Metallic spinal artifacts in magnetic resonance imaging.

A R Vaccaro1, R M Chesnut, G Scuderi, J F Healy, J B Massie, S R Garfin.   

Abstract

STUDY
DESIGN: The magnetic resonance artifact susceptibility of traces of surgical aluminum, titanium, and stainless steel in a human spine model was investigated. Metallic filings were deposited in noncontiguous disc spaces in five human thoracic spines before magnetic resonance imaging with spin echo and gradient echo sequences.
OBJECTIVES: Spin echo and gradient echo sequences were used for quantitate and compare void artifact produced by commonly used surgical metals. This was compared to a liquid paraffin control. SUMMARY OF BACKGROUND DATA: No significant susceptibility artifact was seen with any metal in all spin echo sequences, including T1 (TR 600, TE 12), T2 (TR 2000, TE 30), proton density (TR 2000, TE 80), and fast T2 scanning (TR 3800, TE 96, Ef).
METHODS: Sagittal magnetic resonance imaging permitted void artifact quantification and comparison between different metallic alloys. Two neuroradiologists, working on a blinded basis, evaluated all data and rated the void susceptibility artifact on a scale of 1 (least) to 4 (greatest).
RESULTS: In general, the magnitude of an imaging artifact during magnetic resonance imaging correlated with the magnetism of the metal. Nickel, found in a larger concentration in 316L than in 304 stainless steel, decreases the magnetic resonance artifact of specific metals because of its ability to stabilize iron in a non-magnetic state. Therefore, the 316L stainless steel yielded less artifact production than 304 stainless steel on gradient echo imaging.
CONCLUSION: If upon gradient echo imaging in the postoperative period significant artifact production is noted, stainless steel deposition should be suspected as the causative agent. In this situation, spin echo techniques should be the first approach for attempting optimal visualization of the spinal cord and soft tissue structures.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8073315     DOI: 10.1097/00007632-199405310-00008

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  7 in total

1.  Long-term stability after multilevel cervical laminectomy for spinal cord tumor resection in von Hippel-Lindau disease.

Authors:  Ashok R Asthagiri; Gautam U Mehta; John A Butman; Martin Baggenstos; Edward H Oldfield; Russell R Lonser
Journal:  J Neurosurg Spine       Date:  2011-01-28

2.  The Contribution of Common Surgically Implanted Hardware to Functional MR Imaging Artifacts.

Authors:  A A Desai; M K Strother; C C Faraco; V L Morgan; T R Ladner; L M Dethrage; L C Jordan; M J Donahue
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

3.  The interobserver-validated relevance of intervertebral spacer materials in MRI artifacting.

Authors:  T Ernstberger; G Heidrich; T Bruening; S Krefft; G Buchhorn; H M Klinger
Journal:  Eur Spine J       Date:  2006-02-07       Impact factor: 3.134

4.  Implant detectibility of intervertebral disc spacers in post fusion MRI: evaluation of the MRI scan quality by using a scoring system--an in vitro study.

Authors:  Thorsten Ernstberger; Gabert Heidrich; Wolfgang Schultz; Eckhardt Grabbe
Journal:  Neuroradiology       Date:  2006-11-04       Impact factor: 2.804

5.  A standardized evaluation of artefacts from metallic compounds during fast MR imaging.

Authors:  Shumei Murakami; Rinus G Verdonschot; Miyoshi Kataoka; Naoya Kakimoto; Hiroaki Shimamoto; Sven Kreiborg
Journal:  Dentomaxillofac Radiol       Date:  2016-08-19       Impact factor: 2.419

6.  Artifacts in spine magnetic resonance imaging due to different intervertebral test spacers: an in vitro evaluation of magnesium versus titanium and carbon-fiber-reinforced polymers as biomaterials.

Authors:  Thorsten Ernstberger; Gottfried Buchhorn; Gabert Heidrich
Journal:  Neuroradiology       Date:  2009-05-26       Impact factor: 2.804

7.  Magnetic resonance imaging evaluation of intervertebral test spacers: an experimental comparison of magnesium versus titanium and carbon fiber reinforced polymers as biomaterials.

Authors:  T Ernstberger; G Buchhorn; G Heidrich
Journal:  Ir J Med Sci       Date:  2009-08-20       Impact factor: 1.568