-ray beam.

Authors:  T N Padikal; J A Deye
Journal:  Phys Med Biol       Date:  1978-11       Impact factor: 3.609

9.  Build-up cap materials for measurement of photon head-scatter factors.

Authors:  L Weber; P Nilsson; A Ahnesjö
Journal:  Phys Med Biol       Date:  1997-10       Impact factor: 3.609

Review 10.  Radiation-induced second cancers: the impact of 3D-CRT and IMRT.

Authors:  Eric J Hall; Cheng-Shie Wuu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-05-01       Impact factor: 7.038

View more
  3 in total

1.  Comparison of surface dose delivered by 7 MV-unflattened and 6 MV-flattened photon beams.

Authors:  Ashokkumar Sigamani; Arunai Nambiraj
Journal:  Rep Pract Oncol Radiother       Date:  2017-04-25

2.  Interpretation of In-air Output Ratio of Wedged Fields in Different Measurement Conditions.

Authors:  Parinaz Mehnati; Farideh Biglari; Ali Jomehzadeh
Journal:  J Med Signals Sens       Date:  2019 Apr-Jun

3.  Surface dose measurements and comparison of unflattened and flattened photon beams.

Authors:  Ashokkumar Sigamani; Arunai Nambiraj; Girigesh Yadav; Ananda Giribabu; Karthikeyan Srinivasan; Venkadamanickam Gurusamy; Kothanda Raman; Kaviarasu Karunakaran; Rajesh Thiyagarajan
Journal:  J Med Phys       Date:  2016 Apr-Jun
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.