We performed filtration surgery with a unidirectional, pressure-sensitive valve implant in 40 eyes with neovascular glaucoma. The device consisted of an open Supramid tube (outside diameter 0.58 mm, inside diameter 0.38 mm) sealed to a Silastic tube with a slit valve. The Supramid tube was inserted at the corneoscleral limbus 1 to 4 mm into the anterior chamber and the Silastic portion was located under a lamellar scleral flap. Twenty-seven of 40 eyes (68%) had a postoperative IOP less than or equal to 24 mm Hg with a mean follow-up of 13.8 months. Ten of these 27 eyes required postoperative medical therapy to achieve this level of IOP control. The glaucoma valve implant failed to return IOP to normal in 13 of the 40 eyes (32%), 11 of these as a consequence of scarring of the external bleb.

Entities:  

Mesh:

Year:  1980        PMID: 6154418     DOI: 10.1016/0002-9394(80)90002-1

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  14 in total

1.  A one-piece ocular drainage implant for glaucoma surgery: a preliminary report.

Authors:  M Kivalo; C Raitta; T Mononen
Journal:  Acta Vet Scand       Date:  1992       Impact factor: 1.695

2.  Pars plana vitrectomy with pars plana tube implantation in eyes with intractable glaucoma.

Authors:  S Kaynak; N F Tekin; I Durak; A T Berk; A O Saatci; M F Soylev
Journal:  Br J Ophthalmol       Date:  1998-12       Impact factor: 4.638

3.  Efficacy and safety of the Ahmed glaucoma valve implant in Chinese eyes with complicated glaucoma.

Authors:  J S Lai; A S Poon; J K Chua; C C Tham; A T Leung; D S Lam
Journal:  Br J Ophthalmol       Date:  2000-07       Impact factor: 4.638

4.  Experimental studies of aqueous filtration using the Molteno implant.

Authors:  D S Minckler; A Shammas; M Wilcox; T E Ogden
Journal:  Trans Am Ophthalmol Soc       Date:  1987

5.  Investigations of the reasons for success and failure in the anterior shunt-to-the-encircling-band procedure in the treatment of refractory glaucoma.

Authors:  S S Schocket
Journal:  Trans Am Ophthalmol Soc       Date:  1986

6.  Single stage insertion of the Molteno tube for glaucoma and modifications to reduce postoperative hypotony.

Authors:  J E Hoare Nairne; D Sherwood; J S Jacob; W J Rich
Journal:  Br J Ophthalmol       Date:  1988-11       Impact factor: 4.638

7.  A new method for the evacuation of aqueous humor in uncontrolled glaucoma. The vitreo-tenonian tube.

Authors:  J Haut; P Larricart; Y Le Mer; E Abboud
Journal:  Int Ophthalmol       Date:  1987-02       Impact factor: 2.031

8.  Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification.

Authors:  C E Traverso; F De Feo; A Messas-Kaplan; P Denis; S Levartovsky; E Sellem; F Badalà; Z Zagorski; A Bron; S Gandolfi; M Belkin
Journal:  Br J Ophthalmol       Date:  2005-04       Impact factor: 4.638

9.  The Gonioseton, a surgical treatment for chronic glaucoma.

Authors:  G van der Veen; W L Jongebloed; J G Worst
Journal:  Doc Ophthalmol       Date:  1990-10       Impact factor: 2.379

10.  The outcome of the Ahmad Glaucoma Valve implantation for refractory glaucoma in Kuwait.

Authors:  Adel M Aljazzaf; Sidky M A Abdelmoaty; Abdelmutalib H Behbehani; Ahmad A Abdulmuez; Hadeel A Aljazzaf
Journal:  Saudi J Ophthalmol       Date:  2012-10-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.