Distortion of intravascular ultrasound images because of nonuniform angular velocity of mechanical-type transducers.

B J Kimura1, V Bhargava, W Palinski, R J Russo, A N DeMaria.   

Abstract

The purpose of this study was to quantify nonuniform rotation in a current mechanical intravascular ultrasound (IVUS) instrument and its effect on arc, area, and diameter measurements. The accurate reconstruction of IVUS two-dimensional images is dependent on uniform rotation of the catheter tip. Prior investigations suggested that bends in the catheter driveshaft may be responsible for poor torque transmission, nonuniform rotation, and consequent errors in IVUS measurements. Eight 30 MHz mechanically driven IVUS catheters were evaluated in a model simulating the catheter course through the aorta and coronary ostium in a clinical study. Angular velocity and posi-ion profiles of the transducer, image angle, and diameter and area measurement errors were obtained from each catheter by imaging a vascular phantom with eight equispaced echogenic markers from concentric and eccentric positions. Six catheters also were tested for comparison in a simple curvature model. Rotational error was found in all catheters tested and worsened in the aortic model. Maximal angular error, defined as the largest angle between actual and presumed transducer direction, increased when measured in the aortic model as compared with the simple curvature model (17 +/- 12 degrees to 45 +/- 25 degrees; p < 0.05). Angles of 45 degrees were misrepresented with a mean range of values of 26 to 63 degrees. With eccentric catheter placement, area and diameters had average maximal absolute errors of 26% +/- 7.8% and 23% +/- 10%, respectively. In conclusion, nonuniform rotation of mechanical IVUS transducers constitutes a significant potential source of error in IVUS measurement of arcs of calcification, and lumen shape, area, and diameter.

Entities:  

Mesh:

Year:  1996        PMID: 8701894     DOI: 10.1016/s0002-8703(96)90429-9

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  7 in total

1.  Assessment of the mechanical properties of coronary arteries using intravascular ultrasound: an in vivo study.

Authors:  M J Williams; R A Stewart; C J Low; G T Wilkins
Journal:  Int J Card Imaging       Date:  1999-08

2.  Reliability of mechanical and phased-array designs for serial intravascular ultrasound examinations--animal and clinical studies in stented and non-stented coronary arteries.

Authors:  J C Tardif; O F Bertrand; R Mongrain; J Lespérance; J Grégoire; P Paiement; R Bonan
Journal:  Int J Card Imaging       Date:  2000-10

3.  A novel realistic three-layer phantom for intravascular ultrasound imaging.

Authors:  J Brunette; R Mongrain; G Cloutier; M Bertrand; O F Bertrand; J C Tardif
Journal:  Int J Cardiovasc Imaging       Date:  2001-10       Impact factor: 2.357

4.  Towards a geometrically correct 3-D reconstruction of tortuous coronary arteries based on biplane angiography and intravascular ultrasound.

Authors:  G P Prause; S C DeJong; C R McKay; M Sonka
Journal:  Int J Card Imaging       Date:  1997-12

5.  Factors influencing the intravascular ultrasound dimensions and areas measured using mechanical transducers.

Authors:  V Bhargava; W F Penny; A Arbab-Zadeh
Journal:  Int J Card Imaging       Date:  1996-12

6.  Investigating the impact of spatial priors on the performance of model-based IVUS elastography.

Authors:  M S Richards; M M Doyley
Journal:  Phys Med Biol       Date:  2011-10-28       Impact factor: 3.609

7.  Non-rigid image registration based strain estimator for intravascular ultrasound elastography.

Authors:  Michael S Richards; Marvin M Doyley
Journal:  Ultrasound Med Biol       Date:  2012-12-15       Impact factor: 2.998

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.