Literature DB >> 8404955

Quantification of technetium-99m hexamethylpropylene amine oxime brain uptake in routine clinical practice using calibrated point sources as an external standard: phantom and human studies.

A Dobbeleir1, R Dierckx.   

Abstract

Quantitative methods for calculation of regional cerebral blood flow with technetium-99m hexamethylpropylene amine oxime (99mTc-HMPAO) have been proposed. These methods are very labour intensive and therefore are not useful in routine clinical practice. We describe a simple alternative method, using calibrated point sources as a scaling factor, whereby the tomographic slices are displayed as regional 99mTc-HMPAO brain uptake per cm3 brain tissue in 10(-6) of the injected lipophilic dose. The method was validated on Jaszczak and Hoffman phantoms using a three-detector system with HR parallel and HR fan-beam collimators. Under the optimal conditions described in this paper, the measured to real activity ratio was 1.00 (SD = 0.06). The reproducibility of the cerebellar uptake in a group of ten normal volunteers and five patients was studied. Intra-individually a mean deviation of 12.6% was observed for the total group. For those persons with a heart rate difference of less than 5 units between the two studies, a mean deviation of 7.2% was obtained. Quantitative 99mTc-HMPAO brain uptake images can be useful for longitudinal studies, especially for follow-up, activation and pharmacological studies.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8404955     DOI: 10.1007/bf00181759

Source DB:  PubMed          Journal:  Eur J Nucl Med        ISSN: 0340-6997


  14 in total

1.  A quantitative approach to technetium-99m hexamethylpropylene amine oxime.

Authors:  H Matsuda; S Tsuji; N Shuke; H Sumiya; N Tonami; K Hisada
Journal:  Eur J Nucl Med       Date:  1992

2.  Effect of spatial resolution on SPECT quantification values.

Authors:  A Kojima; M Matsumoto; M Takahashi; Y Hirota; H Yoshida
Journal:  J Nucl Med       Date:  1989-04       Impact factor: 10.057

3.  The in vivo distribution of 99Tcm-HM-PAO in normal man.

Authors:  D C Costa; P J Ell; I D Cullum; P H Jarritt
Journal:  Nucl Med Commun       Date:  1986-09       Impact factor: 1.690

4.  The retention of [99mTc]-d,l-HM-PAO in the human brain after intracarotid bolus injection: a kinetic analysis.

Authors:  N A Lassen; A R Andersen; L Friberg; O B Paulson
Journal:  J Cereb Blood Flow Metab       Date:  1988-12       Impact factor: 6.200

5.  Quantification of regional cerebral blood flow (rCBF) using 99Tcm-HMPAO and SPECT: choice of the reference region.

Authors:  G M Syed; S Eagger; B K Toone; R Levy; J J Barrett
Journal:  Nucl Med Commun       Date:  1992-11       Impact factor: 1.690

6.  Kinetic analysis of technetium-99m d,1-HM-PAO decomposition in aqueous media.

Authors:  J C Hung; M Corlija; W A Volkert; R A Holmes
Journal:  J Nucl Med       Date:  1988-09       Impact factor: 10.057

7.  The behavior of 99mTc-hexamethylpropyleneamineoxime (99mTc-HMPAO) in blood and brain.

Authors:  K Nakamura; Y Tukatani; A Kubo; S Hashimoto; Y Terayama; T Amano; F Goto
Journal:  Eur J Nucl Med       Date:  1989

8.  Development of a 99Tcm-labelled radiopharmaceutical for cerebral blood flow imaging.

Authors:  D P Nowotnik; L R Canning; S A Cumming; R C Harrison; B Higley; G Nechvatal; R D Pickett; I M Piper; V J Bayne; A M Forster
Journal:  Nucl Med Commun       Date:  1985-09       Impact factor: 1.690

9.  An analysis of the arterial input curve for technetium-99m-HMPAO: quantification of rCBF using single-photon emission computed tomography.

Authors:  A Pupi; M T De Cristofaro; L Bacciottini; D Antoniucci; A R Formiconi; M Mascalchi; U Meldolesi
Journal:  J Nucl Med       Date:  1991-08       Impact factor: 10.057

10.  Technetium-99m HM-PAO stereoisomers as potential agents for imaging regional cerebral blood flow: human volunteer studies.

Authors:  P F Sharp; F W Smith; H G Gemmell; D Lyall; N T Evans; D Gvozdanovic; J Davidson; D A Tyrrell; R D Pickett; R D Neirinckx
Journal:  J Nucl Med       Date:  1986-02       Impact factor: 10.057

View more
  3 in total

1.  External reference markers for the correction of head rotation in brain single-photon emission tomography.

Authors:  W D Leslie; A Borys; D McDonald; J O Dupont; A E Peterdy
Journal:  Eur J Nucl Med       Date:  1995-04

2.  Technetium-99m HMPAO SPET in acute supratentorial ischaemic infarction, expressing deficits as millilitre of zero perfusion.

Authors:  R A Dierckx; A Dobbeleir; B A Pickut; L Timmermans; I Dierckx; A Vervaet; J Vandevivere; W Deberdt; P P De Deyn
Journal:  Eur J Nucl Med       Date:  1995-05

3.  Parameters influencing SPET regional brain uptake of technetium-99m hexamethylpropylene amine oxine measured by calibrated point sources as an external standard.

Authors:  R A Dierckx; A Dobbeleir; M Maes; B A Pickut; A Vervaet; P P De Deyn
Journal:  Eur J Nucl Med       Date:  1994-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.