Literature DB >> 7744901

Failure of the Mecring screw-ring acetabular component in total hip arthroplasty. A three to seven-year follow-up study.

J D Bruijn1, J L Seelen, R M Feenstra, B E Hansen, F P Bernoski.   

Abstract

We prospectively studied the results of 411 consecutive total hip arthroplasties with a Mecring screw-ring acetabular component inserted without cement combined with a Stanmore femoral stem inserted with cement. The duration of follow-up ranged from three to seven years (mean, four years and six months). Three hundred and thirty-one patients (378 hips) were available for physical examination and had a complete set of radiographs. The clinical result was good or excellent for 82 per cent (309) of the 378 hips. However, the rate of radiographic loosening of the acetabular component, as evidenced by migration at the most recent follow-up examination, was alarmingly high: 25 per cent (ninety-five) of the 378 hips. In general, these patients did not have serious clinical symptoms. The cups in women migrated significantly more often (p = 0.003) than those in men. Migration was also more frequent in patients who were less than fifty-one years old and in patients in whom the index procedure was a revision arthroplasty, but these differences were not significant. Twenty-one (6 per cent) of the acetabular cups were revised for aseptic loosening. The high rate of radiographic loosening has led us to abandon the use of the Mecring screw-ring acetabular component.

Entities:  

Mesh:

Year:  1995        PMID: 7744901     DOI: 10.2106/00004623-199505000-00012

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  7 in total

1.  Early polyethylene wear and osteolysis with ABG acetabular cups (7- to 12-year follow-up).

Authors:  Sachin Badhe; Peter Livesley
Journal:  Int Orthop       Date:  2005-11-08       Impact factor: 3.075

2.  Modes of failure of a threaded acetabular cup : a radiographic study with EBRA of 42 revised cups.

Authors:  Thomas Ilchmann; Severin Neher; Franz Maurer; Kuno Weise
Journal:  Int Orthop       Date:  2006-06-08       Impact factor: 3.075

3.  [Five-year survival rate of the Allofit titanium press-fit cup].

Authors:  K Schroeder; A Moehlenbruch; M Zimmermann-Stenzel; D Parsch
Journal:  Orthopade       Date:  2010-01       Impact factor: 1.087

4.  Long-term results of the threaded Mecron cup in primary total hip arthroplasty : A 15-20-year follow-up study.

Authors:  Michael Clarius; Alexander W Jung; Marcus R Streit; Christian Merle; Patric Raiss; Peter R Aldinger
Journal:  Int Orthop       Date:  2009-07-24       Impact factor: 3.075

5.  Long-term results of the threaded Weill cup in primary total hip arthroplasty: a 15-20-year follow-up study.

Authors:  Michael Clarius; Alexander W Jung; Patric Raiss; Marcus R Streit; Christian Merle; Peter R Aldinger
Journal:  Int Orthop       Date:  2009-07-24       Impact factor: 3.075

6.  A comparative study of "plasmacup" and "porous-coated" acetabular components: survival after 10 to 12 years of follow-up.

Authors:  José Ricardo Negreiros Vicente; Carlos Antonio Soares Ulhoa; Marcio Katz; Renato Dainesi Addeo; Alberto Tesconi Croci
Journal:  Clinics (Sao Paulo)       Date:  2010       Impact factor: 2.365

7.  Failure of Polyethylene Inlays in Cementless Total Hip Arthroplasty: A Retrieval Analysis.

Authors:  Ulrike Mueller; Christoph Lee; Christian Heisel; Marc Thomsen; Rudi G Bitsch; J Philippe Kretzer
Journal:  Biomed Res Int       Date:  2016-08-31       Impact factor: 3.411

  7 in total