PURPOSE: To evaluate the radiation risk to the operator and the patient during endovascular surgery of the head and neck.
METHODS: The dose was measured using thermoluminescence dosimeters attached at the body surface of the operator and the patient during 15 endovascular surgeries (3 for arteriovenous malformation, 8 for dural arteriovenous fistulas, and 4 for other disorders of the head and neck). The dose was measured at seven sites on the operator and at five sites on the patient.
RESULTS: The mean number of digital subtraction angiography studies and fluoroscopy time were 21 +/- 10 and 73 +/- 24 minutes, respectively. The equivalent dose range at each site in the operator was 0.12 to 0.88 mSv (glabella), 0.06 to 1.1 and 0 to 0.09 mSv (neck, outside and inside the protector, respectively), 0 to 0.20 mSv (left should, inside the protector), 0.09 to 1.99 mSv (left arm), 0.05 to 3.55 mSv (left hand), and 0 to 0.49 mSv (pubis, inside the protector). Those in the patients were 3.1 to 136 mSv (glabella), 13 to 5441 mSv (right temporal area), 4 to 186 mSv (left temporal area), 0.1 to 51 mSv (neck), and 0 to 0.62 mSv (pubis).
CONCLUSIONS: The total doses at the operator's eyes and left hand during the course of a year may exceed the dose limits recommended by the International Commission on Radiological Protection. Operators should wear not only body protectors, but also thyroid protectors and lead glass spectacles. The equivalent dose at the right temporal area of the patient may exceed the deterministic dose for transient erythema or alopecia of the skin even in one endovascular procedure.

Entities:  

Mesh:

Year:  1994        PMID: 7863926      PMCID: PMC8334275     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  10 in total

1.  Practical protective tools for occupational exposure: 1) double focus spectacles for the aged with highly refracted glass lens 2) remodeled barrier for radiation protection.

Authors:  S Kurokawa; S Yabe; A Takamura; H Ishizaki; S Aizawa
Journal:  Interv Neuroradiol       Date:  2001-05-15       Impact factor: 1.610

2.  The principle of digital subtraction angiography and radiological protection.

Authors:  K Okamoto; J Ito; K Sakai; S Yoshimura
Journal:  Interv Neuroradiol       Date:  2001-05-15       Impact factor: 1.610

Review 3.  Occupational radiation doses to operators performing fluoroscopically-guided procedures.

Authors:  Kwang Pyo Kim; Donald L Miller; Amy Berrington de Gonzalez; Stephen Balter; Ruth A Kleinerman; Evgenia Ostroumova; Steven L Simon; Martha S Linet
Journal:  Health Phys       Date:  2012-07       Impact factor: 1.316

4.  Reduction of radiation dose for cerebral angiography using flat panel detector of direct conversion type: a vascular phantom study.

Authors:  Y Hatakeyama; S Kakeda; N Ohnari; J Moriya; N Oda; K Nishino; W Miyamoto; Y Korogi
Journal:  AJNR Am J Neuroradiol       Date:  2007-04       Impact factor: 3.825

5.  Dose measurement on both patients and operators during neurointerventional procedures using photoluminescence glass dosimeters.

Authors:  T Moritake; Y Matsumaru; T Takigawa; K Nishizawa; A Matsumura; K Tsuboi
Journal:  AJNR Am J Neuroradiol       Date:  2008-08-21       Impact factor: 3.825

6.  Treatment of obstructive epiphora in adults by balloon dacryocystoplasty.

Authors:  Z Yazici; B Yazici; M Parlak; H Erturk; G Savci
Journal:  Br J Ophthalmol       Date:  1999-06       Impact factor: 4.638

7.  Radiation dermatitis after spinal arteriovenous malformation embolization: case report.

Authors:  G J Carstens; M B Horowitz; P D Purdy; A G Pandya
Journal:  Neuroradiology       Date:  1996-05       Impact factor: 2.804

8.  Patient skin dose during neuroembolization by multiple-point measurement using a radiosensitive indicator.

Authors:  S Suzuki; S Furui; Y Matsumaru; S Nobuyuki; M Ebara; T Abe; D Itoh
Journal:  AJNR Am J Neuroradiol       Date:  2008-04-03       Impact factor: 3.825

9.  Local skin and eye lens equivalent doses in interventional neuroradiology.

Authors:  Michael Sandborg; Sandro Rossitti; Håkan Pettersson
Journal:  Eur Radiol       Date:  2009-09-02       Impact factor: 5.315

Review 10.  Genetic and epigenetic mechanisms in the development of arteriovenous malformations in the brain.

Authors:  Jaya Mary Thomas; Sumi Surendran; Mathew Abraham; Arumugam Rajavelu; Chandrasekharan C Kartha
Journal:  Clin Epigenetics       Date:  2016-07-22       Impact factor: 6.551

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.