Arterial line for monitoring SpO2 in patients with ischemic peripheries.

Goneppanavar Umesh1, Kaur Jasvinder, Shetty Nanda, Shetty Ranjan.   

Abstract

Monitoring the oxygenation status of patients with poor peripheral perfusion or ischemic peripheries is challenging in view of unreliable or unrecordable pulse oximeter data. In this article we describe a very simple and innovative technique of using the arterial line for reliable recording of arterial oxygen saturation (SpO(2)) in such patients. We conclude that the arterial line can be used as an extension of the artery and SpO(2) may be reliably measured using the arterial line in such patients as long as the blood in the arterial tubing is pulsatile and a good contact is ensured between the arterial tubing and the sensor of the pulse oximeter.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20635123     DOI: 10.1007/s10877-010-9245-1

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  15 in total

1.  Digital pulp space infiltration of local anesthetic enhances pulse oximeter signal

Authors: 
Journal:  Anesth Analg       Date:  2000-09       Impact factor: 5.108

2.  A simple solution.

Authors:  A Gupta; M Vegfors
Journal:  Anaesthesia       Date:  1992-09       Impact factor: 6.955

3.  Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers.

Authors:  R J Falconer; B J Robinson
Journal:  Br J Anaesth       Date:  1990-10       Impact factor: 9.166

4.  Digital nerve blocks and pulse oximeter signal detection.

Authors:  D W Eastwood
Journal:  Anesth Analg       Date:  1992-06       Impact factor: 5.108

5.  Digital nerve blocks can restore pulse oximeter signal detection.

Authors:  D L Bourke; R F Grayson
Journal:  Anesth Analg       Date:  1991-12       Impact factor: 5.108

6.  Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery.

Authors:  H Pälve; A Vuori
Journal:  Crit Care Med       Date:  1989-01       Impact factor: 7.598

7.  Pulse oximeter failure thresholds in hypotension and vasoconstriction.

Authors:  J W Severinghaus; M J Spellman
Journal:  Anesthesiology       Date:  1990-09       Impact factor: 7.892

8.  Digital block of the flexor tendon sheath can restore pulse oximeter signal detection.

Authors:  M E Gentili; E Chevaleraud; E Viel
Journal:  Reg Anesth       Date:  1995 Jan-Feb

9.  Intraarterial vasodilator administration to restore pulse oximeter function.

Authors:  K Holroyd; M Lui; C Beattie
Journal:  Anesthesiology       Date:  1993-08       Impact factor: 7.892

10.  Continuous assessment of arterial blood gases.

Authors: 
Journal:  Crit Care       Date:  1997       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.