Randomized evaluation of pulse oximetry in 20,802 patients: I. Design, demography, pulse oximetry failure rate, and overall complication rate.

J T Moller1, T Pedersen, L S Rasmussen, P F Jensen, B D Pedersen, O Ravlo, N H Rasmussen, K Espersen, N W Johannessen, J B Cooper.   

Abstract

BACKGROUND: Although pulse oximetry is currently in widespread use, there are few data documenting improvement in patient outcome as a result of the use of oximetry. The authors describe the study design, patient demographic findings, data validation, pulse oximetry failure rate, and overall postoperative complication rates in the first large prospective randomized multicenter clinical trial on perioperative pulse oximetry monitoring.
METHODS: In five Danish hospitals, by random assignment, monitoring did or did not include pulse oximetry for patients 18 yr of age and older, whether scheduled for elective or emergency operations, or for regional or general anesthesia, except during cardiac and neurosurgical procedures. Operational definitions were established for perioperative events and postoperative complications. The data were collected preoperatively, during anesthesia, in the postanesthesia care unit, and until the day of discharge from the hospital or the seventh postoperative day.
RESULTS: Of 20,802 patients, 10,312 were assigned to the oximetry group and 10,490, to the control group. In general, the demographic data, patient factors, and anesthetic agents used were distributed evenly. A slight intergroup difference was found in the distribution of age, duration of surgery, some types of surgery, and some types of anesthesia. The total failure rate of the oximetry was 2.5%, but it increased to 7.2% in patients with American Society of Anesthesiologists physical status 4 (P < 0.00001). In 14.9% of the patients, one or more events occurred in the operating room and 13.5% in the postanesthesia care unit. The overall postoperative complication rate was 9.7%. The total rates of cardiovascular and respiratory complications were 2.78% and 3.50%, respectively. Within the first seven postoperative days, 0.47% of the patients died. Anesthesia was not thought to have been solely responsible for any death, but in 7 patients (1 per 3,365), it was a possible contributory factor.
CONCLUSIONS: Despite the finding of a few significant inter-group differences, the randomization was well balanced with a high validity of data. The overall postoperative complication rate was similar to that in other recent morbidity and mortality studies.

Entities:  

Mesh:

Year:  1993        PMID: 8457044     DOI: 10.1097/00000542-199303000-00006

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  30 in total

1.  Qualitative versus quantitative research--balancing cost, yield and feasibility. 1993.

Authors:  W B Runciman
Journal:  Qual Saf Health Care       Date:  2002-06

2.  Comparison of a prototype esophageal oximetry probe with two conventional digital pulse oximetry monitors in aortocoronary bypass patients.

Authors:  R C Prielipp; P E Scuderi; M H Hines; J L Atlee; J F Butterworth
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

3.  Pulse oximetry.

Authors:  Amal Jubran
Journal:  Intensive Care Med       Date:  2004-07-24       Impact factor: 17.440

4.  Arterial line for monitoring SpO2 in patients with ischemic peripheries.

Authors:  Goneppanavar Umesh; Kaur Jasvinder; Shetty Nanda; Shetty Ranjan
Journal:  J Clin Monit Comput       Date:  2010-07-16       Impact factor: 2.502

Review 5.  Using EEG to monitor anesthesia drug effects during surgery.

Authors:  Leslie C Jameson; Tod B Sloan
Journal:  J Clin Monit Comput       Date:  2006-12       Impact factor: 2.502

Review 6.  What type of monitoring has been shown to improve outcomes in acutely ill patients?

Authors:  Gustavo A Ospina-Tascón; Ricardo L Cordioli; Jean-Louis Vincent
Journal:  Intensive Care Med       Date:  2008-01-05       Impact factor: 17.440

7.  Pro: is pulse oximetry still worthwhile?

Authors:  E C Pierce
Journal:  J Clin Monit Comput       Date:  1998-07       Impact factor: 2.502

8.  Con: is pulse oximetry still worthwhile?

Authors:  F K Orkin
Journal:  J Clin Monit Comput       Date:  1998-07       Impact factor: 2.502

9.  Influence of pulse oximeter settings on the frequency of alarms and detection of hypoxemia: Theoretical effects of artifact rejection, alarm delay, averaging, median filtering or a lower setting of the alarm limit.

Authors:  A T Rheineck-Leyssius; C J Kalkman
Journal:  J Clin Monit Comput       Date:  1998-04       Impact factor: 2.502

Review 10.  Technology assessment: a Canadian perspective.

Authors:  R J Byrick
Journal:  Can J Anaesth       Date:  1996-05       Impact factor: 5.063

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.