Accuracy of laboratory and portable monitor international normalized ratio determinations. Comparison with a criterion standard.

S S Kaatz1, R H White, J Hill, E Mascha, J E Humphries, D M Becker.   

Abstract

BACKGROUND: Portable instruments that measure the prothrombin time and automatically calculate the international normalized ratio (INR) with the use of a drop of whole blood have simplified the treatment of patients who are receiving warfarin therapy. The accuracy of these portable monitors has never been determined by comparing INR results with a criterion (gold) standard INR determination.
METHODS: Duplicate whole-blood INR determinations were made with two commercially available portable INR monitors. Duplicate frozen-plasma samples were measured with four different thromboplastin reagents, each with a different international sensitivity index. The criterion standard INR was determined by using an international reference thromboplastin and the manual tilt-tube technique. Agreement was evaluated by determining how accurately laboratory and monitor INR determinations matched criterion standard values in designating a sample to be within or outside of currently recommended INR target ranges.
RESULTS: Two of the laboratory methods, which used relatively sensitive thromboplastins, showed close agreement with the criterion standard, whereas two laboratory methods that used less sensitive thromboplastin reagents showed poor agreement. Both of the portable monitors fell between these extremes. The two best laboratory methods ere significantly better (P < .003) than both monitors, which in turn were better (P < .003) than the remaining two laboratories.
CONCLUSIONS: There is large interlaboratory variation in the accuracy of INR determinations. Laboratory methods that used insensitive (high international sensitivity index) thromboplastins performed poorly. Accuracy of monitor measurements appears satisfactory.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7677552

Source DB:  PubMed          Journal:  Arch Intern Med        ISSN: 0003-9926


  15 in total

Review 1.  The role of point-of-care anticoagulation monitoring in arterial and venous thromboembolic disorders.

Authors:  C R Zimmerman
Journal:  J Thromb Thrombolysis       Date:  2000-04       Impact factor: 2.300

2.  Accuracy and clinical utility of the CoaguChek XS portable international normalised ratio monitor in a pilot study of warfarin home-monitoring.

Authors:  Luke R Bereznicki; Shane L Jackson; Gregory M Peterson; Ella C Jeffrey; Katherine A Marsden; David M Jupe
Journal:  J Clin Pathol       Date:  2006-06-02       Impact factor: 3.411

3.  Validity of CoaguChek S for home monitoring of anticoagulant therapy in pediatrics.

Authors:  James Hill; Sébastien Perreault; Michel Dorval
Journal:  Can J Cardiol       Date:  2007-01       Impact factor: 5.223

Review 4.  Oral anticoagulant therapy: Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines.

Authors:  Walter Ageno; Alexander S Gallus; Ann Wittkowsky; Mark Crowther; Elaine M Hylek; Gualtiero Palareti
Journal:  Chest       Date:  2012-02       Impact factor: 9.410

5.  Warfarin monitoring: point-of-care INR testing limitations and interpretation of the prothrombin time.

Authors:  Alan K Jacobson
Journal:  J Thromb Thrombolysis       Date:  2007-11-15       Impact factor: 2.300

Review 6.  Accuracy, precision, and quality control for point-of-care testing of oral anticoagulation.

Authors:  A M van den Besselaar
Journal:  J Thromb Thrombolysis       Date:  2001-09       Impact factor: 2.300

7.  Reliability of international normalised ratios from two point of care test systems: comparison with conventional methods.

Authors:  Leon Poller; Michelle Keown; Nikhil Chauhan; Anton M H P van den Besselaar; Armando Tripodi; Caroline Shiach; Jorgen Jespersen
Journal:  BMJ       Date:  2003-07-05

8.  Reliability, validity and ease of use of a portable point-of-care coagulation device in a pharmacist-managed anticoagulation clinic.

Authors:  Annie Lizotte; Isabelle Quessy; Marie-Claude Vanier; Josée Martineau; Stéphanie Caron; Martin Darveau; Alain Dubé; Edith Gilbert; Normand Blais; Lyne Lalonde
Journal:  J Thromb Thrombolysis       Date:  2002-12       Impact factor: 2.300

Review 9.  Home monitoring of anticoagulation.

Authors:  Julie Hambleton
Journal:  J Thromb Thrombolysis       Date:  2003 Aug-Oct       Impact factor: 2.300

10.  A randomized trial comparing INR monitoring devices in patients with anticoagulation self-management: evaluation of a novel error-grid approach.

Authors:  Lars G Hemkens; Kristian M Hilden; Stephan Hartschen; Thomas Kaiser; Ulrike Didjurgeit; Roland Hansen; Ralf Bender; Peter T Sawicki
Journal:  J Thromb Thrombolysis       Date:  2007-10-26       Impact factor: 2.300

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.