Measurements of dose from secondary radiation outside a treatment field.

K R Kase, G K Svensson, A B Wolbarst, M A Marks.   

Abstract

Radiation dose to organs outside the radiotherapy treatment field can be significant and therefore is of clinical interest. We have made measurements of dose at distances up to 70 cm from the central axes of 5 X 5, 15 X 15 and 25 X 25 cm radiation fields of 300 kVp, 4 MV and 8 MV X rays, and 60Co gamma rays, at the surface and at depths in water of 5 and 10 cm. Contributions to the total secondary radiation dose from water scatter, machine (collimator) scatter and leakage radiation have been separated. We have found that the component of dose from water scatter can be described by a simple exponential function of distance from the central axis of the radiation field for all energies and field sizes. Machine scatter contributes 20 to 40% of the total secondary dose depending on machine, field size and distance from the field. Leakage radiation contributes very little dose, but becomes the dominant component at distances beyond 60 cm from the central axis. Estimates of the risk of second tumors in long term survivors indicate a small incremental increase above the natural incidence rate based on information from the 1980 BEIR Committee report.

Entities:  

Mesh:

Substances:

Year:  1983        PMID: 6409854     DOI: 10.1016/0360-3016(83)90177-3

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  27 in total

Review 1.  Exposing the thyroid to radiation: a review of its current extent, risks, and implications.

Authors:  Bridget Sinnott; Elaine Ron; Arthur B Schneider
Journal:  Endocr Rev       Date:  2010-07-21       Impact factor: 19.871

2.  Real-time dosimetry in external beam radiation therapy.

Authors:  Ramachandran Prabhakar
Journal:  World J Radiol       Date:  2013-10-28

3.  Variations in photon energy spectra of a 6 MV beam and their impact on TLD response.

Authors:  Sarah B Scarboro; David S Followill; Rebecca M Howell; Stephen F Kry
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

4.  Excessive applicator radiation leakage for a common therapeutic kilovoltage system.

Authors:  Brad Beeksma; Joerg Lehmann
Journal:  Br J Radiol       Date:  2018-11-15       Impact factor: 3.039

5.  Technical Report: Evaluation of peripheral dose for flattening filter free photon beams.

Authors:  E L Covington; T A Ritter; J M Moran; A M Owrangi; J I Prisciandaro
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

6.  [The gonadal loading during the irradiation of lymph outflow in operated seminomas. In-vivo dosimetry].

Authors:  G Gruber; N Schwegler
Journal:  Strahlenther Onkol       Date:  1999-04       Impact factor: 3.621

7.  Analytical model for out-of-field dose in photon craniospinal irradiation.

Authors:  Phillip J Taddei; Wassim Jalbout; Rebecca M Howell; Nabil Khater; Fady Geara; Kenneth Homann; Wayne D Newhauser
Journal:  Phys Med Biol       Date:  2013-10-08       Impact factor: 3.609

Review 8.  A review of dosimetry studies on external-beam radiation treatment with respect to second cancer induction.

Authors:  X George Xu; Bryan Bednarz; Harald Paganetti
Journal:  Phys Med Biol       Date:  2008-06-09       Impact factor: 3.609

9.  Use of new radiochromic devices for peripheral dose measurement: potential in-vivo dosimetry application.

Authors:  S-T Chiu-Tsao; Mf Chan
Journal:  Biomed Imaging Interv J       Date:  2009-10-01

10.  Peripheral dose measurements with diode and thermoluminescence dosimeters for intensity modulated radiotherapy delivered with conventional and un-conventional linear accelerator.

Authors:  Rajesh Kinhikar; Poonam Gamre; Chandrashekhar Tambe; Sudarshan Kadam; George Biju; C S Magai; Dipak Dhote; Shyam Shrivastava; Deepak Deshpande
Journal:  J Med Phys       Date:  2013-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.