Limitations of forehead pulse oximetry.

J S Jørgensen1, E R Schmid, V König, K Faisst, A Huch, R Huch.   

Abstract

During initial clinical tests to calibrate our reflectance pulse oximetry system, we observed serious physiologic limitations to the use of pulse oximetry in the forehead region. We present a case of simultaneous reflectance and transmission mode pulse oximetry monitoring in a child undergoing cardiac surgery for congenital cyanotic heart disease with a large intracardiac shunt. During general anesthesia, when the patient was endotracheally intubated and mechanically ventilated, the transmission mode saturation agreed well with arterial oxygen saturation measurements; but, our reflectance pulse oximeter, with the sensor applied to the forehead, displayed spuriously lower (-18%) oxygen saturations. Before and after anesthesia and surgery, there was fine agreement between reflectance and transmission mode saturation values. We suggest that the difference was caused by vasodilatation and pooling of venous blood due to compromised venous return to the heart, and a combination of arterial and venous pulsations in the forehead region. This means that the reflectance pulse oximeter measured a mixed arterial-venous oxygen saturation.

Entities:  

Mesh:

Year:  1995        PMID: 7561999     DOI: 10.1007/BF01617520

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  11 in total

Review 1.  Pulse oximetry.

Authors:  K K Tremper; S J Barker
Journal:  Anesthesiology       Date:  1989-01       Impact factor: 7.892

Review 2.  Recent developments in pulse oximetry.

Authors:  J W Severinghaus; J F Kelleher
Journal:  Anesthesiology       Date:  1992-06       Impact factor: 7.892

3.  Normal and respiratory variations of the hepatic and portal venous duplex Doppler waveforms with simultaneous electrocardiographic correlation.

Authors:  M M Abu-Yousef
Journal:  J Ultrasound Med       Date:  1992-06       Impact factor: 2.153

4.  Adaptation of pulse oximetry for fetal monitoring during labour.

Authors:  J O Gardosi; C M Schram; E M Symonds
Journal:  Lancet       Date:  1991-05-25       Impact factor: 79.321

5.  Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry.

Authors:  H M Sami; B S Kleinman; V A Lonchyna
Journal:  J Clin Monit       Date:  1991-10

6.  Systolic venous waves cause spurious signs of arterial hemoglobin desaturation.

Authors:  J B Mark
Journal:  Anesthesiology       Date:  1989-07       Impact factor: 7.892

7.  Pulse oximetry and circulatory kinetics associated with pulse volume amplitude measured by photoelectric plethysmography.

Authors:  J M Kim; K Arakawa; K T Benson; D K Fox
Journal:  Anesth Analg       Date:  1986-12       Impact factor: 5.108

8.  Falsely normal saturation reading with the pulse oximeter.

Authors:  A T Costarino; D A Davis; T P Keon
Journal:  Anesthesiology       Date:  1987-11       Impact factor: 7.892

9.  An evaluation of the effect of vasodilatation on oxygen saturations measured by pulse oximetry and venous blood gas analysis.

Authors:  I J Broome; G H Mills; P Spiers; C S Reilly
Journal:  Anaesthesia       Date:  1993-05       Impact factor: 6.955

10.  Experimental and clinical evaluation of a noninvasive reflectance pulse oximeter sensor.

Authors:  S Takatani; C Davies; N Sakakibara; A Zurick; E Kraenzler; L R Golding; G P Noon; Y Nose; M E DeBakey
Journal:  J Clin Monit       Date:  1992-10
View more
  4 in total

1.  The influence of larger subcutaneous blood vessels on pulse oximetry.

Authors:  Paul D Mannheimer; Neil Michael P O'; Ewald Konecny
Journal:  J Clin Monit Comput       Date:  2004-06       Impact factor: 2.502

2.  Forehead reflectance oximetry: a clinical comparison with conventional digit sensors during laparotomic and laparoscopic abdominal surgery.

Authors:  Andrea Casati; Grazia Squicciarini; Marco Baciarello; Marta Putzu; Alessandra Salvadori; Guido Fanelli
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

3.  Effects of passive smoking on students at College of Applied Medical Sciences, King Saud Bin Abdulaziz University for Health Sciences, Riyadh.

Authors:  Abdullah Alanazi; Farhan Al Enezi; Mohammd Mesfer Alqahtani; Turki Faleh Alshammari; Mumtaz Ahmed Ansari; Saleh Al-Oraibi; Shoeb Qureshi
Journal:  J Nat Sci Biol Med       Date:  2015 Jan-Jun

Review 4.  Application of Modern Multi-Sensor Holter in Diagnosis and Treatment.

Authors:  Erik Vavrinsky; Jan Subjak; Martin Donoval; Alexandra Wagner; Tomas Zavodnik; Helena Svobodova
Journal:  Sensors (Basel)       Date:  2020-05-07       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.