Literature DB >> 7869161

Internal fixation of the unstable anterior pelvic ring: a biomechanical comparison of standard plating techniques and the retrograde medullary superior pubic ramus screw.

P T Simonian1, M L Routt, R M Harrington, A F Tencer.   

Abstract

The purpose of this study was to evaluate pubic ramus fracture fixation. This biomechanical evaluation compared standard plating techniques with retrograde medullary screw fixation of a superior pubic ramus fracture in a pelvic fracture model. Six fresh-frozen, cadaveric pelvic specimens with a mean age of 79 years were harvested. These specimens were physiologically loaded according to the following modifications and instrumentations: (a) intact; (b) an APC-II unstable pelvic injury, specifically, unilateral superior and inferior rami osteotomies combined with ipsilateral anterior sacroiliac (SI) joint, sacrospinous, and sacrotuberous ligamentous disruptions, without fixation; (c) disrupted as in (b) but fixed anteriorly with a 10-hole 3.5-mm reconstruction plate contoured to the superior ramus and secured with four 3.5-mm cortical screws; (d) disrupted as in (b) but fixed anteriorly with a 10-hole 3.5-mm reconstruction plate contoured to the superior ramus and secured with six 3.5-mm cortical screws; (e) disrupted as in (b) but fixed anteriorly with a 4.5-mm retrograde medullary superior pubic ramus cortical screw 80 mm long (medial to the hip joint); and (f) disrupted as in (b) but fixed anteriorly with a 4.5-mm retrograde medullary superior pubic ramus cortical screw 130 mm long that was extraarticular and engaged the lateral iliac cortex cephalad to the ipsilateral hip joint. The posterior disruptions of the pelvic ring were not fixed. The APC-II injury created in this study resulted in significant (p < 0.05) motion at the disrupted rami and the injured SI joint, compared with the intact pelvic specimen.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1994        PMID: 7869161

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  18 in total

1.  Physiological in vitro sacroiliac joint motion: a study on three-dimensional posterior pelvic ring kinematics.

Authors:  Niels Hammer; Mario Scholze; Thomas Kibsgård; Stefan Klima; Stefan Schleifenbaum; Thomas Seidel; Michael Werner; Ronny Grunert
Journal:  J Anat       Date:  2018-12-09       Impact factor: 2.610

2.  Screw placement in percutaneous acetabular surgery: gender differences of anatomical landmarks in a cadaveric study.

Authors:  Thomas Dienstknecht; Michael Müller; Richard Sellei; Michael Nerlich; Franz Josef Müller; Bernd Fuechtmeier; Arne Berner
Journal:  Int Orthop       Date:  2012-12-19       Impact factor: 3.075

3.  Anterior internal fixator versus a femoral distractor and external fixation for sacroiliac joint compression and single stance gait testing: a mechanical study in synthetic bone.

Authors:  Jonathan M Vigdorchik; Amanda O Esquivel; Xin Jin; King H Yang; Rahul Vaidya
Journal:  Int Orthop       Date:  2013-05-04       Impact factor: 3.075

4.  Biomechanical analysis of anterior ring fixation of the ramus in type C pelvis fractures.

Authors:  S McLachlin; M Lesieur; D Stephen; H Kreder; C Whyne
Journal:  Eur J Trauma Emerg Surg       Date:  2017-04-08       Impact factor: 3.693

5.  [Percutaneous internal fixation of pelvic fractures. German version].

Authors:  A J Stevenson; B Swartman; A T Bucknill
Journal:  Unfallchirurg       Date:  2016-10       Impact factor: 1.000

Review 6.  Percutaneous internal fixation of pelvic fractures.

Authors:  A J Stevenson; B Swartman; A T Bucknill
Journal:  Unfallchirurg       Date:  2017-12       Impact factor: 1.000

7.  Preliminary results of cannulated screw fixation for isolated pubic ramus fractures.

Authors:  Jasper Winkelhagen; Michel P J van den Bekerom; Hugo W Bolhuis; Mike Hogervorst
Journal:  Strategies Trauma Limb Reconstr       Date:  2012-05-01

8.  Biomechanical stability of a supra-acetabular pedicle screw internal fixation device (INFIX) vs external fixation and plates for vertically unstable pelvic fractures.

Authors:  Jonathan M Vigdorchik; Amanda O Esquivel; Xin Jin; King H Yang; Ndidi A Onwudiwe; Rahul Vaidya
Journal:  J Orthop Surg Res       Date:  2012-09-27       Impact factor: 2.359

9.  Biomechanical Comparison of Five Fixation Techniques for Unstable Fragility Fractures of the Pelvic Ring.

Authors:  Moritz F Lodde; J Christoph Katthagen; Clemens O Schopper; Ivan Zderic; Geoff Richards; Boyko Gueorguiev; Michael J Raschke; René Hartensuer
Journal:  J Clin Med       Date:  2021-05-26       Impact factor: 4.241

10.  Experimental study on treatment of acetabular anterior column fractures: applyment of a minimally invasive percutaneous lag screw guide apparatus.

Authors:  Li-hai Zhang; Li-cheng Zhang; Qing-hua Si; Yuan Gao; Xiu-yun Su; Zhe Zhao; Pei-fu Tang
Journal:  BMC Musculoskelet Disord       Date:  2016-01-15       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.