The effects of pressure and flow on hemolysis caused by Bio-Medicus centrifugal pumps and roller pumps. Guidelines for choosing a blood pump.

Y Tamari1, K Lee-Sensiba, E F Leonard, V Parnell, A J Tortolani.   

Abstract

Two Bio-Medicus BP-50 centrifugal pumps and two roller pumps were tested simultaneously with porcine blood at 21 degrees +/- 1 degree C in four in vitro circuits to determine the effect of four combinations of flow and pressure conditions on blood damage. Flows of 300 ml/min (1/4-inch inner-diameter tubing in the roller pump) and 1775 ml/min (1/2-inch inner-diameter tubing in the roller pump) and pressure differences across the pump (delta P = outlet pressure--inlet pressure) of 215 mm Hg (n = 6) and 345 mm Hg (n = 5) were examined. The index of hemolysis (milligrams plasma hemoglobin per 100 L blood pumped) for the BP-50 pump was higher at a flow of 300 ml/min than at a flow of 1775 ml/min (p < 0.0002). At 300 ml/min, the index of hemolysis for the BP-50 pump tended to be higher at 345 mm Hg than at 215 mm Hg (mean +/- standard error of the mean, 135 +/- 22 versus 88 +/- 9, p = 0.059). At 1775 ml/min, there was no difference in the index of hemolysis for the BP-50 pump between 215 and 345 mm Hg (37 +/- 7 versus 29 +/- 5, p = 0.32). With the roller pump, the index of hemolysis was higher at a flow of 300 ml/min than at a flow of 1775 ml/min (p < 0.036), but there was no difference in the indexes of hemolysis between 215 and 345 mm Hg at 300 ml/min (60 +/- 9 versus 61 +/- 11, p = 0.93) or at 1775 ml/min (40 +/- 6 versus 36 +/- 6, p = 0.61). Comparison between the two types of pumps showed that the index of hemolysis was significantly higher for the BP-50 than for the roller pump at a flow of 300 ml/min and a delta P of 215 mm Hg (88 +/- 9 versus 60 +/- 9, p = 0.009), as well as at a flow of 300 ml/min and a delta P of 345 mm Hg (135 +/- 22 versus 61 +/- 11, p = 0.001). At a flow of 1775 ml/min, there was no difference in the index of hemolysis between the two pumps at either pressure condition.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8246582

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  4 in total

1.  Performance evaluation of a pediatric viscous impeller pump for Fontan cavopulmonary assist.

Authors:  Guruprasad A Giridharan; Steven C Koenig; Jeffrey Kennington; Michael A Sobieski; Jun Chen; Steven H Frankel; Mark D Rodefeld
Journal:  J Thorac Cardiovasc Surg       Date:  2012-03-14       Impact factor: 5.209

Review 2.  Extracorporeal life support in critically ill adults.

Authors:  Corey E Ventetuolo; Christopher S Muratore
Journal:  Am J Respir Crit Care Med       Date:  2014-09-01       Impact factor: 21.405

3.  Membrane Lung and Blood Pump Use During Prolonged Extracorporeal Membrane Oxygenation: Trends From 2002 to 2017.

Authors:  McKenzie M Hayes; Brian P Fallon; Ryan P Barbaro; Niki Manusko; Robert H Bartlett; John M Toomasian
Journal:  ASAIO J       Date:  2021-09-01       Impact factor: 3.826

4.  Clinical evaluation of the Spiral Pump® after improvements to the original project in patients submitted to cardiac surgeries with cardiopulmonary bypass.

Authors:  Jarbas Jakson Dinkhuysen; Aron Jose Pazin de Andrade; Juliana Leme; Cibele Silva; Claudia Sanches Medina; Cristiane Célia Pereira; José Francisco Biscegli
Journal:  Rev Bras Cir Cardiovasc       Date:  2014 Jul-Sep
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.