PURPOSE: To prospectively compare use of a phased-array multicoil and a conventional body coil in abdominal MR imaging.
MATERIALS AND METHODS: Thirteen patients (seven men, six women; mean age, 55 years) underwent imaging with a phased-array multicoil and with a conventional body coil. Four pulse sequences were used: T2-weighted spin echo (SE), magnetization-prepared gradient-recalled echo (GRE), breath-hold fast SE, and echo planar (EP).
RESULTS: Lesion detection improved the most on fast SE, multicoil-acquired images. Signal-to-noise ratio (S/N) increased 64% with fast SE (P = .0005) and EP (P < .0109) sequences. Contrast-to-noise ratio (C/N) doubled (P < .05) with T2-weighted SE sequences. Lesion conspicuity improved on multicoil-acquired images with all fast sequences (magnetization-prepared GRE, P = .015; fast SE, P = .002; EP imaging, P = .013). There was little difference in respiratory and vascular artifact. Depiction of most abdominal structures improved (P < .01).
CONCLUSION: Use of the phased-array multicoil provides better MR images of the abdomen than does use of a conventional body coil.

Entities:  

Mesh:

Year:  1995        PMID: 7754009     DOI: 10.1148/radiology.195.3.7754009

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

1.  Determination of appropriate RF blocking impedance for MRI surface coils and arrays.

Authors:  A Kocharian; P J Rossman; T C Hulshizer; J P Felmlee; S J Riederer
Journal:  MAGMA       Date:  2000-06       Impact factor: 2.310

2.  T2-weighted breathold imaging of the liver: a quantitative and qualitative comparison of fast spin echo and half Fourier single shot fast spin echo imaging.

Authors:  T K Helmberger; J Schröder; N Holzknecht; M Gregor; A Heuck; R Petsch; M F Reiser
Journal:  MAGMA       Date:  1999-10       Impact factor: 2.310

3.  High-resolution MRI of pancreatic masses with a new circularly polarized body phased-array coil.

Authors:  K Engelhard; H P Hollenbach
Journal:  Eur Radiol       Date:  1997       Impact factor: 5.315

4.  Signal-to-noise ratio and parallel imaging performance of commercially available phased array coils in 3.0 T brain magnetic resonance imaging.

Authors:  Tsukasa Yoshida; Kensei Shirata; Atsushi Urikura; Michitoshi Ito; Yoshihiro Nakaya
Journal:  Radiol Phys Technol       Date:  2015-05-26

5.  MRI of gastric carcinoma: results of T and N-staging in an in vitro study.

Authors:  Il Young Kim; Sang Won Kim; Hyeong Cheol Shin; Moon Soo Lee; Dong Jun Jeong; Chang Jin Kim; Young Tong Kim
Journal:  World J Gastroenterol       Date:  2009-08-28       Impact factor: 5.742

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.