L G Carlson, C J Delaney, J J Plorde.
Abstract
We report two potential liabilities of the homogeneous enzyme immunoassay for the determination of serum gentamicin. One is a considerable loss of precision and accuracy at the ends of the calibration curve, and the other is an apparent loss of gentamicin with storage at -60 degrees C.Entities:
Mesh:
Substances:
Year: 1982 PMID: 7044298 PMCID: PMC181853 DOI: 10.1128/AAC.21.1.192
Source DB: PubMed Journal: Antimicrob Agents Chemother ISSN: 0066-4804 Impact factor: 5.191