Literature DB >> 7894555

Accuracy of fingerstick glucose values in shock patients.

H F Sylvain1, M E Pokorny, S M English, N H Benson, T W Whitley, C J Ferenczy, J G Harrison.   

Abstract

BACKGROUND: Fingerstick blood glucose measurement has become widespread in both hospital and prehospital settings.
OBJECTIVE: To determine the accuracy of fingerstick blood glucose measurement in patients with poor peripheral perfusion (shock).
METHOD: Results obtained during three methods of glucose analysis (fingerstick blood glucose measurement; bedside and laboratory glucose analysis) were examined prospectively on 38 patients from inpatient medical and surgical critical care units or the emergency department of a large tertiary care referral center.
RESULTS: The means of the three glucose measurements were significantly different. Univariate analysis of the mean laboratory glucose value versus the mean fingerstick glucose value was significantly different. The mean venipuncture glucose measured by the bedside glucose meter versus the mean venous laboratory glucose was not significantly different.
CONCLUSION: These results suggest that fingerstick blood samples should not be used for bedside glucose analysis in patients who may have inadequate tissue perfusion.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7894555

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  20 in total

1.  Suitability of capillary blood glucose analysis in patients receiving vasopressors.

Authors:  Myra F Ellis; Kesi Benjamin; Morgan Cornell; Kelsey Decker; Debra Farrell; Lynn McGugan; Gloria P Porter; Helen Shearin; Yanfang Zhao; Bradi B Granger
Journal:  Am J Crit Care       Date:  2013-09       Impact factor: 2.228

2.  Capillary versus venous bedside blood glucose estimations.

Authors:  R Boyd; B Leigh; P Stuart
Journal:  Emerg Med J       Date:  2005-03       Impact factor: 2.740

Review 3.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 4.  Challenges of inpatient blood glucose monitoring: standards, methods, and devices to measure blood glucose.

Authors:  Kathie L Hermayer; Aundrea S Loftley; Sumana Reddy; Satya Nandana Narla; Nina A Epps; Yusheng Zhu
Journal:  Curr Diab Rep       Date:  2015-03       Impact factor: 4.810

5.  Analysis: New point-of-care blood glucose monitoring system for the hospital demonstrates satisfactory analytical accuracy using blood from critically ill patients--an important step toward improved blood glucose control in the hospital.

Authors:  Jeffrey I Joseph
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

6.  Multicenter Observational Study of the First-Generation Intravenous Blood Glucose Monitoring System in Hospitalized Patients.

Authors:  Grant V Bochicchio; Brian R Hipszer; Michelle F Magee; Richard M Bergenstal; Anthony P Furnary; Angela M Gulino; Michael J Higgins; Peter C Simpson; Jeffrey I Joseph
Journal:  J Diabetes Sci Technol       Date:  2015-06-01

7.  Blood glucose level and outcome after cardiac arrest: insights from a large registry in the hypothermia era.

Authors:  Fabrice Daviaud; Florence Dumas; Nadège Demars; Guillaume Geri; Adrien Bouglé; Tristan Morichau-Beauchant; Yên-Lan Nguyen; Wulfran Bougouin; Frédéric Pène; Julien Charpentier; Alain Cariou
Journal:  Intensive Care Med       Date:  2014-03-25       Impact factor: 17.440

Review 8.  Challenges to glycemic measurement in the perioperative and critically ill patient: a review.

Authors:  Andrew D Pitkin; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

9.  Accuracy of bedside capillary blood glucose measurements in critically ill patients.

Authors:  C Dana Critchell; Vincent Savarese; Amy Callahan; Christine Aboud; Serge Jabbour; Paul Marik
Journal:  Intensive Care Med       Date:  2007-09-01       Impact factor: 17.440

10.  Evaluation of the VIA Blood Chemistry Monitor for Glucose in Healthy and Diabetic Volunteers.

Authors:  Arjunan Ganesh; Brian Hipszer; Navdeep Loomba; Barbara Simon; Marc C Torjman; Jeffrey Joseph
Journal:  J Diabetes Sci Technol       Date:  2008-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.