Studies were performed to determine the possible influence of physiologic motion on the parenchymal intensity of organs in magnetic resonance (MR) images. It is known that periodic motion associated with respiration and cardiac function causes characteristic artifacts in spin-warp images. The present study shows that bulk motion can also cause striking intensity changes at velocities equivalent to the craniocaudal respiratory excursion of organs in the upper abdomen. The magnitude of the effect depends on the velocity and direction of motion with respect to the three orthogonal axes of the imager and on the technical details of the imager and pulse sequence. Large systematic errors in calculated tissue relaxation times are possible due to this phenomenon. The findings have important implications for clinical imaging because motion can cause artifactual changes in the gray-scale relationships among tissues. Some pulse sequences are much less sensitive to these effects. These results provide guidance for selecting MR techniques that reduce the detrimental effect of respiratory and other physiologic motion on examinations of the upper abdomen and thorax.

Mesh:

Year:  1986        PMID: 3704156     DOI: 10.1148/radiology.159.3.3704156

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  13 in total

1.  High-resolution human diffusion tensor imaging using 2-D navigated multishot SENSE EPI at 7 T.

Authors:  Ha-Kyu Jeong; John C Gore; Adam W Anderson
Journal:  Magn Reson Med       Date:  2012-05-16       Impact factor: 4.668

2.  Feasibility of self-gated isotropic radial late-phase MR imaging of the liver.

Authors:  Jakob Weiss; Jana Taron; Ahmed E Othman; Robert Grimm; Matthias Kuendel; Petros Martirosian; Christer Ruff; Christina Schraml; Konstantin Nikolaou; Mike Notohamiprodjo
Journal:  Eur Radiol       Date:  2016-06-08       Impact factor: 5.315

3.  Young Investigator Award presentation at the 13th annual meeting of the ESMRMB, September 1996, Prague. Quantification of pulmonary water compartments by magnetic resonance.

Authors:  J Lehmann; J C Böck; P Podrabsky; W Wlodarczyk; R Felix
Journal:  MAGMA       Date:  1997-03       Impact factor: 2.310

4.  Exploring the sensitivity of magnetic resonance fingerprinting to motion.

Authors:  Zidan Yu; Tiejun Zhao; Jakob Assländer; Riccardo Lattanzi; Daniel K Sodickson; Martijn A Cloos
Journal:  Magn Reson Imaging       Date:  2018-09-05       Impact factor: 2.546

5.  Efficacy of the radial acquisition regime (RADAR) for acquiring head and neck MR images.

Authors:  Tsukasa Kojima; Hidetake Yabuuchi; Hiroshi Narita; Seiji Kumazawa; Yuzo Yamasaki; Yuki Yano; Noriyuki Sakai; Yoichi Kurihara; Kiyoshi Hisada; Masafumi Masaki; Hiroshi Kimura
Journal:  Br J Radiol       Date:  2016-09-16       Impact factor: 3.039

6.  Subject-specific estimation of respiratory navigator tracking factor for free-breathing cardiovascular MR.

Authors:  Mehdi H Moghari; Peng Hu; Kraig V Kissinger; Beth Goddu; Lois Goepfert; Long Ngo; Warren J Manning; Reza Nezafat
Journal:  Magn Reson Med       Date:  2011-08-29       Impact factor: 4.668

7.  Halting the effects of flow enhancement with effective intermittent zeugmatographic encoding (HEFEWEIZEN) in SSFP.

Authors:  Jamal J Derakhshan; Mark A Griswold; Jeffrey L Sunshine; Jeffrey L Duerk
Journal:  J Magn Reson Imaging       Date:  2009-05       Impact factor: 4.813

8.  Liver MR imaging: comparison of respiratory triggered fast spin echo with T2-weighted spin-echo and inversion recovery.

Authors:  M T Keogan; C E Spritzer; E K Paulson; S S Paine; L Harris; J L Dahlke; J R MacFall
Journal:  Abdom Imaging       Date:  1996 Sep-Oct

9.  Primary malignant liver tumors in childhood: assessment of resectability with high-field MR and comparison with CT.

Authors:  J P Finn; M A Hall-Craggs; C Dicks-Mireaux; L Spitz; E R Howard; J Pritchard; G M Vergani
Journal:  Pediatr Radiol       Date:  1990

10.  Brain imaging in the unsedated pediatric patient: comparison of periodically rotated overlapping parallel lines with enhanced reconstruction and single-shot fast spin-echo sequences.

Authors:  Kirsten P Forbes; James G Pipe; John P Karis; Victoria Farthing; Joseph E Heiserman
Journal:  AJNR Am J Neuroradiol       Date:  2003-05       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.