Literature DB >> 8289713

Investigation of the convolution method for polyenergetic spectra.

N Papanikolaou1, T R Mackie, C Meger-Wells, M Gehring, P Reckwerdt.   

Abstract

The distribution of absolute dose per unit fluence from polyenergetic photon beams impinging upon a water phantom was calculated using two convolution approaches that properly account for beam hardening effects. Dose deposition kernels calculated previously using the EGS4 Monte Carlo code are convolved with the primary terma to give the dose for monoenergetic photon beams of energies ranging from 100 kev to 50 MeV. A polyenergetic dose distribution is composed of separately calculated monoenergetic components, which are appropriately weighted with the fluence spectrum to yield the polyenergetic dose distribution. Alternatively, a single convolution for the polyenergetic beam is considered, where a composite polyenergetic kernel is convolved with the respective polyenergetic terma. The effects of the polyenergetic kernel variance due to beam hardening as well as the effect of tilting the kernels for a diverging beam geometry were also examined. The depth dose data produced using the two proposed methods were compared with measured data and Monte Carlo simulations and showed good agreement.

Entities:  

Mesh:

Year:  1993        PMID: 8289713     DOI: 10.1118/1.597154

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  GPU-accelerated Monte Carlo convolution/superposition implementation for dose calculation.

Authors:  Bo Zhou; Cedric X Yu; Danny Z Chen; X Sharon Hu
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

2.  Investigation of various energy deposition kernel refinements for the convolution∕superposition method.

Authors:  Jessie Y Huang; David Eklund; Nathan L Childress; Rebecca M Howell; Dragan Mirkovic; David S Followill; Stephen F Kry
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

3.  High-precision radiotherapy for craniospinal irradiation: evaluation of three-dimensional conformal radiotherapy, intensity-modulated radiation therapy and helical TomoTherapy.

Authors:  D S Sharma; T Gupta; R Jalali; Z Master; R D Phurailatpam; R Sarin
Journal:  Br J Radiol       Date:  2009-07-06       Impact factor: 3.039

4.  A convolution/superposition method using primary and scatter dose kernels formed for energy bins of X-ray spectra reconstructed as a function of off-axis distance: a theoretical study on 10-MV X-ray dose calculations in thorax-like phantoms.

Authors:  Akira Iwasaki; Shigenobu Kimura; Kohji Sutoh; Kazuo Kamimura; Makoto Sasamori; Fumio Komai; Morio Seino; Singo Terashima; Mamoru Kubota; Junichi Hirota; Yoichiro Hosokawa
Journal:  Radiol Phys Technol       Date:  2011-06-15

5.  Comparison of the helical tomotherapy against the multileaf collimator-based intensity-modulated radiotherapy and 3D conformal radiation modalities in lung cancer radiotherapy.

Authors:  P Mavroidis; C Shi; G A Plataniotis; M G Delichas; B Costa Ferreira; S Rodriguez; B K Lind; N Papanikolaou
Journal:  Br J Radiol       Date:  2010-09-21       Impact factor: 3.039

6.  Analytical model for out-of-field dose in photon craniospinal irradiation.

Authors:  Phillip J Taddei; Wassim Jalbout; Rebecca M Howell; Nabil Khater; Fady Geara; Kenneth Homann; Wayne D Newhauser
Journal:  Phys Med Biol       Date:  2013-10-08       Impact factor: 3.609

7.  Dosimetric evaluation of heterogeneity corrections for RTOG 0236: stereotactic body radiotherapy of inoperable stage I-II non-small-cell lung cancer.

Authors:  Ying Xiao; Lech Papiez; Rebecca Paulus; Robert Timmerman; William L Straube; Walter R Bosch; Jeff Michalski; James M Galvin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-03-15       Impact factor: 7.038

8.  Validation of the Pinnacle³ photon convolution-superposition algorithm applied to fast neutron beams.

Authors:  Alan M Kalet; George A Sandison; Mark H Phillips; Upendra Parvathaneni
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

9.  Comparison of monitor unit calculations performed with a 3D computerized planning system and independent "hand" calculations: results of three years clinical experience.

Authors:  Jackson Chan; David Russell; Victor G Peters; Thomas J Farrell
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

10.  Comparing two strategies of dynamic intensity modulated radiation therapy (dIMRT) with 3-dimensional conformal radiation therapy (3DCRT) in the hypofractionated treatment of high-risk prostate cancer.

Authors:  Jasper Yuen; George Rodrigues; Kristina Trenka; Terry Coad; Slav Yartsev; David D'Souza; Michael Lock; Glenn Bauman
Journal:  Radiat Oncol       Date:  2008-01-07       Impact factor: 3.481

View more