Truncation artifact reduction in transmission CT for improved SPECT attenuation compensation.

D J Kadrmas1, R J Jaszczak, J W McCormick, R E Coleman, C B Lim.   

Abstract

Transmission computed tomography (TCT) has been shown to be an accurate method of acquiring non-uniform attenuation maps for single-photon emission computed tomography (SPECT) attenuation compensation. One commonly encountered problem, especially for convergent beam geometries, is image truncation. We describe two methods for reducing associated truncation artifacts, with the goal of improving SPECT attenuation compensation without unduly increasing imaging or reconstruction times: (i) the two-scan method, which reduces the degree of truncation by combining two short-duration, patient-shifted scans, and (ii) a quantitative extrapolation method, which fills in truncated projections accounting for the correct amount of attenuating medium in the slice. The methods are evaluated by imaging two phantoms on a fan beam TCT system. Projection sets are truncated to various degrees with software, and reconstructed images are compared to both untruncated filtered backprojection and iterative reconstructions of truncated data. A fundamental analysis of SPECT attenuation factors is performed, and attenuation compensation of a cardiac insert is analysed. Results indicate the two-scan method can effectively reduce the degree of truncation in many cases, and the quantitative extrapolation method greatly improves SPECT attenuation compensation over using truncated maps.

Entities:  

Mesh:

Year:  1995        PMID: 7659732     DOI: 10.1088/0031-9155/40/6/009

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  C-SPECT - a Clinical Cardiac SPECT/Tct Platform: Design Concepts and Performance Potential.

Authors:  Wei Chang; Caesar E Ordonez; Haoning Liang; Yusheng Li; Jingai Liu
Journal:  IEEE Trans Nucl Sci       Date:  2009-10-06       Impact factor: 1.679

2.  Reconstruction from truncated projections in CT using adaptive detruncation.

Authors:  K Sourbelle; M Kachelriess; W A Kalender
Journal:  Eur Radiol       Date:  2005-02-09       Impact factor: 5.315

3.  Effects of airway tree asymmetry on the emergence and spatial persistence of ventilation defects.

Authors:  D Leary; T Winkler; A Braune; G N Maksym
Journal:  J Appl Physiol (1985)       Date:  2014-06-19

4.  Enhanced accuracy of defect detection by myocardial single-photon emission computed tomography with attenuation correction with gadolinium 153 line sources: evaluation with a cardiac phantom.

Authors:  Z X He; M D Scarlett; J J Mahmarian; M S Verani
Journal:  J Nucl Cardiol       Date:  1997 May-Jun       Impact factor: 5.952

5.  A diffusion-based truncated projection artifact reduction method for iterative digital breast tomosynthesis reconstruction.

Authors:  Yao Lu; Heang-Ping Chan; Jun Wei; Lubomir M Hadjiiski
Journal:  Phys Med Biol       Date:  2013-01-14       Impact factor: 3.609

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.