An automatic and accurate x-ray tube focal spot/grid alignment system for mobile radiography: system description and alignment accuracy.

David M Gauntt1, Gary T Barnes.   

Abstract

PURPOSE: A mobile radiography automatic grid alignment system (AGAS) has been developed by modifying a commercially available mobile unit. The objectives of this article are to describe the modifications and operation and to report on the accuracy with which the focal spot is aligned to the grid and the time required to achieve the alignment.
METHODS: The modifications include an optical target arm attached to the grid tunnel, a video camera attached to the collimator, a motion control system with six degrees of freedom to position the collimator and x-ray tube, and a computer to control the system. The video camera and computer determine the grid position, and then the motion control system drives the x-ray focal spot to the center of the grid focal axis. The accuracy of the alignment of the focal spot with the grid and the time required to achieve alignment were measured both in laboratory tests and in clinical use.
RESULTS: For a typical exam, the modified unit automatically aligns the focal spot with the grid in less than 10 s, with an accuracy of better than 4 mm. The results of the speed and accuracy tests in clinical use were similar to the results in laboratory tests. Comparison patient chest images are presented--one obtained with a standard mobile radiographic unit without a grid and the other obtained with the modified unit and a 15:1 grid. The 15:1 grid images demonstrate a marked improvement in image quality compared to the nongrid images with no increase in patient dose.
CONCLUSIONS: The mobile radiography AGAS produces images of significantly improved quality compared to nongrid images with alignment times of less than 10 s and no increase in patient dose.

Entities:  

Mesh:

Year:  2010        PMID: 21302797      PMCID: PMC3003722          DOI: 10.1118/1.3518085

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Scatter correction in digital radiography using interpolated local sampling of aperture signals.

Authors:  K P Maher
Journal:  Australas Phys Eng Sci Med       Date:  2001-06       Impact factor: 1.430

2.  X-ray scattering in full-field digital mammography.

Authors:  Kirsi Nykänen; Samuli Siltanen
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

3.  Device for facilitating precise alignment in bedside radiography.

Authors:  P B O'Donovan; G J Skipper; J C Litchney; A J Salupo; J R Bortnick
Journal:  Radiology       Date:  1992-07       Impact factor: 11.105

4.  Laser alignment system for high-quality portable radiography.

Authors:  H MacMahon; N J Yasillo; M Carlin
Journal:  Radiographics       Date:  1992-01       Impact factor: 5.333

5.  Dual-energy x-ray projection imaging: two sampling schemes for the correction of scattered radiation.

Authors:  F C Wagner; A Macovski; D G Nishimura
Journal:  Med Phys       Date:  1988 Sep-Oct       Impact factor: 4.071

6.  Scatter correction algorithm for digitally acquired radiographs: theory and results.

Authors:  J M Boone
Journal:  Med Phys       Date:  1986 May-Jun       Impact factor: 4.071

7.  A regional convolution kernel algorithm for scatter correction in dual-energy images: comparison to single-kernel algorithms.

Authors:  D G Kruger; F Zink; W W Peppler; D L Ergun; C A Mistretta
Journal:  Med Phys       Date:  1994-02       Impact factor: 4.071

8.  New method for the experimental evaluation of x-ray grids.

Authors:  C E Dick; J W Motz
Journal:  Med Phys       Date:  1978 Mar-Apr       Impact factor: 4.071

9.  Effect of reduced scatter on radiographic information content and patient exposure: a quantitative demonstration.

Authors:  R F Wagner; G T Barnes; B S Askins
Journal:  Med Phys       Date:  1980 Jan-Feb       Impact factor: 4.071

10.  Grid removal and impact on population dose in full-field digital mammography.

Authors:  Gisella Gennaro; Luc Katz; Henri Souchay; Remy Klausz; Claudio Alberelli; Cosimo di Maggio
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

View more
  3 in total

1.  Development of a mini-mobile digital radiography system by using wireless smart devices.

Authors:  Chang-Won Jeong; Su-Chong Joo; Jong-Hyun Ryu; Jinseok Lee; Kyong-Woo Kim; Kwon-Ha Yoon
Journal:  J Digit Imaging       Date:  2014-08       Impact factor: 4.056

2.  The Potential Role of Grid-Like Software in Bedside Chest Radiography in Improving Image Quality and Dose Reduction: An Observer Preference Study.

Authors:  Su Yeon Ahn; Kum Ju Chae; Jin Mo Goo
Journal:  Korean J Radiol       Date:  2018-04-06       Impact factor: 3.500

3.  Air gap technique is recommended in axiolateral hip radiographs.

Authors:  Susanne Kivistö; Antti Kotiaho; Anja Henner; Terhi Nevala; Jaakko Niinimäki; Miika T Nieminen; Matti Hanni
Journal:  J Appl Clin Med Phys       Date:  2020-09-21       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.