Comparison of Vitek and Cobas Micro systems with a semiautomated conventional microsystem for identification and susceptibility testing of gram negative bacilli.

A M Simoons-Smit1, D M Maclaren.   

Abstract

AIMS: To compare the sensitivity and specificity of two semiautomated systems against a conventional (MIC 2000) test system for the identification and antibiotic susceptibility of Gram negative bacteria.
METHODS: Clinical isolates of Gram negative bacilli (188 urinary and 229 non-urinary strains) were identified and tested for antibiotic susceptibility in the Cobas Micro and MIC 2000 systems. Of these, 359 strains were then tested in the Vitek and MIC 2000 systems. Two hundred and forty three strains were tested in all three systems immediately after isolation. Forty three were also tested only in the Vitek and MIC 2000 systems immediately after isolation. The remaining 174 strains were tested after storage at -20 degrees C for several months.
RESULTS: The Cobas Micro and MIC 2000 systems agreed on the identification of 310 of the 417 (74.3%) strains; the Vitek and MIC 2000 systems agreed on 338 of the 359 (94.2%) strains. The Cobas Micro system correctly identified 86.8% of strains tested after storage and 65.4% of those immediately after isolation. Organism-antibiotic combinations (non-urinary isolates) were tested in the Cobas Micro and MIC 2000 systems (n = 2335), in the Vitek and MIC 2000 systems (n = 999). Essential correlation (complete agreement plus minor errors) was observed in 98% (with 90% complete agreement) in the former and in 97% (with 86% complete agreement) in the latter. For the urinary isolates, 1949 organism-antibiotic combinations were analysed in the Cobas Micro and MIC 2000 systems where complete agreement was observed in 92% (with 3% very major discrepancies), for 1382 urinary organism-antibiotic combinations tested in the Vitek and MIC 2000 systems, the figures were 95% and 2%, respectively.
CONCLUSIONS: The Vitek system is highly accurate in the identification and antibiotic susceptibility testing of Gram negative bacteria. The Cobas Micro system has many shortcomings in its identification of Gram negative rods, especially freshly isolated strains, but it is comparable with the Vitek system in antibiotic susceptibility testing.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8132814      PMCID: PMC501761          DOI: 10.1136/jcp.47.1.71

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  9 in total

1.  Evaluation of a new commercial system for the identification of Enterobacteriaceae and non-fermentative bacteria.

Authors:  H K Geiss; M Geiss
Journal:  Eur J Clin Microbiol Infect Dis       Date:  1992-07       Impact factor: 3.267

Review 2.  Automated systems for identification of microorganisms.

Authors:  C E Stager; J R Davis
Journal:  Clin Microbiol Rev       Date:  1992-07       Impact factor: 26.132

3.  Comparison of the autoSCAN-W/A and Vitek Automicrobic systems for identification and susceptibility testing of bacteria.

Authors:  M R Visser; L Bogaards; M Rozenberg-Arska; J Verhoef
Journal:  Eur J Clin Microbiol Infect Dis       Date:  1992-11       Impact factor: 3.267

4.  Description and evaluation of the semiautomated 4-hour ATB 32E method for identification of members of the family Enterobacteriaceae.

Authors:  J Freney; C Herve; M Desmonceaux; F Allard; J M Boeufgras; D Monget; J Fleurette
Journal:  J Clin Microbiol       Date:  1991-01       Impact factor: 5.948

5.  Comparison of AMS-Vitek, MicroScan, and Autobac Series II for the identification of gram-negative bacilli.

Authors:  A L Truant; E Starr; C A Nevel; M Tsolakis; E F Fiss
Journal:  Diagn Microbiol Infect Dis       Date:  1989 May-Jun       Impact factor: 2.803

6.  Accuracy of microdilution and the AutoMicrobic System in detection of beta-lactam resistance in gram-negative bacterial mutants with derepressed beta-lactamase.

Authors:  J A Washington; C C Knapp; C C Sanders
Journal:  Rev Infect Dis       Date:  1988 Jul-Aug

7.  Rapid antimicrobial susceptibility testing of Gram-negative clinical isolates with the AutoMicrobic system.

Authors:  B A Backes; S J Cavalieri; J T Rudrik; E M Britt
Journal:  J Clin Microbiol       Date:  1984-06       Impact factor: 5.948

8.  Evaluation of the AutoMicrobic system for identification and susceptibility testing of gram-negative bacilli.

Authors:  B F Woolfrey; R T Lally; M N Ederer; C O Quall
Journal:  J Clin Microbiol       Date:  1984-12       Impact factor: 5.948

9.  Evaluation of the AutoMicrobic system for susceptibility testing of aminoglycosides and gram-negative bacilli.

Authors:  J A Hindler; D A Bruckner
Journal:  J Clin Microbiol       Date:  1987-03       Impact factor: 5.948

  9 in total
  2 in total

1.  Evaluation of the Wider system, a new computer-assisted image-processing device for bacterial identification and susceptibility testing.

Authors:  R Cantón; M Pérez-Vázquez; A Oliver; B Sánchez Del Saz; M O Gutiérrez; M Martínez-Ferrer; F Baquero
Journal:  J Clin Microbiol       Date:  2000-04       Impact factor: 5.948

2.  Evaluation of a liquid-phase colorimetric method for rapid antibacterial susceptibility testing.

Authors:  Siavosh Salmanzadeh-Ahrabi; Tanıl Kocagöz
Journal:  Iran J Microbiol       Date:  2013-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.