OBJECTIVE: The objective of this paper is to present the results from Spiral Pump clinical trial after design modifications performed at its previous project. This pump applies axial end centrifugal hydraulic effects for blood pumping during cardiopulmonary bypass for patients under cardiac surgery. METHODS: This study was performed in 52 patients (51% males), between 20 to 80 (67±14.4) years old weighing 53 to 102 (71.7±12.6) kg, mostly under myocardial revascularization surgery (34.6%) and valvular surgery (32.8%). Besides the routine evaluation of the data observed in these cases, we monitored pump rotational speed, blood flow, cardiopulmonary bypass duration, urine free hemoglobin for blood cell trauma analysis (+ to 4+), lactate desidrogenase (UI/L), fibrinogen level (mg/dL) and platelet count (nº/mm3). RESULTS: Besides maintaining appropriate blood pressure and metabolic parameters it was also observed that the Free Hemoglobin levels remained normal, with a slight increase after 90 minutes of cardiopulmonary bypass. The Lactate Dehydrogenase showed an increase, with medians varying between 550-770 IU/L, whereas the decrease in Fibrinogen showed medians of 130-100 mg/dl. The number of platelets showed a slight decrease with the medians ranging from 240,000 to 200,000/mm3. No difficulty was observed during perfusion terminations, nor were there any immediate deaths, and all patients except one, were discharged in good condition. CONCLUSION: The Spiral Pump, as blood propeller during cardiopulmonary bypass, demonstrated to be reliable and safe, comprising in a good option as original and national product for this kind of application.
OBJECTIVE: The objective of this paper is to present the results from Spiral Pump clinical trial after design modifications performed at its previous project. This pump applies axial end centrifugal hydraulic effects for blood pumping during cardiopulmonary bypass for patients under cardiac surgery. METHODS: This study was performed in 52 patients (51% males), between 20 to 80 (67±14.4) years old weighing 53 to 102 (71.7±12.6) kg, mostly under myocardial revascularization surgery (34.6%) and valvular surgery (32.8%). Besides the routine evaluation of the data observed in these cases, we monitored pump rotational speed, blood flow, cardiopulmonary bypass duration, urine free hemoglobin for blood cell trauma analysis (+ to 4+), lactate desidrogenase (UI/L), fibrinogen level (mg/dL) and platelet count (nº/mm3). RESULTS: Besides maintaining appropriate blood pressure and metabolic parameters it was also observed that the Free Hemoglobin levels remained normal, with a slight increase after 90 minutes of cardiopulmonary bypass. The Lactate Dehydrogenase showed an increase, with medians varying between 550-770 IU/L, whereas the decrease in Fibrinogen showed medians of 130-100 mg/dl. The number of platelets showed a slight decrease with the medians ranging from 240,000 to 200,000/mm3. No difficulty was observed during perfusion terminations, nor were there any immediate deaths, and all patients except one, were discharged in good condition. CONCLUSION: The Spiral Pump, as blood propeller during cardiopulmonary bypass, demonstrated to be reliable and safe, comprising in a good option as original and national product for this kind of application.
The Division of Bioengineering from Instituto Dante Pazzanese de Cardiologia (IDPC)
started in 1992 a research aiming to develop a blood pump for cardiopulmonary bypass
(CPB), called Spiral Pump® (SP). The SP was idealized as a device to drive
blood minimizing damage to blood cells during CPB[, and also, providing
lower cost by being produced in Brazil protected by a Patent of Invention number PI
91,022,215 from the National Institute of Industrial Property.SP uses two hydraulic pumping principles simultaneously, centrifugal and axial, provided
by conically shaped rotor with threads on its surface, increasing the hydrodynamic
pumping performance without increasing rates of blood cell damage such as
hemolysis[.Between 1994 and 1999, a pilot number of devices was produced, for in
vivo tests (sheep), based on an experimental protocol, approved by IDPC's
Experimental Research Ethics Committee. SP was used as a pumping device for CPB. From
1999 to 2002, clinical evaluations were performed on 43 patients during cardiac surgery
with CPB[. After approved in
clinical evaluations, the device was registered at the National Health Surveillance
Agency (ANVISA) under number 10264470020.In 2007, the development of a new rotor design had started, using two entrances at the
threads of the spiral impeller pumps and also, modifying the device external base,
aiming to have same coupling system as other devices, allowing its use routinely in
cardiac surgeries. During this improvement process, many additional in
vitro and in vivo tests were performed[.A new in vivo tests were performed according to a new experiment
protocol, number 2010/003, approved by the IDPC's Ethics Committee for Animal Usage, and
with those new results, SP was able to undergo to a new clinic evaluation phase. For
this phase, new devices were assembled inside controlled environment, according to ISO
14644, ISO 8 and ISO 13485, and sterilized by ethylene oxide (ETO)[.As the original prototype (Figure 1A), this new
model uses centrifugal and axial forces simultaneously as hydraulic principles,
improving its hydrodynamic performance without increasing blood damage. This pump has in
its interior a spiral conic rotor to generate pumping pressures. Axial pumping effect is
produced due to double entrance threads at impeller surface, and centrifugal pumping
effect occurs because of impeller conical shape. At the top of the housing, there is the
inlet port and, at the bottom, the outlet port, both with 3/8-inch connections (Figure 1B). Housing base consists of a plastic disc
with one stainless steel shaft that holds all internal pump components. Housing base
provides total isolation between pump inside and outside. Motor torque is transmitted to
spiral impeller by magnetic coupling, using two annular magnets, one inside the impeller
and another fixed to the shaft of the driver motor.
Fig. 1
A Original prototype, developed in 1992 and 1B - The new model, developed in 2007,
with two thread entrances and with the modified device's external base[.
Entrada=Intelet port; Cone externo=Outer spire; Fuso=Rotor; Saída=Outlet port;
Base externa do cone=Housing base
A Original prototype, developed in 1992 and 1B - The new model, developed in 2007,
with two thread entrances and with the modified device's external base[.Entrada=Intelet port; Cone externo=Outer spire; Fuso=Rotor; Saída=Outlet port;
Base externa do cone=Housing basePreliminary studies have shown that changes in rotor design caused important changes in
device performance and positive impact was noted during hemolysis tests due to the new
conical shape and new rotor design, as well as making lower the pump outlet port (Figures 2 and 3)[. Previous studies, conducted in 1998, showed that the
original pump performance was 3000 rpm under a pressure of 350 mmHg, and under similar
pumping flow as the reference model Biopump® (BPX80, Medtronic, USA).
However, it was around 2500 rpm for the same pressure with the new SP model[.
Fig. 2
A First model and 2B - Rotor final model
Fig. 3
A Housing initial model and 3B - Housing final model with change in output port
position
A First model and 2B - Rotor final modelA Housing initial model and 3B - Housing final model with change in output port
position
Objectives
This work aims to conduct clinical evaluation with Spiral Pump (SP) during
cardiopulmonary bypass (CPB), as an impeller blood pump for patients under cardiac
surgery.
METHODS
Inclusion criteria:• Indication for cardiac surgery with CPB in most myocardial revascularization and
orovalvular lesions;• With or without cardioplegia;• Age between 40 and 80 years;• Male and female;• Body weight between 55 and 120 kg.Exclusion criteria:• Cardiac surgery without CPB;• Abnormalities in laboratory tests (coagulogram);• Surgery for patients with underweight (<55 kg).Study was conducted in a group of 52 patients, 27 male (52%), age of 20-80 (67.0±14.4)
years, weight of 53-102 (71.7±12.6) kg.Indications for surgery were:• Surgical Myocardial Revascularization (SMR) 19 (36.5%);• Orovalvar surgery 17 (32.6%), SMR + orovalvar 11 (21.1%);• Other 2 (3.8%);• Valvar + other 1 (1.9%);• Congenital + orovalvar 1 (1.9%);• Congenital 1 (1.9%).CPB circuit components were: SP, drive module (Bioconsole 550®, Medtronic,
Minneapolis, USA), flowmeter (Bioprobe® TX50, Medtronic, Minneapolis, USA),
CPB machine (Macchi, Sao Paulo, Brazil), tube kit with blood filter (Nipro Medical,
Sorocaba, Brazil), Vital adult membrane oxygenator (Nipro Medical, Sorocaba, Brazil) (in
18 cases) and Quadrox adult oxygenator - Softlibe Coating and adult hemoconcentrator -
HPH 1000 (Maquet, Germany) (in 34 cases).All clinical evaluation procedures were performed by the same surgeon, principal
investigator, and two perfusionists to maintain reliable results.Routine procedures for cardiac surgeries were used, such as general anesthesia with
endotracheal intubation for pulmonary ventilation, assessment of body temperature by an
esophageal thermometer, assessment of average medial arterial pressure (MAP) by
catheterization of peripheral arterial line and central venous pressure (CVP) by
catheterization of central vein, urinary catheterization to evaluate urinary flow and
heart rate monitoring with specific monitor.The exposure of the heart was performed by median thoracotomy and, after systemic
heparinization (5 mg/kg), CPB was installed by the ascending aorta or femoral artery for
oxygenated blood intake from the oxygenator, pumped by the SP, and, from the right
atrium, with single cannula or selective of higher and lower cava to drain venous blood
toward the membrane oxygenator reservoir to CO2/O2 exchange.After CPB initiation and stabilization of metabolic and perfusion parameters with
esophageal temperature around 32ºC, pulmonary ventilation was stopped and the
perfusionist, every 5 minutes, took balance notes from MAP, CVP, diuresis, temperature,
flow and occasional events. As usual, at the beginning, middle and end of the procedure,
samples for laboratory tests were collected (blood gases, A/V, K+,
Na+, hematocrit, glucose) that were used as standard for possible
therapeutic interventions during the procedure. The anticoagulation control during CPB
were performed by measuring Activated Coagulation Time (ACT) in an activated coagulation
monitor MCA2000 (Adib Jatene Foundation, Sao Paulo, Brazil), trying to keep it above 500
seconds. At the end of CPB, reversion of heparin was performed with protamine sulfate,
as previously established standard. Completing the surgical procedure, the patient was
taken to Postoperative Recovery Unit under mechanical ventilation, ECG monitoring and
MAP, diuresis and temperature controls.In order to evaluate the interactions between SP and blood, specific investigated
parameters in this study were: LDH (U/L), fibrinogen (g/L), platelet count
(nº/mm3), ACT (seconds) and the urinalysis tape in pre CPB (5 minutes
before), every 15 minutes during the procedure and after CPB (5 minutes after infusion)
in all patients in the operating room before being referred to the Postoperative
Recovery Unit.For subsequent evaluation, ACT was determined for the period that preceded the CPB,
during (30, 60, 90 and 120 minutes) and after anticoagulation reversing at the end of
the procedure.All patients have been previously contacted by researchers in the preoperative period,
according to surgery schedule, and were told about the study objectives, risks and
benefits, in accordance to Informed Consent Statement signed by each participant of the
Clinical Evaluation Protocol, No. 4072 of April 5, 2011, approved by IDPC´s Research
Ethics Committee.
Statistical analysis
Describing patient samples, attendance and percentage were used for qualitative
variables and average, standard deviation, median and interquartile range for
quantitative variables.Two-tailed Fisher's exact tests were made for qualitative variables and Kruskal
Wallis for quantitative ones. For the variables that have association
(P<0,05) in the Kruskal Wallis tests, multiple comparisons
were made (2 by 2 tests), through the non parametric Turkey method, to verify which
groups are different between themselves.ANOVA non-parametric tests were also performed, with time being the only factor, with
n greater than or equal to 5. Charts profiles were used with 95% confidence intervals
for the averages.
RESULTS
Time of CPB ranged from 30 to 240 (111.2±47.7) minutes in all cases and no problem was
observed during perfusion outcome.There are 12 graphs in Figure 4 with observed
parameters from device clinical evaluation. First graph shows rotational speed of SP
during CPB, always related to weight (body surface area) for each patient, small
variations between 2000 - 2500 rpm were observed. The flow provided by SP (second graph)
shows oscillations from 3.7 to 3.3 L/min with a higher standard deviation at final
moments during the procedure.
Fig. 4
Chart with observed parameters during procedures
Chart with observed parameters during proceduresConsidering as appropriate the MAP value of 60 mmHg during CPB, we observed that the
curve oscillates right below this level with higher standard deviation at final moments
during the procedure (third graph), this fact has not caused any organic damage due to
the lowering of esophageal temperature near 30ºC, reducing the metabolic and perfusion
requirements. Urinary volume generated during CPB varied around 150 ml at first moments,
decreasing around 100 ml from 30 to 150 minutes (fourth graph). Related to laboratory
data, we emphasize that all time scales of the following graphs ranged from pre CPB,
0-120 minutes and post CPB.The pH values ranged close to normal values of 7.45 (fifth graph), with larger
oscillation after 75 minutes until the end of the procedure. The Sodium (Na) levels
(sixth graph) ranged below 139 mg/L and increased after 90 minutes of perfusion,
maintaining this level until the end of the procedure, so did the Potassium (K) levels
(seventh graph). As expected, the hematocrit (Ht) showed significant decrease after
perfusion initiation, this is due to hemodilution imposed by oxygenator priming, usually
with saline solutions, and remained around 30% until 105 minutes, increasing at the end
of the procedure (eighth graph), the same occurs to hemoglobin (Hb) values (ninth
graph). Blood glucose (tenth graph) varied around 130 mg/L initially to levels near 200
mg/L at final moments of CPB and considering infusion of serum glycosylated during
anesthesia.ACT analysis (eleventh graph) shows adequate levels of anticoagulation during procedure,
ranging from 1500 to 1800 seconds. Finally, free hemoglobin analysis from urine, which
represents the mechanical trauma impact to the blood (twelfth graph), shows levels close
to 0 within 90 minutes, when we observed an increase of values from 0 to 1 with
increased standard deviation, common fact observed at the end of CPB.Considering the values of lactic dehydrogenase, we observed an increase of the medians
ranging between 500-770 IU/L (Figure 5 and Chart
1). The levels of fibrinogen showed a decrease of medians between 130-100 mg/dl (Figure 5 and Chart 2), both proportional to higher
CPB time. Related to number of platelets (Figure 5
and Chart 3), the medians ranged between 240,000/m3 for the 0 to 60 minutes of CPB group
and 20,000/mm3 for the 0 to >120 minutes group, denoting small traumatic
impact to the blood.
Fig. 5
Box Plot with observed parameters during procedures
Box Plot with observed parameters during proceduresThe values of lactic dehydrogenase (VN 305UI/L) in pre CPB ranged between 238-994
(427.6±134.5) and post CPB between 380-1954 (926.7±350.1) IU/L. Fibrinogen values (VN
200-400 mg/dl) before CPB ranged between 164.2 to 403.2 (267±56.4) and after CPB between
145.5 to 439.3 (215.3±48.1) mg/dl, and Platelet (VN 150000-400000/mm3) showed
before CPB between 103000-307000 (209.100±64.900/mm3) and after CPB between
66000-233000 (146,000 ± 43,000/m3).Figure 5 shows the Box Plot analyzing the
percentage of Lactate Dehydrogenase (Chart 1), fibrinogen (Chart 2) and Platelet Count
(Chart 3) for three groups of patients, a CPB group from 0 to 60 minutes, a group from 0
to 120 minutes and a group with time above 120 minutes, observing the expected results
for this procedure.There were no immediate deaths, observing ICU stay ranged from 1 to 4 (average 2.4)
days. One patient remained 13 and another remained 44 days for presenting postoperative
complications that required intensive controls, the patients in question were diabetic
and smokers, complicating the postoperative period. One death occurred on the
1st postoperative day due to bleeding by clotting disorder.Patients were discharged from 3 to 14 (average 6.5) days and only one remained for
longer due to postoperative complications and for being diabetic and smoker, and was
discharged on the 61st postoperative day in good condition.
DISCUSSION
During the project development, the rapid prototyping process helped to make possible
the construction and assembly of different SP models in a short period of time,
moreover, allowed cost reduction in the manufacture of disposable product process, to
avoid costly manufacturing of different plastic injection molds.Various changes were performed to hydrodynamically test the performance curves (pressure
x flow), showing that, for each prototype, changes in pumping characteristics were made,
until we reach the most efficient one, that has been selected for clinical
use[.While developing the design of medical products, such as blood pumps for CPB, patients
can be adversely affected by the use of devices that may cause high levels of plasma
free hemoglobin, especially in the postoperative period. Therefore, the project is based
on consistent in vitro and in vivo tests that can
generate product with good acceptance and competitiveness for clinical
application[.Structural characteristics of a blood pump are important in determining its hydrodynamic
performance and blood trauma. Often, small structural changes cause large changes to the
results, as shown in this work[.Currently, roller pumps are used in cardiac surgery and centrifugal pumps for severe
cases. The roller pump is widely used due to low cost, easy usage and simplicity, its
flow is linear with slight pulsatility, but generates high negative pressure when
sucking blood to the CPB circuit, which can cause major blood damages[. Centrifugal pumps, mostly imported,
have higher costs, so they are not often used. The development of a blood pump with
national technology and low cost will enable more often usage in public and private
hospitals.During CPB, the flow, provided by a centrifugal pump, depends on the motor rotational
speed and on the pump output pressure that depends on the pressure drop of CPB circuit
and on its after load. Thus, when the peripheral resistance of the patient increases,
the blood pressure will increase also, causing decrease in the pump flow. This is a
self-compensatory effect that does not happen in peristaltic roller pumps. Roller pumps
provide a constant flow for a fixed rotation of the motor, in other words, independent
of blood pressure. So, if an increase in peripheral resistance of the patient occurs,
the blood pressure will rise too much, causing risks to the patient[.Compression of blood between collapsed structures, heating, friction during pumping,
irregular surfaces, turbulent flows, the stagnation flow and the rapid pressure
variations are among the main mechanisms of trauma in these devices.Hemolysis is always a reason for conducting research for development of blood pumps, it
is impossible to avoid it completely, so maintaining close to normal values found in the
literature is necessary[.Group under study consisted of adult patients undergoing cardiac surgery with
conventional CPB and SP as blood pump. In the procedures, perfusions during heart
surgery were followed from pre CPB to post CPB and, in this period, the operating data
of SP, such as rotational speed, blood flow, vibration and noise were controlled and
were considered satisfactory according to standards for cardiac surgery. The flow data,
comparing to the theoretical flow calculated for each patient, were within our
expectations and attending the needs for perfusion during surgery, demonstrating that SP
is suitable for perfusion of adult patients.Physiological and laboratory parameters such as diuresis, mean arterial pressure,
temperature and blood gas data were recorded. The selected as hemolysis indicators were
hemoglobin and hematocrit, which were satisfactory by means of comparison with normal
and expected post-perfusion values found in the literature[.In the case of hemoglobin, analysis over time shows that this parameter tends to fall
after the first fifteen minutes of perfusion, stabilizing below the reference value and
returning to normal after the end of perfusion. The same occurs with the hematocrit,
which, before CPB, presents mean values very close to normal minimum value, but still
shows higher than expected during perfusion and after CPB.In most cases, there was no free hemoglobin (Hb) in urine and its presence in urine was
seen only in cases over 115 minutes of perfusion. Specific parameters to this study,
Lactate Dehydrogenase, Platelets and Fibrinogen, have its analysis performed according
to pefusion time. We noted that time has a direct influence on the increase of LDH and
decreased platelets and fibrinogen, these parameters are indicators of hemolysis. But
the values found are still acceptable for the post CPB, which means that the pumping
system was adequate without traumatic impact.It is important to say that hemolysis mechanically caused is not only a consequence from
SP usage, but is a result of a whole perfusion system that includes the pump, arterial
filter, oxygenator, hemoconcentrator and plastic tubes.Cardiac surgery techniques and perfusion time were varied and the SP performance showed
suitable for all cases relating to easy rotation/flow adjustment by the perfusionist and
noise or vibration absence.The results from this study demonstrate that the perfusion system with the usage of SP
showed changes for indicative parameters within acceptable values and satisfactorily
under the requirements for its application in cardiac surgery, being able to be applied
safely.
Authors: Francisco Ubaldo Vieira; Reinaldo Wilson Vieira; Nilson Antunes; Orlando Petrucci; Pedro Paulo de Oliveira; Márcia Milena Pivatto Serra; Karlos Alexandre de Sousa Vilarinho; Marcio Roberto do Carmo Journal: Rev Bras Cir Cardiovasc Date: 2009 Apr-Jun
Authors: A Andrade; J Biscegli; J Dinkhuysen; J E Sousa; Y Ohashi; S Hemmings; J Glueck; K Kawahito; Y Nosé Journal: Artif Organs Date: 1996-06 Impact factor: 3.094
Authors: Juliana Leme; Jeison Fonseca; Eduardo Bock; Cibele da Silva; Bruno Utiyama da Silva; Alex Eugênio Dos Santos; Jarbas Dinkhuysen; Aron Andrade; José F Biscegli Journal: Artif Organs Date: 2011-05 Impact factor: 3.094