Extrafocal radiation: a unified approach to the prediction of beam penumbra and output factors for megavoltage x-ray beams.

M B Sharpe1, D A Jaffray, J J Battista, P Munro.   

Abstract

An extrafocal source model has been developed to explain the dependence of head scatter and beam penumbra on field size. In this model, the x-ray source of a medical linear accelerator is described by two components: a small but intense focal component; and a broadly distributed extrafocal component of low intensity. The extrafocal component is so large that it can be "eclipsed" by the field-defining collimators. Extrafocal radiation was found to account for 12% of the energy fluence on the central axis of the 6 MV x-ray beam from a Varian Clinac 2100 c accelerator. Head scatter factors were calculated "in-air" for symmetric, asymmetric, and half-blocked fields. Calculations agreed with measured values to better than 0.5%, on average. However, head scatter factors for asymmetric fields were underestimated by 1.2% when one of the field dimensions was reduced to 4 cm (the minimum jaw setting that was tested). The extrafocal source model was combined with a convolution/superposition dose calculation algorithm to calculate dose-per-monitor-unit calibration (output) factors and beam dose profiles in water. These dose calculations predict the degradation of the field edge as a function of field size, and calculate output factors to within 0.5%, on average. In the most extreme case of a 4 cm field width, output factors were underestimated by 2%. Dose profiles are predicted without the aid of an empirical fit to measured beam penumbra data. The extrafocal source model will be particularly useful for fields defined by independent jaw and multileaf collimation systems.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8746712     DOI: 10.1118/1.597648

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Multisource modeling of flattening filter free (FFF) beam and the optimization of model parameters.

Authors:  Woong Cho; Kayla N Kielar; Ed Mok; Lei Xing; Jeong-Hoon Park; Won-Gyun Jung; Tae-Suk Suh
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

2.  Estimation of focal and extra-focal radiation profiles based on Gaussian modeling in medical linear accelerators.

Authors:  Shigeo Anai; Hidetaka Arimura; Katsumasa Nakamura; Fujio Araki; Takaomi Matsuki; Hideki Yoshikawa; Satoshi Yoshidome; Yoshiyuki Shioyama; Hiroshi Honda; Nobuo Ikeda
Journal:  Radiol Phys Technol       Date:  2011-03-24

3.  Measurement and comparison of head scatter factor for 7 MV unflattened (FFF) and 6 MV flattened photon beam using indigenously designed columnar mini phantom.

Authors:  Sigamani Ashokkumar; Arunai Nambiraj; Sujit Nath Sinha; Girigesh Yadav; Kothanda Raman; Manindra Bhushan; Rajesh Thiyagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2015-03-01

4.  Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71.

Authors:  John P Gibbons; John A Antolak; David S Followill; M Saiful Huq; Eric E Klein; Kwok L Lam; Jatinder R Palta; Donald M Roback; Mark Reid; Faiz M Khan
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

5.  Modification and validation of an analytical source model for external beam radiotherapy Monte Carlo dose calculations.

Authors:  Scott E Davidson; Jing Cui; Stephen Kry; Joseph O Deasy; Geoffrey S Ibbott; Milos Vicic; R Allen White; David S Followill
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

6.  An Experimental Slope Method for a More Accurate Measurement of Relative Radiation Doses using Radiographic and Radiochromic Films and Its Application to Megavoltage Small-Field Dosimetry.

Authors:  Raghavendra Holla; D Khanna; Bhaskaran K Pillai; K V Jafar Ali; P S Renil Mon; C O Clinto; Tharmarnadar Ganesh
Journal:  J Med Phys       Date:  2019 Jul-Sep

7.  Comparison of Head Scatter Factor for 6MV and 10MV flattened (FB) and Unflattened (FFF) Photon Beam using indigenously Designed Columnar Mini Phantom.

Authors:  Sigamani Ashokkumar; N Arunai Nambi Raj; Sujit Nath Sinha; Girigesh Yadav; Rajesh Thiyagarajan; Kothanda Raman; Manindra Bhushan Mishra
Journal:  J Med Phys       Date:  2014-07
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.