| Literature DB >> 6994639 |
T D O'Leary, R M Ratcliff, T D Geary.
Abstract
The EMIT (Enzyme Multiplied Immunoassay Technique) for serum gentamicin determination was evaluated by a standard procedure. The precision, accuracy, and specificity were assessed and proved satisfactory. Comparison with a bioassay was done, and results for patient samples showed a correlation coefficient of 0.95 between the two methods. Advantages of the enzyme immunoassay system were the provision of results within 15 min of receipt of blood in the laboratory, a requirement of minimal technical expertise, and an applicability to both large and small workloads. The EMIT proved during evaluation to be a practical alternative to current bioassays for the determination of serum gentamicin concentrations.Entities:
Mesh:
Substances:
Year: 1980 PMID: 6994639 PMCID: PMC283874 DOI: 10.1128/AAC.17.5.776
Source DB: PubMed Journal: Antimicrob Agents Chemother ISSN: 0066-4804 Impact factor: 5.191