Computerized determination of pneumotachometer characteristics using a calibrated syringe.

M P Yeh, R M Gardner, T D Adams, F G Yanowitz.   

Abstract

A computerized method has been developed to determine the conductance characteristics of pneumotachometers. Conductance values of the flowmeter, which correspond to all pressure values, can be determined by a weighted averaging technique, when multiple strokes of a precision 3-liter calibrated syringe are used. The conductance values then allow the measured differential pressures to be converted, point by point, into flows. The accuracy of measured volumes is within +/- 0.5% for a 100-stroke calibration process, and calibration with 50 strokes provides +/- 1% accuracy. The method improves the accuracy of the on-line measurements of ventilation. O2 consumption, and CO2 production during exercise.

Entities:  

Mesh:

Year:  1982        PMID: 7118642     DOI: 10.1152/jappl.1982.53.1.280

Source DB:  PubMed          Journal:  J Appl Physiol Respir Environ Exerc Physiol        ISSN: 0161-7567


  25 in total

1.  Error analysis of a natural breathing calibration method for respiratory inductive plethysmography.

Authors:  N O Strömberg
Journal:  Med Biol Eng Comput       Date:  2001-05       Impact factor: 2.602

2.  Effects of respiratory muscle work on blood flow distribution during exercise in heart failure.

Authors:  Thomas P Olson; Michael J Joyner; Niki M Dietz; John H Eisenach; Timothy B Curry; Bruce D Johnson
Journal:  J Physiol       Date:  2010-05-10       Impact factor: 5.182

3.  Accuracy and reliability of the ParvoMedics TrueOne 2400 and MedGraphics VO2000 metabolic systems.

Authors:  Scott E Crouter; Amanda Antczak; Jonathan R Hudak; Diane M DellaValle; Jere D Haas
Journal:  Eur J Appl Physiol       Date:  2006-08-03       Impact factor: 3.078

4.  H1 receptor-mediated vasodilatation contributes to postexercise hypotension.

Authors:  Jennifer M Lockwood; Brad W Wilkins; John R Halliwill
Journal:  J Physiol       Date:  2004-12-23       Impact factor: 5.182

5.  Fluid replacement and heat stress during exercise alter post-exercise cardiac haemodynamics in endurance exercise-trained men.

Authors:  Brenna M Lynn; Christopher T Minson; John R Halliwill
Journal:  J Physiol       Date:  2009-06-02       Impact factor: 5.182

6.  Cardiorespiratory response to exercise on a large therapeutic roll.

Authors:  Eduard Gappmaier; Sima F Tavazoie; Michael G Jacketta
Journal:  Cardiopulm Phys Ther J       Date:  2013-09

7.  Calibration of Fleisch and screen pneumotachographs for use with various oxygen concentrations.

Authors:  M J Turner; I M MacLeod; A D Rothberg
Journal:  Med Biol Eng Comput       Date:  1990-03       Impact factor: 2.602

8.  Cardiovascular effects of anti-G suit inflation at 1 and 2 G.

Authors:  Stéphanie Montmerle; Dag Linnarsson
Journal:  Eur J Appl Physiol       Date:  2005-04-07       Impact factor: 3.078

9.  An indirect calorimetry system for ventilator dependent very low birthweight infants.

Authors:  J S Forsyth; A Crighton
Journal:  Arch Dis Child       Date:  1992-03       Impact factor: 3.791

10.  Transient, steady-state and rebreathing responses to carbon dioxide in man, at rest and during light exercise.

Authors:  M S Jacobi; C P Patil; K B Saunders
Journal:  J Physiol       Date:  1989-04       Impact factor: 5.182

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.