Singles transmission in volume-imaging PET with a 137Cs source.

J S Karp1, G Muehllehner, H Qu, X H Yan.   

Abstract

The feasibility of a new method of attenuation correction in PET has been investigated, using a single-photon emitter for the transmission scan. The transmission scan is predicted to be more than a factor of ten faster with the singles method than the standard coincidence method, for comparable statistics. Thus, a transmission scan be completed in 1-2 min, rather than 10-20 min, as is common practice with the coincidence method. In addition, a potential advantage of using the single-photon source 137Cs, which has an energy of 662 keV, is that postinjection transmission studies can be performed using energy discrimination to separate the transmission from the emission data at 511 keV. In order to compensate for the energy difference of the attenuation coefficients at 662 keV compared to 511 keV, the transmission images are segmented into two compartments, tissue and lung, and known values (for 511 keV) of attenuation are inserted into these compartments. This technique also compensates for the higher amount of scatter present with the singles method, since it is not possible to use a position gate (based on collinearity of the source and two detector positions) as is commonly done with a positron-emitting source. We have demonstrated, with experimental phantom studies, that the singles transmission method combined with segmentation gives results equivalent both qualitatively and quantitatively to the coincidence method, but requires significantly less time.

Mesh:

Substances:

Year:  1995        PMID: 7652016     DOI: 10.1088/0031-9155/40/5/014

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  11 in total

1.  Attenuation correction in emission tomography using the emission data--A review.

Authors:  Yannick Berker; Yusheng Li
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

2.  Performance evaluation of the Philips MOSAIC small animal PET scanner.

Authors:  Marc C Huisman; Sybille Reder; Axel W Weber; Sibylle I Ziegler; Markus Schwaiger
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-11-22       Impact factor: 9.236

3.  Attenuation correction for small animal PET tomographs.

Authors:  Patrick L Chow; Fernando R Rannou; Arion F Chatziioannou
Journal:  Phys Med Biol       Date:  2005-04-06       Impact factor: 3.609

Review 4.  Oncological applications of positron emission tomography with fluorine-18 fluorodeoxyglucose.

Authors:  P Rigo; P Paulus; B J Kaschten; R Hustinx; T Bury; G Jerusalem; T Benoit; J Foidart-Willems
Journal:  Eur J Nucl Med       Date:  1996-12

5.  Comparison of list-mode and DIRECT approaches for time-of-flight PET reconstruction.

Authors:  Margaret E Daube-Witherspoon; Samuel Matej; Matthew E Werner; Suleman Surti; Joel S Karp
Journal:  IEEE Trans Med Imaging       Date:  2012-03-06       Impact factor: 10.048

6.  137Cs transmission imaging and segmented attenuation corrections in a small animal PET scanner.

Authors:  Ying-Hwey Nai; Takayuki Ose; Miho Shidahara; Hiroshi Watabe
Journal:  Radiol Phys Technol       Date:  2017-07-08

7.  Improvement of attenuation correction in time-of-flight PET/MR imaging with a positron-emitting source.

Authors:  Pieter Mollet; Vincent Keereman; Jason Bini; David Izquierdo-Garcia; Zahi A Fayad; Stefaan Vandenberghe
Journal:  J Nucl Med       Date:  2014-01-16       Impact factor: 10.057

8.  Rapid normalization of osseous FDG uptake following traumatic or surgical fractures.

Authors:  Hongming Zhuang; Joseph W Sam; Thomas K Chacko; Paulo S Duarte; Marc Hickeson; Qi Feng; Kozaim Z Nakhoda; Liang Guan; Phillip Reich; Shirley M Altimari; Abass Alavi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-05-22       Impact factor: 9.236

9.  Importance of Attenuation Correction (AC) for Small Animal PET Imaging.

Authors:  Henrik H El Ali; Rasmus Poul Bodholdt; Jesper Tranekjær Jørgensen; Rebecca Myschetzky; Andreas Kjaer
Journal:  Diagnostics (Basel)       Date:  2012-10-09

10.  Attenuation correction for small animal PET images: a comparison of two methods.

Authors:  Daniela D'Ambrosio; Federico Zagni; Antonello E Spinelli; Mario Marengo
Journal:  Comput Math Methods Med       Date:  2013-04-16       Impact factor: 2.238

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.