| Literature DB >> 12383049 |
Jackson Chan1, David Russell, Victor G Peters, Thomas J Farrell.
Abstract
A comparison of the monitor unit calculations of a commercial 3D computerized treatment planning system (TPS) with "hand" calculations from lookup tables was made for a large number of clinical cases (greater than 13 500 treatment fields). Differences were analyzed by treatment site for prostate, rectum, cranium, and breast. The 3D TPS monitor unit calculation was systematically higher than the "hand" calculation by an amount that depended on the complexity of the treatment geometry. For simple geometries the mean difference was 1% and was as high as 3% for more complicated geometries. The higher value was attributed to an accumulation of differences introduced by multiple factors in the monitor unit calculation. Careful attention to factors such as patient contour could reduce the mean difference. "Hand" calculations were shown to be an accurate and useful tool for verification of TPS monitor unit calculations.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12383049 PMCID: PMC5724537 DOI: 10.1120/jacmp.v3i4.2553
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Factors used for “hand” monitor unit calculation.
| Factor | Symbol | Definition (dependence) |
|---|---|---|
| Output calibration |
| Dose in cGy/MU in calibration conditions. 1 cGy/MU at SAD for reference depth and field size. |
| Collimator scatter factor |
| Dose rate in air for a given collimator setting relative to that for the reference collimator setting (field size). |
| Phantom scatter factor |
| Dose rate at reference depth for a given field relative to that at the same depth for the reference field, using the same collimator setting (depth, field size). |
| Tissue phantom ratio | TPR | Dose rate at depth relative to dose rate at the reference depth for the same field size (depth, field size). |
| Tray attenuation factor | TF | Attenuation factor due to shielding tray (none). |
| Wedge attenuation factor | WF | Attenuation due to transmission through physical wedge (depth, field size). |
| Off‐axis ratio | OAR | Dose rate at off‐axis position relative to dose rate at the central axis (off‐axis distance). |
| Wedge off‐axis ratio | WOAR | Attenuation through wedge at off‐axis position relative to that through the central axis (off‐axis distance). |
| Inverse square correction | ISC | Dose rate in air at prescription distance relative to that at standard SAD |
Statistical summary of MU ratios by treatment site.
| Statistic/Site | Prostate | Rectum | Brain | Breast |
|---|---|---|---|---|
| Number of fields | 3577 | 1948 | 741 | 7510 |
| Average MU ratio | 1.010 | 1.011 | 1.013 | 1.012 |
| St. dev. MU ratio | 0.005 | 0.008 | 0.008 | 0.016 |
| Min. MU ratio | 0.992 | 0.984 | 0.977 | 0.954 |
| Max. MU ratio | 1.034 | 1.051 | 1.043 | 1.077 |
Figure 1A histogram showing the distribution of MU ratios by site for Pinnacle TPS compared with “hand” calculations.
Figure 2A histogram showing the distribution of MU ratios for Pinnacle3 TPS compared with “hand” calculations for prostate. The original “hand” calculation did not account for small corner blocks in the field, whereas the recalculation did.
Statistical summary of MU ratios for breast treatments by technique.
| Statistic/Technique | Cobalt | 6 MV | Four‐field tangent | Four‐field supraclav |
|---|---|---|---|---|
| Number of Fields | 4113 | 1906 | 755 | 679 |
| Average MU ratio | 1.007 | 1.013 | 1.017 | 1.032 |
| St. dev. MU ratio | 0.015 | 0.010 | 0.018 | 0.012 |
| Min. MU ratio | 0.966 | 0.954 | 0.957 | 0.987 |
| Max. MU ratio | 1.072 | 1.061 | 1.072 | 1.077 |
Figure 3A histogram showing the distribution of MU ratios by treatment technique for Pinnacle3 TPS compared with “hand” calculations for breast treatment.
Figure 4A histogram showing the distribution of MU ratios for Pinnacle3 TPS compared with “hand” calculations for breast tangent fields. The original hand calculation used an equivalent square based on patient contour whereas the recalculation assumes a flat patient entrance surface and corrects only for beam “splash” beyond the patient contour.
Mean MU ratios for supraclavicular fields with decreasing levels of complexity.
| Average MU ratio | |||
|---|---|---|---|
| Plan Geometry | Component of Complexity | Anterior | Posterior |
| Original Plan | ‐ | 1.028 | 1.027 |
| Wedges Removed | WF and WOAR | 1.024 | 1.021 |
| Flat contour, no shielding | dTAR equivalent square | 1.016 | 1.017 |
| X and Y jaws reversed | Collimator exchange effect | 1.013 | 1.014 |
| Asymmetric field size and Prescription point position | OAR | 1.011 | 1.011 |