| Literature DB >> 6425001 |
M Dudley, W F Nauschuetz, S V Juchau.
Abstract
The AutoMicrobic System GSC Plus Card was compared to a broth microdilution reference method. Of 1660 minimum inhibitory concentration pairs, 86.1% were within 1 log2 dilution range. A total of 89.2% of the GSC Plus results agreed qualitatively (susceptible, moderately susceptible, resistant). If chloramphenicol data are excluded, agreement is 88.8% and 89.9%, respectively. A total of only 0.90% results were very major or major discrepancies. This is significantly fewer major discrepancies than reported with other systems.Entities:
Mesh:
Substances:
Year: 1983 PMID: 6425001 DOI: 10.1016/0732-8893(83)90043-3