A fast and reliable system for microcomputerized DNA cytofluorometry in tumour pathology.

K Bjelkenkrantz, O Stål, O Gröntoft.   

Abstract

A microcomputerized cytofluorometry system based on a Leitz MPV 3 cytophotometer, and intended for DNA measurements in tumour pathology is described. The system has been equipped with a reference channel for correction of excitation light instability. The importance of the adjustment of the epi-illumination for optimal performance of the reference channel is stressed, and a detailed description is provided. An apparatus variability well below 1% CV is obtained even during periods of marked instability of the arc-lamp. Software for conventional cytofluorometry with which about 200 cells can be measured in 20 min is presented. In addition a measuring technique where cells are not positioned, but are just passed through the excitation light spot is described. Preliminary results on cytocentrifuged specimens of human cancers indicate that this system is at least three times faster than conventional fluorometry. The increased speed of measurements considerably extends the possibilities to evaluate cell proliferation using static cytofluorometry. The measuring capacity of course is dependent upon the quality of the specimen. In well prepared cytocentrifuged specimens 1000 cells have been measured in 35 min.

Entities:  

Mesh:

Substances:

Year:  1983        PMID: 6196329     DOI: 10.1007/bf00489777

Source DB:  PubMed          Journal:  Histochemistry        ISSN: 0301-5564


  15 in total

1.  Quantitation of DNA by cytofluorometry of the conventional Feulgen reaction.

Authors:  G Prenna; S Leiva; G Mazzini
Journal:  Histochem J       Date:  1974-09

2.  A microspectrofluorometer with epi-illumination operated under computer control.

Authors:  J S Ploem; J A De Sterke; J Bonnet; H Wasmund
Journal:  J Histochem Cytochem       Date:  1974-07       Impact factor: 2.479

3.  Improvement in microscope photometry by voltage to frequency conversion: analogue measurement and digital processing.

Authors:  M I Freitas; P A Giordano; G Bottiroli
Journal:  J Microsc       Date:  1981-11       Impact factor: 1.758

4.  A parameter for the distribution of fluorophores in cells derived from measurements of inner filter effect and reabsorption phenomenon.

Authors:  H J Tanke; P van Oostveldt; P van Duijn
Journal:  Cytometry       Date:  1982-05

5.  DNA content and survival in mammary carcinoma.

Authors:  G U Auer; T O Caspersson; A S Wallgren
Journal:  Anal Quant Cytol       Date:  1980-09

6.  HISTOSCAN: computer program for cytophotometry in tissue sections and its application in the evaluation of nuclear atypia.

Authors:  K Bjelkenkrantz; O Stål; O Gröntoft; J Olofsson; A Herder; H Hellquist
Journal:  Histochemistry       Date:  1981-12

7.  Optical errors in scanning stage absorbance cytophotometry. II. Application of correction factors for residual distributional error, glare and diffraction error in practical cytophotometry.

Authors:  W A Duijndam; P Van Duijn; S H Riddersma
Journal:  J Histochem Cytochem       Date:  1980-05       Impact factor: 2.479

8.  HIDACSYS: computer programs for interactive scanning cytophotometry.

Authors:  M Van der Ploeg; K Van den Broek; A W Smeulders; A M Vossepoel; P Van Duijn
Journal:  Histochemistry       Date:  1977-12-28

9.  Quantitative aspects of the cytochemical Feulgen-DNA procedure studied on model systems and cell nuclei.

Authors:  A C Van Prooijen-Knegt; C A Redi; M Van der Ploeg
Journal:  Histochemistry       Date:  1980

10.  An evaluation of Feulgen-acriflavine-SO2 and Hoechst 33258 for DNA cytofluorometry in tumour pathology.

Authors:  K Bjelkenkrantz
Journal:  Histochemistry       Date:  1983
View more
  6 in total

1.  Cytometric and histopathologic features of tumors detected in a randomized mammography screening program: correlation and relative prognostic influence.

Authors:  T Hatschek; O Gröntoft; G Fagerberg; O Stål; S Sullivan; J Carstensen; B Nordenskjöld
Journal:  Breast Cancer Res Treat       Date:  1990-05       Impact factor: 4.872

2.  Influence of S-phase fraction on metastatic pattern and post-recurrence survival in a randomized mammography screening trial.

Authors:  T Hatschek; J Carstensen; G Fagerberg; O Stål; O Gröntoft; B Nordenskjöld
Journal:  Breast Cancer Res Treat       Date:  1989-12       Impact factor: 4.872

3.  An evaluation of Feulgen-acriflavine-SO2 and Hoechst 33258 for DNA cytofluorometry in tumour pathology.

Authors:  K Bjelkenkrantz
Journal:  Histochemistry       Date:  1983

4.  A comparison of static cytofluorometry and flow cytometry for the estimation of ploidy and DNA replication in human breast cancer.

Authors:  O Stål; C Klintenberg; G Franzen; B Risberg; S Arvidsson; K Bjelkenkrantz; L Skoog; B Nordenskjöld
Journal:  Breast Cancer Res Treat       Date:  1986       Impact factor: 4.872

5.  DNA measurement--an objective predictor of response to irradiation? A review of 24 squamous cell carcinomas of the oral cavity.

Authors:  G Franzén; C Klintenberg; J Olofsson; B Risberg
Journal:  Br J Cancer       Date:  1986-05       Impact factor: 7.640

6.  DNA content, malignancy grading and prognosis in T1 and T2 oral cavity carcinomas.

Authors:  M Tytor; G Franzén; J Olofsson; U Brunk; B Nordenskjöld
Journal:  Br J Cancer       Date:  1987-11       Impact factor: 7.640

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.